Analysis of Run-in-Stage Wear Behavior and Contact Mechanics of Metal-on-Metal Hip Joint Bearings with Different Radial Clearances

被引:14
|
作者
Chen, Yan [1 ]
Li, Yunping [2 ,3 ]
Kurosu, Shingo [3 ]
Meng, Qingen [4 ]
Tang, Ning [3 ]
Koizumi, Yuichiro [3 ]
Chiba, Akihiko [3 ]
机构
[1] Tohoku Univ, Dept Mat Sci & Engn, Sendai, Miyagi 9808579, Japan
[2] Cent S Univ, Mat Sci & Engn, Changsha 410083, Hunan, Peoples R China
[3] Tohoku Univ, Inst Mat Res, Sendai, Miyagi 9808577, Japan
[4] Univ Leeds, Sch Mech Engn, Inst Med & Biol Engn, Leeds LS2 9JT, W Yorkshire, England
关键词
joint prostheses; CoCrMo alloy; radial clearance; contact mechanics; ELASTOHYDRODYNAMIC LUBRICATION; RESURFACING ARTHROPLASTY; SIMULATOR; DIAMETER; REPLACEMENTS; PROSTHESES; DEBRIS; IMPLANTS; SURFACES; FRICTION;
D O I
10.2320/matertrans.M2014440
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
We systematically elucidate the effects of radial clearance on the wear behavior of hip joint prostheses bearings during run-in-stage. The results indicated that bearings with smaller radial clearances exhibited lower wear rate and less abrasive wear characterized by mild surface roughness and sphericity. The contact mechanics and lubrication regime of MOM bearings with different radial clearances were also analyzed, indicating that bearings with radial clearance of similar to 20 mu m was more beneficial for small contact pressures even the enlarged clearance made the bearings operate under a full-fluid-to-mixed lubrication during wear. The agreement between the experimental results of wear rate (mm(3)/Mc) and contact mechanics and lubrication analysis indicated that the radial clearance significantly affected the wear behavior of the MOM bearings.
引用
收藏
页码:826 / 834
页数:9
相关论文
共 50 条
  • [21] Friction moments of large metal-on-metal hip joint bearings and other modern designs
    Bishop, N. E.
    Waldow, F.
    Morlock, M. M.
    MEDICAL ENGINEERING & PHYSICS, 2008, 30 (08) : 1057 - 1064
  • [22] Hip simulator wear comparison of metal-on-metal, ceramic-on-ceramic and crosslinked UHMWPE bearings
    Essner, A
    Sutton, K
    Wang, A
    WEAR, 2005, 259 : 992 - 995
  • [23] Influence of hip joint simulator design and mechanics on the wear and creep of metal-on-polyethylene bearings
    Ali, Murat
    Al-Hajjar, Mazen
    Partridge, Susan
    Williams, Sophie
    Fisher, John
    Jennings, Louise M.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2016, 230 (05) : 389 - 397
  • [24] Influence of Lubrication Performance on Wear Factor in Metal-on-Metal Hip Joint Replacement using Numerical Analysis
    Harun, Muhamad Noor
    Jin, Zhongmin
    Syahrom, Ardiyansyah
    INTERNATIONAL TRIBOLOGY CONFERENCE MALAYSIA 2013, 2013, 68 : 95 - 101
  • [25] The wear of high-carbon metal-on-metal bearings after different heat treatments
    Kinbrum, A.
    Unsworth, A.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2008, 222 (H6) : 887 - 895
  • [26] Surface-engineered metal-on-metal bearings improve the friction and wear properties of local area contact in total joint arthroplasty
    Koseki, Hironobu
    Shindo, Hiroyuki
    Baba, Koumei
    Fujikawa, Takamasa
    Sakai, Nobuo
    Sawae, Yoshinori
    Murakami, Teruo
    SURFACE & COATINGS TECHNOLOGY, 2008, 202 (19): : 4775 - 4779
  • [27] A Meta-Analysis of Design- and Manufacturing-Related Parameters Influencing the Wear Behavior of Metal-on-Metal Hip Joint Replacements
    Kretzer, Jan Philippe
    Kleinhans, Jennifer A.
    Jakubowitz, Eike
    Thomsen, Marc
    Heisel, Christian
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2009, 27 (11) : 1473 - 1480
  • [28] Analysis of mixed lubrication mechanism in metal-on-metal hip joint replacements
    Jin, ZM
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2002, 216 (H1) : 85 - 89
  • [29] Explant analysis of the Biomet Magnum/ReCap metal-on-metal hip joint
    Scholes, S. C.
    Hunt, B. J.
    Richardson, V. M.
    Langton, D. J.
    Smith, E.
    Joyce, T. J.
    BONE & JOINT RESEARCH, 2017, 6 (02): : 113 - 122
  • [30] Can cobalt levels estimate in-vivo wear of metal-on-metal bearings used in hip arthroplasty?
    Khan, M.
    Kuiper, J. H.
    Richardson, J. B.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H8) : 929 - 942