CD4-T cell enumeration in human immunodeficiency virus (HIV)-infected patients: A laboratory performance evaluation of Muse Auto CD4/CD4% system by World Health Organization prequalification of in vitro diagnostics

被引:3
|
作者
Ceulemans, Ann [1 ]
Bouzahzah, Chaymae [1 ]
Prat, Irena [2 ]
Urassa, Willy [2 ]
Kestens, Luc [1 ,3 ]
机构
[1] Inst Trop Med, Dept Biomed Sci, Immunol Lab, Antwerp, Belgium
[2] WHO, Geneva, Switzerland
[3] Univ Antwerp, Dept Biomed Sci, Antwerp, Belgium
来源
PLOS ONE | 2019年 / 14卷 / 01期
关键词
COUNT; AGREEMENT;
D O I
10.1371/journal.pone.0209677
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background CD4 T-cell counts are still widely used to assess treatment eligibility and follow-up of HIV-infected patients. The World Health Organization (WHO) prequalification of in vitro diagnostics requested a manufacturer independent laboratory evaluation of the analytical performance at the Institute of Tropical Medicine (ITM) Antwerp, Belgium, of the Muse Auto CD4/CD4% system (Millipore), a new small capillary-flow cytometer dedicated to count absolute CD4-T cells and percentages in venous blood samples from HIV-infected patients. Methods Two hundred and fifty (250) patients were recruited from the HIV outpatient clinic at ITM. Accuracy and precision of CD4 T cell counting on fresh EDTA anticoagulated venous blood samples were assessed in the laboratory on a Muse Auto CD4/CD4% system. Extensive precision analyses were performed both on fresh blood and on normal and low stabilized whole blood controls. Accuracy ((bias) was assessed by comparing results from Muse CD4/CD4% to the reference (single-platform FACSCalibur). Clinical misclassification was measured at 500, 350, 200 and 100 cells/mu L thresholds. Results Intra-assay precision was <5%, and inter-assay was <9%. CD4 T cell counts measured on Muse Auto CD4/CD4% System and on the reference instrument resulted in regression slopes of 0.97 for absolute counts and 1.03 for CD4 T cell percentages and a correlation coefficient of 0.99 for both. The average absolute bias as compared to the reference was negligible (4 cells/mu L or 0.5%). The absolute average bias on CD4 T cell percentages was <1%. Clinical misclassification at different CD4 T cell thresholds was small resulting in sensitivities and specificities equal or >90% at all thresholds except at 100 cells/mu L (sensitivity = 87%). All samples could be analyzed as there was no repetitive rejection errors recorded. Conclusions The Muse Auto CD4/CD4% System performed very well on fresh venous blood samples and met all WHO acceptance criteria for analytical performance of CD4 technologies.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Evaluation of the performances of the "MUSE AUTO CD4/CD4%" flow cytometer vs "GUAVA AUTO CD4/CD4%" flow cytometer for measuring the rate of CD4 lymphocytes in patients infected with HIV in Cameroon
    Kalla, Ginette Claude Mireille
    Voundi, Esther Voundi
    Tanghu, Florence Mimo
    Sike, Christiane Medi
    Noubom, Michel
    Nkodo, Joseph Mendimi
    Amougou, Justin
    Ndjolo, Alexis
    Mbopi-Keou, Francois-Xavier
    PAN AFRICAN MEDICAL JOURNAL, 2019, 32
  • [2] AUTOANTIBODIES AGAINST CD4 CELLS ARE ASSOCIATED WITH CD4 HELPER DEFECTS IN HUMAN IMMUNODEFICIENCY VIRUS-INFECTED PATIENTS
    WEIMER, R
    DANIEL, V
    ZIMMERMANN, R
    SCHIMPF, K
    OPELZ, G
    BLOOD, 1991, 77 (01) : 133 - 140
  • [3] Task-shifting of CD4 T cell count monitoring by the touchscreen-based Muse™ Auto CD4/CD4% single-platform system for CD4 T cell numeration: Implication for decentralization in resource-constrained settings
    Kouabosso, Andre
    Mossoro-Kpinde, Christian Diamant
    Bouassa, Ralph-Sydney Mboumba
    Longo, Jean De Dieu
    Simaleko, Marcel Mbeko
    Gresenguet, Gerard
    Belec, Laurent
    JOURNAL OF IMMUNOLOGICAL METHODS, 2018, 455 : 95 - 98
  • [4] SOLUBLE CD4 LEVELS IN HUMAN-IMMUNODEFICIENCY-VIRUS INFECTED PATIENTS
    REDDY, MM
    JOURNAL OF INFECTIOUS DISEASES, 1993, 168 (01): : 246 - 246
  • [5] Association of virus load, CD4 cell count, and treatment with clinical progression in human immunodeficiency virus-infected patients with very low CD4 cell counts
    Miller, V
    Phillips, AN
    Clotet, B
    Mocroft, A
    Ledergerber, B
    Kirk, O
    Ormaasen, V
    Gargalianos-Kakolyris, P
    Vella, S
    Lundgren, JD
    JOURNAL OF INFECTIOUS DISEASES, 2002, 186 (02): : 189 - 197
  • [6] WHO Multicenter Evaluation of FACSCount CD4 and Pima CD4 T-Cell Count Systems: Instrument Performance and Misclassification of HIV-Infected Patients
    Wade, Djibril
    Daneau, Geraldine
    Aboud, Said
    Vercauteren, Gaby H.
    Urassa, Willy S. K.
    Kestens, Luc
    JAIDS-JOURNAL OF ACQUIRED IMMUNE DEFICIENCY SYNDROMES, 2014, 66 (05) : E98 - E107
  • [7] Lenalidomide enhancement of human T cell functions in human immunodeficiency virus (HIV)-infected and HIV-negative CD4 T lymphocytopenic patients
    Lim, H.
    Kane, L.
    Schwartz, J. B.
    Hesdorffer, C. S.
    Deeks, S. G.
    Greig, N.
    Ferrucci, L.
    Goetzl, E. J.
    CLINICAL AND EXPERIMENTAL IMMUNOLOGY, 2012, 169 (02): : 182 - 189
  • [8] Neurophysiological parameters and the number of CD4 lymphocytes in patients infected with human immunodeficiency virus
    Aznar-Bueno, C
    Abad-Alegria, F
    Amiguet-Garcia, JA
    REVISTA DE NEUROLOGIA, 1998, 27 (159) : 789 - 792
  • [9] QUALITY STANDARD FOR THE ENUMERATION OF CD4(+) LYMPHOCYTES IN ADULTS AND ADOLESCENTS INFECTED WITH HUMAN-IMMUNODEFICIENCY-VIRUS
    GROSS, PA
    PHAIR, JP
    KAPLAN, JE
    HOLMES, KK
    MASUR, H
    CLINICAL INFECTIOUS DISEASES, 1995, 21 : S126 - S127
  • [10] CD4 T-Cell Enumeration in a Field Setting: Evaluation of CyFlow Counter Using the CD4 Easy Count Kit-Dry and Pima CD4 Systems
    Wade, Djibril
    Diaw, Papa Alassane
    Daneau, Geraldine
    Camara, Makhtar
    Dieye, Tandakha Ndiaye
    Mboup, Souleymane
    Kestens, Luc
    PLOS ONE, 2013, 8 (09):