X-ray electron probe microanalysis in the vicinity of M absorption edges

被引:1
|
作者
Lavrent'ev, Yu. G. [1 ]
Usova, L. V. [1 ]
机构
[1] Russian Acad Sci, Sobolev Inst Geol & Mineral, Siberian Branch, Novosibirsk 630090, Russia
关键词
Correction Method; PbTe; HgTe; Analytical Line; Quantitative Electron Probe Microanalysis;
D O I
10.1134/S1061934809100116
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Nine correction methods with five methods of calculating absorption coefficients are tested for the cases when the analytical line is located in the vicinity of the M adsorption edges of elements with atomic numbers from 76-83. Calculations were performed using the CARAT program and experiments, on the Camebacs Micro microanalyzer. It was found that none of the correction methods can maintain the appropriate accuracy within all absorption ranges. It was proposed to change the mu/rho ratio in the corresponding inner intervals, which will ensure an accuracy of the analysis at a level of 1 rel %.
引用
收藏
页码:1035 / 1041
页数:7
相关论文
共 50 条