Evaluation of the Abbott Cell-Dyn Sapphire hematology analyzer

被引:8
|
作者
Park, Younhee [1 ]
Song, Jaewoo [1 ]
Song, Sungwook [1 ]
Song, Kyung Soon [1 ]
Ahn, Mee Suk [1 ]
Yang, Mi-Sook [1 ]
Kim, Il [1 ]
Choi, Jong Rak [1 ]
机构
[1] Yonsei Univ, Coll Med, Dept Lab Med, Seoul, South Korea
来源
KOREAN JOURNAL OF LABORATORY MEDICINE | 2007年 / 27卷 / 03期
关键词
Cell-Dyn Sapphire; performance; correlation;
D O I
10.3343/kjlm.2007.27.3.162
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The performance of Cell-Dyn Sapphire (Abbott Diagnostic, USA) was compared to the Bayer Advia 2120 (Bayer Diagnostics, USA), Sysmex XE-2100 (Sysmex Corporation, Japan), and reference microscopy. Methods: Three hundred samples for routine CBC and WBC differentials were randomly chosen for a comparison analysis. The Cell-Dyn Sapphire system was evaluated according to the linearity, imprecision, inter-instrument correlations, and white blood cell differential. Results: The CBC parameters (WBC, RBC, hemoglobin and platelet) showed a significant linearity with correlation coefficients greater than 0.99 (P < 0.0001). Coefficients of variation (CV) for within-run and differential count of WBC were less than 5% except for Total CV for monocytes, eosinophils, and basophils and within-run CV for low valued eosinophils. The correlation coefficients with manual count were lower in monocytes, eosinophils, and basophils than in neutrophils and lymphocytes. The correlation with other hematology anlayzers was significant exclusive of basophils. Conclusions : These results demonstrate that the Cell-Dyn Sapphire has a good linearity, an acceptable reproducibility, a minimal carryover, and a comparable performance with the sysmex XE-2100 and Advia 2120.
引用
收藏
页码:162 / 168
页数:7
相关论文
共 50 条
  • [1] Evaluation of the Abbott CELL-DYN 4000 hematology analyzer
    Grimaldi, E
    Scopacasa, F
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2000, 113 (04) : 497 - 505
  • [2] Evaluation of the Abbott Cell-DYN 3500 hematology analyzer in a University Hospital
    VivesCorrons, JL
    Besson, I
    Jou, JM
    Gutierrez, G
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1996, 105 (05) : 553 - 559
  • [3] Evaluation of the Abbott CELL-DYN 4000 hematology analyzer in a hospital.
    Komiya, M
    Kikawada, M
    Yashiro, I
    Koike, T
    Oikawa, S
    Ieiri, T
    [J]. CLINICAL CHEMISTRY, 1998, 44 : A50 - A50
  • [4] PERFORMANCE EVALUATION OF THE NEW AUTOMATED ABBOTT HEMATOLOGY ANALYZER ALINITY HQ COMPARED TO THE CELL-DYN SAPPHIRE
    Van der Beken, Yolien
    Van Dalem, Annelien
    Van Moer, Guy
    Segers, Eddy
    Damiaens, Serge
    Hoffmann, Johannes
    Jochmans, Kristin
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2018, 40 : 148 - 148
  • [5] Immunophenotypic diagnosis of lymphoproliferative disorders using the Abbott CELL-DYN sapphire hematology analyzer.
    Settegrana, Catherine
    Maloum, Karim
    Souni, Fadila
    Septier, Isabelle
    Vinet, Sandrine
    Bonnemye, Patrick
    Merle-Beral, Helene
    [J]. BLOOD, 2006, 108 (11) : 183B - 183B
  • [6] Detection of imported malaria with the Cell-Dyn Sapphire ™ hematology analyzer
    Chesney, Alden
    Cardoso-Vouzoukos, Fatima
    Phillips, Ann-Marie
    Reis, Marciano
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2007, 29 : 49 - 49
  • [7] EVALUATION OF THE CELL-DYN EMERALD 22 HEMATOLOGY ANALYZER
    Crespo, Santiago
    Antonio Royo, Jose
    Hoffmann, Johannes
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2017, 39 : 30 - 30
  • [8] PERFORMANCE EVALUATION OF THE HEMATOLOGICAL ANALYZER CELL-DYN 3000 (ABBOTT)
    AUBERTLETRILLART, B
    BINET, JL
    BAUDET, S
    AZGUI, Z
    MERLEBERAL, H
    [J]. NOUVELLE REVUE FRANCAISE D HEMATOLOGIE, 1993, 35 (06): : 535 - 540
  • [9] FETOMATERNAL HEMORRHAGE ON THE ABBOTT CELL-DYN SAPPHIRE
    Xiros, Nick
    Brown, Ross
    Joshua, Douglas
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2012, 34 : 138 - 138
  • [10] Analysis of serous body fluids using the CELL-DYN Sapphire hematology analyzer
    Keuren, Jeffrey F. W.
    Hoffmann, Johannes J. M. L.
    Leers, Mathie P. G.
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2013, 51 (06) : 1285 - 1290