Measurement of CO2 by Wavelength Modulated Reinjection Off-Axis Integrated Cavity Output Spectroscopy at 2 μm

被引:9
|
作者
Yuan, Zihao [1 ,2 ,3 ]
Huang, Yinbo [1 ,3 ]
Lu, Xingji [1 ,3 ]
Huang, Jun [1 ,2 ,3 ]
Liu, Qiang [1 ,3 ]
Qi, Gang [1 ,2 ,3 ]
Cao, Zhensong [1 ,2 ,3 ]
机构
[1] Chinese Acad Sci, Anhui Inst Opt & Fine Mech, HFIPS, Key Lab Atmospher Opt, Hefei 230031, Peoples R China
[2] Univ Sci & Technol China, Grad Sch, Sci Isl Branch, Hefei 230026, Peoples R China
[3] Adv Laser Technol Lab Anhui Prov, Hefei 230037, Peoples R China
关键词
off-axis integrated cavity output spectroscopy; reinjection; wavelength modulation spectroscopy; CO2; detection; ENHANCEMENT;
D O I
10.3390/atmos12101247
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
A high sensitivity wavelength modulated reinjection off-axis integrated cavity output spectroscopy (WM-RE-OA-ICOS) experimental setup was built at a 2 mu m band. On the basis of an off-axis integrated output spectroscopy (OA-ICOS), combined with an optical reinjection (RE) approach to improve signal intensity, and wavelength modulation spectroscopy (WMS) to improve the signal-to-noise ratio (SNR) of the system, the experimental study of trace CO2 with high sensitivity was carried out using the setup. The performance was compared and evaluated, and the results show that: Compared with the OA-ICOS, the wavelength modulated reinjection OA-ICOS enhanced the signal intensity by 6.3 times, and the SNR increased 7.2 times from 179 to 1288. The Allan variance results showed that the detection limit of the system is 0.35 ppm when the average system time is 230 s. The setup was used to measure the indoor CO2 concentration for a long time (22 h), and the measured results were in line with the actual concentration change. The proposed method shows good performance enhancement for the OA-ICOS system in trace gas measurements.
引用
下载
收藏
页数:12
相关论文
共 50 条
  • [41] Measurement of δ18O and δ2H of water and ethanol in wine by Off-Axis Integrated Cavity Output Spectroscopy and Isotope Ratio Mass Spectrometry
    Wang, Xing
    Jansen, Henk G.
    Duin, Haico
    Meijer, Harro A. J.
    EUROPEAN FOOD RESEARCH AND TECHNOLOGY, 2021, 247 (08) : 1899 - 1912
  • [42] Measurement of δ18O and δ2H of water and ethanol in wine by Off-Axis Integrated Cavity Output Spectroscopy and Isotope Ratio Mass Spectrometry
    Xing Wang
    Henk G. Jansen
    Haico Duin
    Harro A. J. Meijer
    European Food Research and Technology, 2021, 247 : 1899 - 1912
  • [43] Spectral contaminant identifier for off-axis integrated cavity output spectroscopy measurements of liquid water isotopes
    Leen, J. Brian
    Berman, Elena S. F.
    Liebson, Lindsay
    Gupta, Manish
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2012, 83 (04):
  • [44] Environmental and Atmospheric Monitoring Using Off-Axis Integrated Cavity Output Spectroscopy (OA-ICOS)
    Baer, Douglas
    Gupta, Manish
    Leen, J. Brian
    Berman, Elena
    AMERICAN LABORATORY, 2012, 44 (10) : 20 - 23
  • [45] Optical parametric oscillator based off-axis integrated cavity output spectroscopy for rapid chemical sensing
    Arslanov, Denis D.
    Cristescu, Simona M.
    Harren, Frans J. M.
    OPTICS LETTERS, 2010, 35 (19) : 3300 - 3302
  • [46] Simultaneous δ2H and δ18O analyses of water inclusions in halite with off-axis integrated cavity output spectroscopy
    Fourel, Francois
    Lecuyer, Christophe
    Jame, Patrick
    Guironnet, Alexandre
    Boutier, Antoine
    Barbier, Manon
    Blamey, Nigel
    Brand, Uwe
    Fralick, Philip
    JOURNAL OF MASS SPECTROMETRY, 2020, 55 (10):
  • [47] H2S trace concentration measurements using off-axis integrated cavity output spectroscopy in the near-infrared
    W. Chen
    A.A. Kosterev
    F.K. Tittel
    X. Gao
    W. Zhao
    Applied Physics B, 2008, 90 : 311 - 315
  • [48] H2S trace concentration measurements using off-axis integrated cavity output spectroscopy in the near-infrared
    Chen, W.
    Kosterev, A. A.
    Tittel, F. K.
    Gao, X.
    Zhao, W.
    APPLIED PHYSICS B-LASERS AND OPTICS, 2008, 90 (02): : 311 - 315
  • [49] Near-infrared acetylene sensor system using off-axis integrated-cavity output spectroscopy and two measurement schemes
    Zheng, Kaiyuan
    Zheng, Chuantao
    He, Qixin
    Yao, Dan
    Hu, Lien
    Zhang, Yu
    Wang, Yiding
    Tittel, Frank K.
    OPTICS EXPRESS, 2018, 26 (20): : 26205 - 26216
  • [50] Measurement of Low-Concentration Water Vapor Based on Off-Axis Integrated Cavity Absorption Spectroscopy
    Yang Huinan
    Chen Ning
    Chen Jun
    ACTA OPTICA SINICA, 2018, 38 (02)