Instrumentation for neutron sources; Neutron detectors (cold; thermal; fast neutrons); Radiation monitoring; Scintillators; scintillation and light emission processes (solid; gas and liquid scintillators); PULSE-SHAPE DISCRIMINATION; PLASTIC SCINTILLATORS; NEUTRON DETECTION; GAMMA; ENVIRONMENT; DETECTOR; ARRAYS; CF-252;
D O I
10.1088/1748-0221/12/12/P12007
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Monte-Carlo modelling (MCNPX) methods have been employed to conduct an investigation into a suitable scintillator and coded-aperture material for a scintillator based mixed-field radiation imaging system. Single stilbene crystal, pure and Li-6-loaded plastic scintillators were simulated and their neutron/gamma detection performance compared when exposed to the spontaneous fission spectrum produced by Cf-252. The most suitable candidate was then incorporated into a scintillator based mixed-field coded-aperture imaging system. Coded-aperture models made of three W and Cd-113 compositions were tested in different neutron/gamma environments with a square W collimator modelled around the aperture. Each simulation involved recording the interactions of neutron events in organic solid scintillator, whose neutron/gamma detection performance was assessed prior to the coded-aperture material investigation. Three coded-aperture material compositions have been tested with the simulated Cf-252 spontaneous fission as well as (AmBe)-Am-241 neutron sources. Results generally claim very good detection sensitivity and spatial resolution for the radioactive sources located in the centre of the aperture.
引用
收藏
页数:16
相关论文
共 35 条
  • [31] Design of an Ultra-Near-Field System for Planar Coded Aperture Nuclear Medicine Imaging
    Starfield, D. M.
    Rubin, D. M.
    Marwala, T.
    14TH NORDIC-BALTIC CONFERENCE ON BIOMEDICAL ENGINEERING AND MEDICAL PHYSICS, 2008, 20 : 590 - 593
  • [32] Challenges of Employing a High Resolution X-ray Detector in a Coded-Aperture X-ray Phase-Contrast Imaging system
    Pil-Ali, Abdollah
    Adnani, Sahar
    Scott, Christopher C.
    Karim, Karim S.
    MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [33] DEPTH DOSE DETERMINATION FOR A MIXED RADIATION FIELD USING A THIN PLASTIC SCINTILLATOR DOSIMETRY SYSTEM
    Aydarous, Abdulkadir
    Waker, Anthony
    NUCLEAR TECHNOLOGY & RADIATION PROTECTION, 2012, 27 (01): : 20 - 27
  • [34] Field Trial of a Highly Portable Coded Aperture Gamma Ray and 3D Imaging System
    Boehnen, Chris
    Paquit, Vincent
    Ziock, Klaus
    Guzzardo, Tyler
    Whitaker, Michael
    Raffo-Caiado, Ana
    2011 FUTURE OF INSTRUMENTATION INTERNATIONAL WORKSHOP (FIIW), 2011,
  • [35] Extended depth of field of a diffraction limited imaging system using a spatial light modulator based intensity compensated polarization coded aperture
    Tiwari, Vipin
    Bisht, Nandan S. .
    OPTICS CONTINUUM, 2023, 2 (01): : 1 - 8