Effect of an Axial Magnetic Field on the Performance of a Quadrupole Mass Spectrometer

被引:9
|
作者
Syed, Sarfaraz U. [1 ]
Sreekumar, Jeyan [1 ]
Brkic, Boris [1 ]
Gibson, John R. [1 ]
Taylor, Stephen [1 ]
机构
[1] Univ Liverpool, Dept Elect Engn & Elect, Liverpool L69 3GJ, Merseyside, England
关键词
ION TRAJECTORIES; FILTER; SIMULATION; TRANSMISSION; RESOLUTION; HARMONICS;
D O I
10.1016/j.jasms.2010.08.020
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We consider the case of a quadrupole mass spectrometer (QMS) in which a static magnetic field is applied axially in the z-direction along the length of the mass filter The theoretical approach assumed in the model is that the QMS contains hyperbolic rods as electrodes and that the magnetic field acts over the full length of the mass filter assembly Initial experimental results with argon and helium for a low-resolution instrument confirm the predicted theoretical trends The analysis also predicts for which values of operating parameters an enhancement of the instrument resolution is achieved when an axial magnetic field is applied The model predicts instrument resolution R >3000 for a QMS with a 200 mm long mass filter via application of an axial magnetic field (J Am Soc Mass Spectrom 2010, 21, 2070-2076) (C) 2010 Published by Elsevier Inc on behalf of American Society for Mass Spectrometry
引用
收藏
页码:2070 / 2076
页数:7
相关论文
共 50 条
  • [31] AXIAL INTRODUCTION OF LASER-DESORBED IONS INTO A QUADRUPOLE ION TRAP MASS-SPECTROMETER
    MCINTOSH, A
    DONOVAN, T
    BRODBELT, J
    ANALYTICAL CHEMISTRY, 1992, 64 (18) : 2079 - 2083
  • [32] Cold cathode field emitter array on a quadrupole mass spectrometer: Route to miniaturization
    Felter, TE
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1999, 17 (05): : 1993 - 1996
  • [33] Reduction of measurement error due to the interference effect in a quadrupole mass spectrometer
    Nemanic, Vincenc
    Zumer, Marko
    Lakner, Mitja
    MEASUREMENT, 2021, 184
  • [34] Simulation of the collisional cooling effect in a quadrupole ion trap mass spectrometer
    Wu, HF
    Chen, LW
    Lin, YP
    JOURNAL OF THE CHINESE CHEMICAL SOCIETY, 1999, 46 (06) : 923 - 932
  • [35] Pulse mode of the quadropole mass spectrometer in a magnetic field
    Pavlenko, YG
    Toropova, AI
    RADIOTEKHNIKA I ELEKTRONIKA, 1997, 42 (07): : 890 - 896
  • [36] Theoretic tests on a mass spectrometer without a magnetic field
    Herzog, R
    Mattauch, J
    ANNALEN DER PHYSIK, 1934, 19 (04) : 345 - 386
  • [37] A NOVEL QUADRUPOLE, QUISTOR, QUADRUPOLE TANDEM MASS-SPECTROMETER
    KOFEL, P
    REINHARD, H
    SCHLUNEGGER, UP
    ORGANIC MASS SPECTROMETRY, 1991, 26 (05): : 463 - 467
  • [38] A Quadrupole Mass Spectrometer for Resolution of Low Mass Isotopes
    Sreekumar, Jeyan
    Hogan, Thomas J.
    Taylor, Stephen
    Turner, Phillip
    Knott, Christopher
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2010, 21 (08) : 1364 - 1370
  • [39] MASS DISCRIMINATION EFFECTS IN A QUADRUPOLE MASS-SPECTROMETER
    WOOD, KV
    GRANGE, AH
    TAYLOR, JW
    ANALYTICAL CHEMISTRY, 1978, 50 (12) : 1652 - 1654
  • [40] IMPROVED PERFORMANCE OF A TANDEM QUADRUPOLE TIME-OF-FLIGHT MASS-SPECTROMETER
    GLISH, GL
    MCLUCKEY, SA
    MCKOWN, HS
    ANALYTICAL INSTRUMENTATION, 1987, 16 (01): : 191 - 206