In vitro validation of new Fourier-domain optical coherence tomography

被引:39
|
作者
Tahara, Satoko [1 ,2 ]
Bezerra, Hiram G. [1 ,2 ]
Baibars, Motaz [1 ,2 ]
Kyono, Hiroyuki [1 ,2 ]
Wang, Wei [1 ,2 ]
Pokras, Stan [1 ,2 ]
Mehanna, Emile [1 ,2 ]
Petersen, Christopher L. [3 ]
Costa, Marco A. [1 ,2 ]
机构
[1] Univ Hosp Case Med Ctr, Harrington McLaughlin Heart & Vasc Inst, Cleveland, OH 44106 USA
[2] Case Western Reserve Univ, Cleveland, OH 44106 USA
[3] LightLab Imaging, Westford, MA USA
关键词
Fourier-domain optical coherence tomography; validation study; phantom model; INTRAVASCULAR ULTRASOUND; ELUTING STENTS; COVERAGE; APPOSITION;
D O I
10.4244/EIJV6I7A149
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims: To validate the accuracy and evaluate the intra- and inter-observer variability of vascular measurements using novel Fourier-domain optical coherence tomography (FD-OCT) versus intravascular ultrasound (IVUS) in a coronary phantom model. Methods and results: A dedicated cylindrical phantom comprised of four sections with differing dimensions was used to compare images generated by five FD-OCT and five IVUS catheters. Each FD-OCT catheter was tested using three different consoles, generating 15 imaging pullbacks. Two independent experts, blinded to the phantom data, performed measurements. There were 180 FD-OCT and 60 IVUS cross-sectional measurements of mean lumen diameter (MeanLD), minimal lumen diameter (MLD), maximum lumen diameter (MaxLD) and lumen area (LA) at three points in each section. FD-OCT measurements had excellent correlations with IVUS (concordance correlation coefficient [CCC]: >= 0.9769 for MeanLD, LA or longitudinal length [LL]) and phantom dimensions (CCCs for FD-OCT: >= 0.9958 for MeanLD, LA or LL). FD-OCT measurements were larger than IVUS (p<0.0001), but showed less measurement errors compared to IVUS (p<0.0001). Moreover, FD-OCT caused less discrepancy between MaxLD and MLD versus IVUS (p<0.0001). Intra- and inter-observer variability was low for both FD-OCT (CCCs for MeanLD, LA and LL >= 0.9996) and IVUS (>= 0.9935). Image catheter position did not influence FD-OCT measurements. Conclusions: FD-OCT was more accurate than IVUS and had similar high reproducibility to determine vascular dimensions in vitro. These results support the use of FD-OCT in the clinical setting.
引用
收藏
页码:875 / 882
页数:8
相关论文
共 50 条
  • [1] Fourier phase in Fourier-domain optical coherence tomography
    Uttam, Shikhar
    Liu, Yang
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2015, 32 (12) : 2286 - 2306
  • [2] Pachymetric mapping with Fourier-domain optical coherence tomography
    Li, Yan
    Tang, Maolong
    Zhang, Xinbo
    Salaroli, Camila H.
    Ramos, Jose L.
    Huang, David
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (05): : 826 - 831
  • [3] Time-domain interpolation for Fourier-domain optical coherence tomography
    Zhang, Yudong
    Li, Xiqi
    Wei, Ling
    Wang, Kai
    Ding, Zhihua
    Shi, Guohua
    [J]. OPTICS LETTERS, 2009, 34 (12) : 1849 - 1851
  • [4] Characterization of coated optical fibers by Fourier-domain optical coherence tomography
    Jasapara, J
    Wielandy, S
    [J]. OPTICS LETTERS, 2005, 30 (09) : 1018 - 1020
  • [5] Macular thickness of Chinese with fourier-domain optical coherence tomography
    Yu Weihong
    Zheng Lin
    Dai Rongping
    Zhang Zhiqiao
    Dong Fangtian
    [J]. ACTA OPHTHALMOLOGICA, 2011, 89 (01) : E104 - E105
  • [6] Corneal power measurement with Fourier-domain optical coherence tomography
    Tang, Maolong
    Chen, Alex
    Li, Yan
    Huang, David
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (12): : 2115 - 2122
  • [7] Reproducibility of intracoronary Fourier-domain optical coherence tomography analysis
    Orii, M.
    Kubo, T.
    Kitabata, H.
    Ishibashi, K.
    Yamano, T.
    Tanimoto, T.
    Ino, Y.
    Yamaguchi, T.
    Imanishi, T.
    Akasaka, T.
    [J]. EUROPEAN HEART JOURNAL, 2012, 33 : 1000 - 1000
  • [8] Fourier-domain optical coherence tomography: next step in optical imaging
    Wojtkowski, M
    Kowalczyk, A
    Targowski, P
    Gorczynska, I
    [J]. OPTICA APPLICATA, 2002, 32 (04) : 569 - 580
  • [9] Full-field Fourier-domain optical coherence tomography
    Zvyagin, AV
    Blazkiewicz, P
    Vintrou, J
    [J]. COHERENCE DOMAIN OPTICAL METHODS AND OPTICAL COHERENCE TOMOGRAPHY IN BIOMEDICINE IX, 2005, 5690 : 44 - 53
  • [10] Time-gated Fourier-domain optical coherence tomography
    Muller, Matthew S.
    Webster, Paul J. L.
    Fraser, James M.
    [J]. OPTICS LETTERS, 2007, 32 (22) : 3336 - 3338