The St. Jude medical cardiac valve prosthesis: Long-term follow up of patients having double valve replacement

被引:0
|
作者
Emery, Robert W. [1 ]
Emery, Ann M. [1 ]
Krogh, Christopher [1 ]
Shake, Jay G. [1 ]
Holter, Arlen R. [1 ]
Blake, David P. [1 ]
Raikar, Goya V. [1 ]
机构
[1] Cardiac Surg Associates, St Paul, MN USA
来源
JOURNAL OF HEART VALVE DISEASE | 2007年 / 16卷 / 06期
关键词
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and aim of the study: Between October 1977 and February 2002, a total of 343 patients (mean age 62 13 years; range: 19-91 years) underwent double valve replacement (DVR) with the St. Jude Medical (SJM) heart valve. Among the replacements, 337 (98%) were aortic and mitral in nature. Concomitant coronary artery bypass was performed in 73 patients (21%). Methods: Cardiac Surgical Associates has maintained an independent database of patients undergoing valve replacement with the SJM prosthesis since the valve's first implantation in October 1977. Patients were contacted by questionnaire and/or telephone (94% complete) between November 2002 and June 2003. The patients' hospital course and valve-related events were verified by patient chart review and/or physician contact. Results: Operative mortality was 8% (n = 29); mortality was valve-related in two cases. The mean follow up was 6.5 +/- 6.0 years (range: 1 month to 24 years); total follow up was 2,226 patient-years. Over 25 years, patient freedom from late mortality was 62%, and from valve-related mortality 78%. Freedom from thromboembolic events was 82% (93% from permanent defect), from bleeding events 76%, from endocarditis 98%, from valve thrombosis 99.9%, and from reoperation 98%. Six reoperations were carried out in five patients (2%), valve repair or replacement in five (2%), and suture closure of paravalvular leak in one patient (0.3%). There were no valve structural failures reported. Conclusion: The SJM valve has proven to be an effective and durable heart valve prosthesis. Over the long term, the event rate is low and there is excellent freedom from reoperation in the double valve configuration.
引用
收藏
页码:634 / 640
页数:7
相关论文
共 50 条
  • [31] ST JUDE MEDICAL VALVE PROSTHESIS
    WADA, J
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1981, 82 (03): : 462 - 462
  • [32] Acute mechanical valve thrombosis of the St. Jude medical prosthesis
    Katircioglu, SF
    Ulus, AT
    Yamak, B
    Özsöyler, I
    Birincioglu, L
    Tasdemir, O
    JOURNAL OF CARDIAC SURGERY, 1999, 14 (03) : 164 - 168
  • [33] Tricuspid valve replacement with the St. Jude Medical valve: 19 years of experience
    Kawano, H
    Oda, T
    Fukunaga, S
    Tayama, E
    Kawara, T
    Oryoji, A
    Aoyagi, S
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2000, 18 (05) : 565 - 569
  • [34] TRICUSPID-VALVE REPLACEMENT WITH THE BILEAFLET ST-JUDE MEDICAL VALVE PROSTHESIS
    NAKANO, K
    KOYANAGI, H
    HASHIMOTO, A
    OHTSUKA, G
    NOJIRI, C
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1994, 108 (05): : 888 - 892
  • [35] A NEW CARDIAC PROSTHESIS - ST JUDE MEDICAL CARDIAC-VALVE
    UMEDA, S
    KON, A
    YAMAZAKI, A
    NISHIWAKI, T
    TANOI, H
    MIYAMOTO, A
    OKAZAKI, S
    KITAMURA, S
    KAWANO, K
    KUMABE, T
    TAKAHASHI, K
    OKAMOTO, I
    RIKUKAWA, H
    SHINDOW, S
    AKIYAMA, K
    SHIONO, M
    SEZAI, K
    ARTIFICIAL ORGANS, 1981, 5 (03) : 341 - 341
  • [36] 10-YEAR FOLLOW-UP-STUDY OF PATIENTS WHO HAD DOUBLE VALVE-REPLACEMENT WITH THE ST-JUDE MEDICAL PROSTHESIS
    AROM, KV
    NICOLOFF, DM
    KERSTEN, TE
    NORTHRUP, WF
    LINDSAY, WG
    EMERY, RW
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1989, 98 (05): : 1008 - 1016
  • [37] Thrombogenicity and Reoperation of the St. Jude Medical Silzone® valve:: A comparison with the conventional St. Jude medical valve
    Stalenhoef, JE
    Mellema, EC
    Veeger, NJGM
    Ebels, T
    JOURNAL OF HEART VALVE DISEASE, 2003, 12 (05): : 635 - 639
  • [38] Aortic valve replacement with a small mechanical valve prosthesis - A clinical and echocardiographic study of a St Jude Medical valve prosthesis
    Aoyagi, S
    Arinaga, K
    Fukunaga, S
    Kawano, H
    Kawara, T
    JAPANESE CIRCULATION JOURNAL-ENGLISH EDITION, 1998, 62 (04): : 244 - 248
  • [39] VALVE-REPLACEMENT WITH A ST JUDE MEDICAL PROSTHESIS IN CHILDREN AND ADOLESCENTS
    CHAMPSAUR, G
    FRIEH, JP
    NINET, J
    ELKIRAT, M
    GRESSIER, M
    BRULE, P
    BOZIO, A
    SASSOLAS, F
    JOCTEURMONROZIER, D
    NORMAND, J
    ARCHIVES DES MALADIES DU COEUR ET DES VAISSEAUX, 1985, 78 (04): : 663 - 663
  • [40] Risk Factors for Cardiac Valve Replacement with St Jude Mechanical Prosthesis
    Sant'Anna, Joao R. M.
    De Bacco, Matheus
    Sant'Anna, Roberto
    Mallmann Gil, Allon Gregory
    Sartori, Ana Paula
    Santos, Marisa Fatima
    Pereira, Edemar
    Reis, Altamiro
    Kalil, Renato A. K.
    Nesralla, Ivo A.
    CIRCULATION, 2010, 122 (02) : E145 - E145