共 50 条
Microwave-assisted dilute acid extraction of trace metals from biological samples for atomic absorption spectrometric determination
被引:23
|作者:
Zhou, CY
Wong, MK
Koh, LL
Wee, YC
机构:
[1] NATL UNIV SINGAPORE,DEPT CHEM,SINGAPORE 119260,SINGAPORE
[2] NATL UNIV SINGAPORE,DEPT BOT,SINGAPORE 119260,SINGAPORE
关键词:
microwave-assisted acid extraction;
trace metal;
biological sample;
atomic absorption spectrometry;
D O I:
10.1039/ja9961100585
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
A microwave extraction method with dilute nitric acid for biological samples was developed using a pressure feedback microwave extraction system, The concentration of nitric acid, pressure setting and extraction time were optimized by a mixed level orthogonal array design, The interactions among the selected parameters for the extraction were investigated, The effect of sample mass was also studied, Six replicate analyses of MST SRM 1515 Apple Leaves were performed under the optimized extraction conditions (5 ml of 14% v/v nitric acid; pressure setting, 1104 kPa; extraction time, 30 min; sample mass, 0.3 g), Eight elements, viz., Ca, Cu, Fe, K, Mg, Mn, Ni and Zn, were determined by FAAS or ETAAS, Recoveries of 96-103% mere achieved for all the elements, The RSDs of the test elements were less than 3.8%, except for Ni, The extraction method was also employed to extract four other biological SRMs and CRMs. With the exception of Ni, the recoveries of the eight test elements were 91-107% in all the reference materials.
引用
收藏
页码:585 / 590
页数:6
相关论文