PRECiSe (priming reduced extracorporeal circulation setup): Results of a safety study

被引:0
|
作者
Beholz, S [1 ]
Kessler, M [1 ]
Konertz, WF [1 ]
机构
[1] Humboldt Univ, Charite, Dept Cardiovasc Surg, D-10117 Berlin, Germany
来源
HEART SURGERY FORUM | 2003年 / 6卷 / 05期
关键词
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Different low-priming systems limited to coronary artery bypass grafting (CABG) have been introduced. We describe the priming reduced extracorporeal circulation setup (PRECiSe), a new low-priming system that supplies all of the features of cardiopulmonary bypass. Methods: PRECiSe incorporates the DeltaStream diagonal pump, which pumps blood from the right atrium to the aorta via a membrane oxygenator and a filter. The system is placed beneath the patient's head, resulting in extremely short tubing lengths. A reservoir allows the use of suckers and vents. Autologous blood priming further reduces hemodilution. Results: In the safety study, the system was used for extracorporeal circulation in 11 patients undergoing CABG without the occurrence of any adverse events. By using the PRECiSe system, we reduced the mean priming volume to 268.5 mL, resulting in minimal hemodilution and transfusion requirements. Conclusion: The use of PRECiSe for extracorporeal circulation in CABG is safe and reduces priming volume as well as transfusion requirements. Further studies are necessary to investigate the clinical benefit for patients as well as the use of the system in open heart procedures.
引用
收藏
页码:311 / 315
页数:5
相关论文
共 50 条
  • [1] Priming reduced extracorporeal circulation setup (PRECiSe) with the DeltaStream diagonal pump
    Beholz, S
    Kessler, M
    Thölke, R
    Konertz, WF
    [J]. ARTIFICIAL ORGANS, 2003, 27 (12) : 1110 - 1115
  • [2] A new PRECiSe (priming reduced extracorporeal circulation setup) minimizes the need for blood transfusions: First clinical results in coronary artery bypass grafting
    Beholz, S
    Zheng, LL
    Kessler, M
    Rusche, M
    Konertz, W
    [J]. HEART SURGERY FORUM, 2005, 8 (03): : E132 - E135
  • [3] GLYCEROL AS PRIMING SOLUTION FOR EXTRACORPOREAL CIRCULATION
    MANDAL, AK
    CALLAGHAN, JC
    DOLAN, AM
    STERNS, LP
    [J]. CANADIAN JOURNAL OF SURGERY, 1970, 13 (01) : 86 - +
  • [4] USE OF GELATINOUS PRIMING SOLUTION FOR EXTRACORPOREAL CIRCULATION
    SILVAY, J
    SUJANSKY, E
    SCHNORRER, M
    HRUBISKOVA, K
    SLEZAK, J
    GABAUER, I
    STYK, J
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1968, 55 (03): : 350 - +
  • [5] NON-HEMIC PRIMING FLUIDS FOR EXTRACORPOREAL CIRCULATION
    PATON, BC
    ROSENKRA.J
    [J]. DISEASES OF THE CHEST, 1965, 48 (03): : 311 - &
  • [6] Effect of ultrafiltration on priming solution with preserved blood for extracorporeal circulation in infants
    Shimpo, Hideto
    Shimamoto, Akira
    Miyake, Yoichiro
    Onoda, Koji
    Tani, Kazuhiro
    Shimono, Takatsugu
    Yuasa, Hiroshi
    Yada, Isao
    [J]. ASAIO Journal, 42 (05):
  • [7] Effect of ultrafiltration on priming solution with preserved blood for extracorporeal circulation in infants
    Shimpo, H
    Shimamoto, A
    Miyake, Y
    Onoda, K
    Tani, K
    Shimono, T
    Yuasa, H
    Yada, I
    [J]. ASAIO JOURNAL, 1996, 42 (05) : M792 - M794
  • [8] Prediction of volatile anaesthetic solubility in blood and priming fluids for extracorporeal circulation
    Yu, RG
    Zhou, JX
    Liu, J
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2001, 86 (03) : 338 - 344
  • [9] CLINICAL RESULTS WITH CONTINUOUSLY MONITORED EXTRACORPOREAL CIRCULATION
    PATON, BC
    SWAN, H
    [J]. CIRCULATION, 1960, 22 (04) : 795 - 795
  • [10] Reduced amount of gaseous microemboli in the arterial line of minimized extracorporeal circulation systems compared with conventional extracorporeal circulation
    Bauer, Adrian
    Schaarschmidt, Jan
    Anastasiadis, Kyriakos
    Carrel, Thierry
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2014, 46 (01) : 152 - 152