Micro thermal conductivity detector with flow compensation using a dual MEMS device

被引:14
|
作者
de Graaf, G. [1 ]
Prouza, A. Abarca [2 ]
Ghaderi, M. [2 ]
Wolffenbuttel, R. F. [2 ]
机构
[1] Delft Univ Technol, Fac Elect Engn, Delft, Netherlands
[2] Delft Univ Technol, Delft, Netherlands
关键词
Thermal conductivity detector; Flow compensation; Hydrogen sensor; Helium sensor; CO2; sensor; MEMS; TCD; GAS; SENSORS;
D O I
10.1016/j.sna.2016.08.019
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A generic method to reduce the in-line flow dependence of thermal conductivity detectors (TCDs) is presented. The principle is based on a dual-MEMS device configuration. Two thin-film sensors on membranes in parallel in the gas stream on the same chip are differentially operated. Both micro-TCDs are designed to be identical in terms of contact with the main gas flow, however a different depth of the detection chamber results in a different response to the thermal conductivity of the sample gas. Static and dynamic simulations have been performed to characterize the design of the fabricated structures. Devices have been fabricated in a MEMS process using a combined surface- and bulk micromachining process. The devices have been characterized statically and dynamically. Measurements on prototypes show that depending on the range of gases, device size and flow range device the effect of flow on the thermal conductivity can be reduced by a factor 4-15. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:186 / 198
页数:13
相关论文
共 50 条
  • [41] Dual-Purpose Photometric-Conductivity Detector for Simultaneous and Sequential Measurements in Flow Analysis
    Mantim, Thitirat
    Chaisiwamongkhol, Korbua
    Uraisin, Kanchana
    Hauser, Peter C.
    Wilairat, Prapin
    Nacapricha, Duangjai
    MOLECULES, 2020, 25 (10):
  • [42] Development of Wall Flow Sensor Using Micro Imaging Device
    Lee, Seung Hwan
    Kim, Byung Soo
    Kim, Hyoung-Bum
    TRANSACTIONS OF THE KOREAN SOCIETY OF MECHANICAL ENGINEERS B, 2012, 36 (12) : 1217 - 1222
  • [43] DETECTION OF INDOMETHACIN, AFTER HALOGENATION AND ESTERIFICATION, USING A HALL MICRO-CONDUCTIVITY DETECTOR
    WILKINSON, DR
    AUTREY, C
    HOLT, A
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1977, 173 (MAR20): : 12 - 12
  • [44] Flow control using MEMS pulsed micro-jets on the Ahmed body
    Joseph, Pierric
    Amandolese, Xavier
    Edouard, Christophe
    Aider, Jean-Luc
    EXPERIMENTS IN FLUIDS, 2013, 54 (01)
  • [45] Flow control using MEMS pulsed micro-jets on the Ahmed body
    Pierric Joseph
    Xavier Amandolese
    Christophe Edouard
    Jean-Luc Aider
    Experiments in Fluids, 2013, 54
  • [46] Simultaneous thermal conductivity and diffusivity sensing in liquids using a micromachined device
    Kuntner, J.
    Kohl, F.
    Jakoby, B.
    SENSORS AND ACTUATORS A-PHYSICAL, 2006, 130 : 62 - 67
  • [47] GAS CHROMATOGRAPHIC DETECTOR RESPONSE USING CARRIER GASES OF LOW THERMAL CONDUCTIVITY
    MILLER, JM
    LAWSON, AE
    ANALYTICAL CHEMISTRY, 1965, 37 (11) : 1348 - &
  • [48] Using infrared opacifiers to reduce the thermal conductivity of micro and nanocellular polymethylmethacrylate
    Martin-de Leon, J.
    Sillero, A.
    Rodriguez-Perez, M. A.
    POLYMER, 2024, 290
  • [49] Two-Port Static Coated Micro Gas Chromatography Column With an Embedded Thermal Conductivity Detector
    Shree Narayanan
    Alfeeli, Bassam
    Agah, Masoud
    IEEE SENSORS JOURNAL, 2012, 12 (06) : 1893 - 1900
  • [50] Thermostatted Micro Gas Chromatography Column with On-chip Thermal Conductivity Detector for Elevated Temperature Separation
    Narayanan, Shree
    Alfeeli, Bassam
    Agah, Masoud
    2010 IEEE SENSORS, 2010, : 2504 - 2507