Edge-enhancement densenet for X-ray fluoroscopy image denoising in cardiac electrophysiology procedures

被引:5
|
作者
Luo, Yimin [1 ]
Ma, Yingliang [2 ]
O' Brien, Hugh [1 ]
Jiang, Kui [3 ]
Kohli, Vikram [1 ]
Maidelin, Sesilia [1 ]
Saeed, Mahrukh [1 ]
Deng, Emily [1 ]
Pushparajah, Kuberan [1 ]
Rhode, Kawal S. [1 ]
机构
[1] Kings Coll London, Sch Biomed Engn & Imaging Sci, London SE1 7EH, England
[2] Coventry Univ, Sch Comp Elect & Math, Coventry, W Midlands, England
[3] Wuhan Univ, Sch Comp Sci, Wuhan, Peoples R China
基金
英国工程与自然科学研究理事会;
关键词
cardiac electrophysiology procedures; convolutional neural network; denoising; edge enhancement; X-ray fluoroscopy; NOISE-REDUCTION; NETWORK;
D O I
10.1002/mp.15426
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Reducing X-ray dose increases safety in cardiac electrophysiology procedures but also increases image noise and artifacts which may affect the discernibility of devices and anatomical cues. Previous denoising methods based on convolutional neural networks (CNNs) have shown improvements in the quality of low-dose X-ray fluoroscopy images but may compromise clinically important details required by cardiologists. Methods In order to obtain denoised X-ray fluoroscopy images whilst preserving details, we propose a novel deep-learning-based denoising framework, namely edge-enhancement densenet (EEDN), in which an attention-awareness edge-enhancement module is designed to increase edge sharpness. In this framework, a CNN-based denoiser is first used to generate an initial denoising result. Contours representing edge information are then extracted using an attention block and a group of interacted ultra-dense blocks for edge feature representation. Finally, the initial denoising result and enhanced edges are combined to generate the final X-ray image. The proposed denoising framework was tested on a total of 3262 clinical images taken from 100 low-dose X-ray sequences acquired from 20 patients. The performance was assessed by pairwise voting from five cardiologists as well as quantitative indicators. Furthermore, we evaluated our technique's effect on catheter detection using 416 images containing coronary sinus catheters in order to examine its influence as a pre-processing tool. Results The average signal-to-noise ratio of X-ray images denoised with EEDN was 24.5, which was 2.2 times higher than that of the original images. The accuracy of catheter detection from EEDN denoised sequences showed no significant difference compared with their original counterparts. Moreover, EEDN received the highest average votes in our clinician assessment when compared to our existing technique and the original images. Conclusion The proposed deep learning-based framework shows promising capability for denoising interventional X-ray fluoroscopy images. The results from the catheter detection show that the network does not affect the results of such an algorithm when used as a pre-processing step. The extensive qualitative and quantitative evaluations suggest that the network may be of benefit to reduce radiation dose when applied in real time in the catheter laboratory.
引用
收藏
页码:1262 / 1275
页数:14
相关论文
共 50 条
  • [1] Ultra-DenseNet for Low-Dose X-Ray Image Denoising in Cardiac Catheter-Based Procedures
    Luo, Yimin
    Toth, Daniel
    Jiang, Kui
    Pushparajah, Kuberan
    Rhode, Kawal
    [J]. STATISTICAL ATLASES AND COMPUTATIONAL MODELS OF THE HEART: MULTI-SEQUENCE CMR SEGMENTATION, CRT-EPIGGY AND LV FULL QUANTIFICATION CHALLENGES, 2020, 12009 : 31 - 42
  • [2] A Device Enhancing and Denoising Algorithm for X-ray Cardiac Fluoroscopy
    Bismuth, Vincent
    Vaillant, Regis
    [J]. 19TH INTERNATIONAL CONFERENCE ON PATTERN RECOGNITION, VOLS 1-6, 2008, : 1802 - 1805
  • [3] A Comparison of Denoising Algorithms for Effective Edge Detection in X-Ray Fluoroscopy
    Andreozzi, Emilio
    Pirozzi, Maria Agnese
    Sarno, Antonio
    Esposito, Daniele
    Cesarelli, Mario
    Bifulco, Paolo
    [J]. XV MEDITERRANEAN CONFERENCE ON MEDICAL AND BIOLOGICAL ENGINEERING AND COMPUTING - MEDICON 2019, 2020, 76 : 405 - 413
  • [4] Improvement of quantitative performance of imaging x-ray microscope by reduction of edge-enhancement effect
    Takeuchi, A.
    Uesugi, K.
    Suzuki, Y.
    [J]. X-RAY MICROSCOPY CONFERENCE 2016 (XRM 2016), 2017, 849
  • [5] Characterization of the phase-contrast radiography edge-enhancement effect in a cabinet x-ray system
    Donnelly, EF
    Lewis, KG
    Wolske, KM
    Pickens, DR
    Price, RR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (01): : 21 - 30
  • [6] Novel registration-based image enhancement for x-ray fluoroscopy
    Dixon, Adam
    Areste, Romain
    Jabri, Kadri N.
    Walimbe, Vivek
    [J]. MEDICAL IMAGING 2010: IMAGE PROCESSING, 2010, 7623
  • [7] Edge-enhancement fourier transform X-ray microscopy using a Laguerre-Gaussian zoneplate
    Nándor Bokor
    Andrew Domondon
    Yoshinori Iketaki
    [J]. Optical Review, 2010, 17 : 79 - 83
  • [8] Edge-enhancement fourier transform X-ray microscopy using a Laguerre-Gaussian zoneplate
    Bokor, Nandor
    Domondon, Andrew
    Iketaki, Yoshinori
    [J]. OPTICAL REVIEW, 2010, 17 (02) : 79 - 83
  • [9] LEARNING A MULTISCALE PATCH-BASED REPRESENTATION FOR IMAGE DENOISING IN X-RAY FLUOROSCOPY
    Matviychuk, Yevgen
    Mailhe, Boris
    Chen, Xiao
    Wang, Qiu
    Kiraly, Atilla
    Strobel, Norbert
    Nadar, Mariappan
    [J]. 2016 IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING (ICIP), 2016, : 2330 - 2334
  • [10] MRI to X-ray Fluoroscopy Overlay for Guidance of Cardiac Resynchronization Therapy Procedures
    Ma, YingLiang
    Duckett, Simon
    Chinchapatnam, Phani
    Gao, Gang
    Shetty, Anoop
    Rinaldi, C. Aldo
    Schaeffter, Tobias
    Rhode, Kawal S.
    [J]. COMPUTING IN CARDIOLOGY 2010, VOL 37, 2010, 37 : 229 - 232