The BIPM Compact Josephson Voltage Standard

被引:2
|
作者
Solve, Stephane [1 ]
Chayramy, Regis [1 ]
机构
[1] Int Bur Weights & Measures BIPM, F-92312 Sevres, France
关键词
International Bureau of Weights and Measures (BIPM) direct on-site comparison; Josephson voltage standard (JVS);
D O I
10.1109/TIM.2011.2108610
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Fifteen National Metrology Institutes have recently confirmed their interest in participating in the International Bureau of Weights and Measures (BIPM) direct on-site comparison of Josephson array voltage standards at the level of 10 V. This comparison has been ongoing since 1994, and the BIPM primary standard involved needs to be replaced by a new standard built from components based on modern technology. We present here in detail the components of the new BIPM Josephson voltage standard. The main goal during its development was to realize a standard that is more easily transportable, more convenient to use, and as efficient and reliable as its predecessor. Both primary standards have already shown agreement within a total combined relative uncertainty of 1.4 x 10(-10) (k = 1).
引用
收藏
页码:2366 / 2371
页数:6
相关论文
共 50 条
  • [21] Comparison of the Josephson voltage standards of the NIM (China) and the BIPM
    Reymann, D
    Xue, SQ
    Liu, RM
    Li, HH
    Gao, J
    Zhou, YQ
    Witt, TJ
    METROLOGIA, 1996, 33 (05) : 475 - 478
  • [22] Recent developments in BIPM voltage standard comparisons
    Reymann, D
    Witt, TJ
    Vrabcek, P
    Tang, YH
    Hamilton, CA
    Katkov, AS
    Jeanneret, B
    Power, O
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2001, 50 (02) : 206 - 209
  • [23] Recent developments in BIPM voltage standard comparisons
    Reymann, D
    Witt, TJ
    Vrabcek, P
    Tang, YH
    Hamilton, CA
    Katkov, A
    Mendeleyev, DI
    Jeanneret, B
    Power, O
    2000 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2000, : 253 - 254
  • [24] A compact comparison standard based on the Josephson effect with an output voltage of 10 V
    A. S. Katkov
    V. E. Lovtsyus
    Measurement Techniques, 2011, 54 : 801 - 807
  • [25] A compact high-precision voltage standard based on programmable Josephson Array
    Gao, Y.
    Li, H.
    2008 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2008, : 364 - 365
  • [26] A COMPACT COMPARISON STANDARD BASED ON THE JOSEPHSON EFFECT WITH AN OUTPUT VOLTAGE OF 10 V
    Katkov, A. S.
    Lovtsyus, V. E.
    MEASUREMENT TECHNIQUES, 2011, 54 (07) : 801 - 807
  • [27] Josephson voltage standard - a review
    Hamilton, Clark A.
    Burroughs, Charles J.
    Benz, Samuel P.
    IEEE Transactions on Applied Superconductivity, 1997, 7 (2 pt 3): : 3756 - 3761
  • [28] Transportable Josephson voltage standard
    Katkov, A
    Lovtsus, V
    Behr, R
    Niemeyer, J
    2002 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS, CONFERENCE DIGEST, 2002, : 488 - 489
  • [29] Josephson voltage standard - A review
    Hamilton, CA
    Burroughs, CJ
    Benz, SP
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1997, 7 (02) : 3756 - 3761
  • [30] Comparison of the Josephson Voltage Standards of the INTI and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.a)
    Solve, S.
    Chayramy, R.
    Stock, M.
    Iuzzolino, R.
    Tonina, A.
    METROLOGIA, 2012, 49