Expanding a single-lead mobile electrocardiographic device to multiple-lead recordings improves diagnostic accuracy and confidence

被引:1
|
作者
Junarta, Joey [1 ]
Frisch, Daniel R. [1 ]
Dikdan, Sean [1 ]
Weiss, Max [1 ]
Khan, Oneib [1 ]
Sarkar, Kumar [1 ]
机构
[1] Thomas Jefferson Univ Hosp, Dept Med, Philadelphia, PA 19107 USA
关键词
KardiaMobile; ECG; Mobile electrocardiogram; Electrophysiology; ALIVECOR HEART MONITOR;
D O I
10.1016/j.jelectrocard.2021.05.006
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Mobile electrocardiographic (mECG) devices that record ECG lead I have been used to detect atrial fibrillation. Other arrhythmias may not be readily diagnosed with one lead. Obtaining multi-lead tracings from an mECG (MLmECG) to simulate a 12-lead ECG may lead to more accurate diagnoses. Methods: We developed a method to generate multi-lead ECGs using a mECG device by attaching it with alligator clips connected to an insulated copper wire to adhesive electrodes on the patient's limbs and torso according to standard lead configurations. Different rhythm and conduction abnormalities from a sample of inpatients were collected. Arrhythmias were recorded in three ways (single lead, MLmECG, and standard 12-lead) and grouped by category. Recordings were sent to cardiology fellows in the form of a multiple choice survey. Participants were asked for their diagnosis and confidence in their decision. Results: Survey response rate was 100%. Single-lead, MLmECG, and 12-lead yielded 48.2%, 81.6%, and 88.6% of agreement with the correct diagnosis, respectively (single-lead vs. MLmECG or 12-lead; p < 0.01). Overall mean confidence scores were 3.34, 4.35, and 4.53 out of 5, for single-lead, MLmECG, and 12-lead ECG, respectively (single-lead vs. MLmECG or 12-lead; p < 0.01, MLmECG vs. 12-lead; p = 0.09). Conclusion: The diagnostic accuracy of MLmECGs were similar to that of a standard 12-lead ECG. Fellows' confi-dence in their diagnosis were similar between MLmECG or 12-lead ECG, and higher with both modalities com-pared to a single-lead tracing. The ability to recreate, as fully as possible, a standard 12-lead ECG is a reasonable goal for mobile technology. (c) 2021 Elsevier Inc. All rights reserved.
引用
收藏
页码:77 / 83
页数:7
相关论文
共 24 条
  • [1] Improved accuracy and confidence with multiple-lead recordings from a single-lead mobile electrocardiographic device
    Frisch, Daniel R.
    Weiss, Max
    Dikdan, Sean J.
    Keith, Scott W.
    Sarkar, Kumar
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2019, 42 (09): : 1191 - 1196
  • [2] Improved Diagnostic Accuracy and Reader Confidence with Multiple-Lead Recordings from a Single-Lead Mobile Electrocardiographic Device
    Frisch, Daniel R.
    Weiss, Max
    Dikdan, Sean
    Sarkar, Kumar
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (03) : 790 - 790
  • [3] Multiple-lead recordings from a commercial single-lead mobile electrocardiographic device
    Sarkar, Kumar
    Frisch, Daniel
    Weiss, Max
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2019, 30 (09) : 1755 - 1756
  • [4] RELIABILITY OF SINGLE-LEAD AND MULTIPLE-LEAD ELECTROCARDIOGRAPHY DURING AND AFTER EXERCISE
    SKETCH, MH
    NAIR, CK
    ESTERBROOKS, DJ
    MOHIUDDIN, SM
    CHEST, 1978, 74 (04) : 394 - 401
  • [5] Diagnostic accuracy of a single-lead portable ECG device for measuring QTc prolongation
    Bekker, Charlotte L.
    Noordergraaf, Fauve
    Teerenstra, Steven
    Pop, Gheorghe
    van den Bemt, Bart J. F.
    ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2020, 25 (01)
  • [6] Accuracy of the Apple Watch single-lead ECG recordings in pre-term neonates
    Paech, Christian
    Kobel, Maria
    Michaelis, Anna
    Gebauer, Roman Antonin
    Kalden, Philipp
    Dahnert, Ingo
    Thome, Ulrich
    Markel, Franziska
    Rutzel, Sebastian
    CARDIOLOGY IN THE YOUNG, 2022, 32 (10) : 1633 - 1637
  • [7] Accuracy of a smartwatch based single-lead electrocardiogram device in detection of atrial fibrillation
    Rajakariar, Kevin
    Koshy, Anoop N.
    Sajeev, Jithin K.
    Nair, Sachin
    Roberts, Louise
    Teh, Andrew W.
    HEART, 2020, 106 (09) : 665 - 670
  • [8] Accuracy of a smartwatch based single-lead electrocardiogram device in screening for atrial fibrillation
    Rajakariar, K.
    Koshy, A. N.
    Sajeev, J. K.
    Nair, S.
    Roberts, L.
    Teh, A. W.
    EUROPEAN HEART JOURNAL, 2019, 40 : 185 - 185
  • [9] Diagnostic Accuracy of a Smartphone-Operated, Single-Lead Electrocardiography Device for Detection of Rhythm and Conduction Abnormalities in Primary Care
    Himmelreich, Jelle C. L.
    Karregat, Evert P. M.
    Lucassen, Wirn A. M.
    van Weert, Henk C. P. M.
    de Groot, Joris R.
    Handoko, M. Louis
    Nijveldt, Robin
    Harskamp, Ralf E.
    ANNALS OF FAMILY MEDICINE, 2019, 17 (05) : 403 - 411
  • [10] Diagnostic validation of smart wearable device embedded with single-lead electrocardiogram for arrhythmia detection
    Niu, Yonghong
    Wang, Hao
    Wang, Hong
    Zhang, Hui
    Jin, Zhigeng
    Guo, Yutao
    DIGITAL HEALTH, 2023, 9