Selective determination of ammonium ions by high-speed ion-exclusion chromatography on a weakly basic anion-exchange resin column

被引:30
|
作者
Mori, M
Tanaka, K
Helaleh, MIH
Xu, Q
Ikedo, M
Ogura, Y
Sato, S
Hu, WZ
Hasebe, K
机构
[1] Natl Inst & Adv Ind Sci & Technol Seto, Aichi 4890884, Japan
[2] Chubu Univ, Natl Inst Adv Ind Sci & Technol Seto, Aichi 4890884, Japan
[3] Chubu Univ, Grad Sch Engn, Aichi 4890884, Japan
[4] Tosoh Corp, Yamaguchi 7468501, Japan
[5] Hokkaido Univ, Kita Ku, Sapporo, Hokkaido 0608010, Japan
关键词
water analysis; environmental analysis; ammonium ions; inorganic cations; tetrabutylammonium hydroxide;
D O I
10.1016/S0021-9673(03)00625-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This paper describes an ion-exclusion chromatographic system for the rapid and selective determination of ammonium ion. The optimized ion-exclusion chromatographic system was established with a polymethacrylate-based weakly basic anion-exchange resin column (TSKgel DEAE-5PW) as the separation column, an aqueous solution containing 0.05 mM tetramethylammonium hydroxide (pH 9.10) as eluent with conductimetric detection for the analyte determination. Under the optimum chromatographic conditions, ammonium ion was determined within 2.3 min with a detection limit (S/N = 3) better than 0.125 muM. Ammonium ion in rain and river waters was precisely determined using this ion-exclusion chromatographic system. (C) 2003 Elsevier Science B.V. All rights reserved.