Synthetic Aperture Imaging in Breast Ultrasound: A Preliminary Clinical Study

被引:11
|
作者
Kim, Won Hwa [1 ]
Chang, Jung Min [1 ]
Kim, Choye [2 ]
Park, Jongho [2 ]
Yoo, Yangmo [3 ]
Moon, Woo Kyung [1 ]
Cho, Nariya [1 ]
Choi, Byung Ihn [1 ]
机构
[1] Seoul Natl Univ Hosp, Dept Radiol, Seoul 100744, South Korea
[2] Sogang Univ, Dept Elect Engn, Seoul, South Korea
[3] Sogang Univ, Interdisciplinary Program Integrated Biotechnol, Seoul, South Korea
关键词
Ultrasound; synthetic aperture imaging; breast lesions; CODED EXCITATION; SONOGRAPHY; SYSTEMS; LESIONS; MOTION;
D O I
10.1016/j.acra.2012.04.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: To compare the image quality between conventional and synthetic aperture (SA) imaging in breast ultrasound (US). Materials and Methods: Twenty-four patients with 31 breast lesions were included in our study. The US data were processed with SA algorithm. For quantitative analysis, contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR) were calculated. For qualitative analysis, conventional and SA images were reviewed by three radiologists and diagnostic preference (conspicuity, margin sharpness, and contrast) was assessed. The radiologists also determined whether artifacts were present. Parameters were analyzed using a paired t-test, Wilcoxon signed-rank test, and chi-square test. Results: The mean CNRs were higher in SA images compared with conventional images (mean, 2.56 versus 2.28, P=.004). The mean SNRs were higher in SA images compared with conventional images (31.62 versus 25.26, P <.0001). SA images were considered as being "better" or "much better" in 16-23 (51.6-74.2%) lesions of total 31 lesions for conspicuity, in 17-24 (45.2%-77.4%) for margin sharpness, and in 13-23 (41.9%-74.2%) for contrast. Significant preferences in SA images were demonstrated (conspicuity, P <.05 for all radiologists; margin sharpness and contrast in two radiologists). Refraction and speckle artifacts were less frequently observed in SA images (refraction, P <.05 for all radiologists; speckle, P <.05 for two radiologists), whereas reflection artifacts were more frequent in SA images (P <.05 in two radiologists). Conclusion: SA imaging provides better image quality than conventional imaging in patients with focal breast lesions in breast US.
引用
收藏
页码:923 / 929
页数:7
相关论文
共 50 条
  • [41] Aberration correction by polynomial approximation for synthetic aperture ultrasound imaging
    Leonov, Denis
    Kulberg, Nicholas
    Yakovleva, Tatyana
    MEDICAL PHYSICS, 2024, 51 (05) : 3292 - 3308
  • [42] Signal Eigenvalue Factor for Synthetic Transmit Aperture Ultrasound Imaging
    Zheng, Chichao
    Zha, Quanchao
    Zhang, Lunan
    Peng, Hu
    IEEE ACCESS, 2018, 6 : 495 - 503
  • [43] Study on aperture configuration of optical synthetic aperture imaging system
    Research Center for Space Optical Engineering, Harbin Institute of Technology, Harbin 150001, China
    不详
    J. Harbin Inst. Technol., 2006, 1 (20-22):
  • [44] Preliminary Result of Spiral Synthetic Aperture Based Ultrasound Computed Tomography
    Zhang, Qiude
    Song, Junjie
    Liu, Zhaohui
    Liu, Kuolin
    Quan, Zhaohui
    Yang, Qi
    Ding, Mingyue
    Ming Yuchi
    MEDICAL IMAGING 2021: ULTRASONIC IMAGING AND TOMOGRAPHY, 2021, 11602
  • [45] A study of synthetic-aperture imaging with virtual source elements in B-mode ultrasound imaging systems
    Bae, MH
    Jeong, MK
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2000, 47 (06) : 1510 - 1519
  • [46] DESIGN OF A LOW COST ULTRASOUND SYSTEM USING DIVERGING BEAMS AND SYNTHETIC APERTURE APPROACH: PRELIMINARY STUDY
    Lokesh, B.
    Thittai, Arun K.
    2017 IEEE 14TH INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI 2017), 2017, : 1108 - 1111
  • [47] Research on the Multi-Element Synthetic Aperture Focusing Technique in Breast Ultrasound Imaging, Based on the Ring Array
    Wu, Yang
    Zhang, Wendong
    Shao, Xingling
    Yang, Yuhua
    Zhang, Tian
    Lei, Miao
    Wang, Zhihao
    Gao, Bizhen
    Hu, Shumin
    MICROMACHINES, 2022, 13 (10)
  • [48] An Adaptive Synthetic Aperture Method Applied to Ultrasound Tissue Harmonic Imaging
    Varnosfaderani, Mehdi Haji Heidari
    Asl, Babak Mohammadzadeh
    Faridsoltani, Saman
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2018, 65 (04) : 557 - 569
  • [49] Design of Orthogonal Coded Excitation for Synthetic Aperture Imaging in Ultrasound Systems
    Yang, Ming
    Chakrabarti, Chaitali
    2012 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS (ISCAS 2012), 2012, : 113 - 116
  • [50] Multi-Element Synthetic Transmit Aperture in Medical Ultrasound Imaging
    Trots, Ihor
    Nowicki, Andrzej
    Lewandowski, Marcin
    Tasinkevych, Yuriy
    ARCHIVES OF ACOUSTICS, 2010, 35 (04) : 687 - 699