Performance study of the Temporal Resolution Improvement using Prior Image Constrained Compressed Sensing (TRI-PICCS)

被引:1
|
作者
Tang, Jie [1 ]
Hsieh, Jiang [4 ]
Chen, Guang-Hong [1 ,2 ,3 ]
机构
[1] Univ Wisconsin, Dept Med Phys, Madison, WI 53705 USA
[2] Univ Wisconsin, Dept Radiol, Madison, WI 53705 USA
[3] Univ Wisconsin, Dept Human Oncol, Madison, WI 53705 USA
[4] GE Healthcare, Waukesha, WI 53188 USA
来源
MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING | 2010年 / 7622卷
关键词
cardiac; CT; temporal resolution; COMPUTED-TOMOGRAPHY; CT; RECONSTRUCTION; SCANS;
D O I
10.1117/12.844537
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A technique for temporal resolution improvement using prior image constrained compressed sensing (TRI-PICCS) in multi-detector computed tomography (MDCT) cardiac imaging is proposed. In this work, the performance of TRI-PICCS was studied using a hybrid phantom which consists of realistic cardiac anatomy and objects moving with designed trajectories. Several simulated moving vessels were added to different locations in the heart. Different motion directions and simulated heart rates were investigated using half of the projection data of the short-scan angular range in TRI-PICCS. Different angular ranges of projection data were also investigated in TRI-PICCS to evaluate the highest achievable temporal resolution. The results showed that the temporal improvement of TRI-PICCS is independent of the locations of the moving objects and motion directions. The motion artifacts at 100 bmp simulated heart rate can be significantly improved using TRI-PICCS compared with conventional filtered back projection (FBP). The minimum angular range requirement of TRI-PICCS is about 90 degrees, corresponding to a temporal resolution improvement factor of 2.6 compared with the standard short-scan FBP reconstruction.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] Temporal resolution improvement using PICCS in MDCT cardiac imaging
    Chen, Guang-Hong
    Tang, Jie
    Hsieh, Jiang
    MEDICAL PHYSICS, 2009, 36 (06) : 2130 - 2135
  • [32] A Performance Comparison between Prior Image Constrained Compressed Sensing and ASiR in Pediatric CT
    Tang, J.
    Chen, G-H
    MEDICAL PHYSICS, 2011, 38 (06)
  • [33] Prospective evaluation of prior image constrained compressed sensing (PICCS) algorithm in abdominal CT: a comparison of reduced dose with standard dose imaging
    Lubner, Meghan G.
    Pickhardt, Perry J.
    Kim, David H.
    Tang, Jie
    del Rio, Alejandro Munoz
    Chen, Guang-Hong
    ABDOMINAL IMAGING, 2015, 40 (01): : 207 - 221
  • [34] Prospective evaluation of prior image constrained compressed sensing (PICCS) algorithm in abdominal CT: a comparison of reduced dose with standard dose imaging
    Meghan G. Lubner
    Perry J. Pickhardt
    David H. Kim
    Jie Tang
    Alejandro Munoz del Rio
    Guang-Hong Chen
    Abdominal Imaging, 2015, 40 : 207 - 221
  • [35] Deep learning enabled prior image constrained compressed sensing (DL-PICCS) reconstruction framework for sparse-view reconstruction
    Zhang, Chengzhu
    Li, Yinsheng
    Chen, Guang-Hong
    MEDICAL IMAGING 2020: PHYSICS OF MEDICAL IMAGING, 2020, 11312
  • [36] Dramatic Noise Reduction and Potential Radiation Dose Reduction in Breast Cone-Beam CT Imaging Using Prior Image Constrained Compressed Sensing (PICCS)
    Qi, Z.
    Huang, S.
    Nett, B.
    Tang, J.
    Yang, K.
    Boone, J.
    Chen, G.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [37] Accurate Tumor Motion Delineation with 4D Cone Beam CT Using Prior Image Constrained Compressed Sensing (PICCS-4DCBCT)
    Chen, G.
    Qi, Z.
    MEDICAL PHYSICS, 2011, 38 (06) : 3770 - +
  • [38] Prior Image Constrained Compressed Sensing Metal Artifact Reduction (PICCS-MAR): 2D and 3D Image Quality Improvement with Hip Prostheses at CT Colonography
    Bannas, Peter
    Li, Yinsheng
    Motosugi, Utaroh
    Li, Ke
    Lubner, Meghan
    Chen, Guang-Hong
    Pickhardt, Perry J.
    EUROPEAN RADIOLOGY, 2016, 26 (07) : 2039 - 2046
  • [39] Prior Image Constrained Compressed Sensing Metal Artifact Reduction (PICCS-MAR): 2D and 3D Image Quality Improvement with Hip Prostheses at CT Colonography
    Peter Bannas
    Yinsheng Li
    Utaroh Motosugi
    Ke Li
    Meghan Lubner
    Guang-Hong Chen
    Perry J. Pickhardt
    European Radiology, 2016, 26 : 2039 - 2046
  • [40] Exploitation of temporal redundancy in compressed sensing reconstruction of fMRI studies with a prior-based algorithm (PICCS)
    Chavarrias, C.
    Abascal, J. F. P. J.
    Montesinos, P.
    Desco, M.
    MEDICAL PHYSICS, 2015, 42 (07) : 3814 - 3821