Bioengineering reasons for the failure of metal-on-metal hip prostheses AN ENGINEER'S PERSPECTIVE

被引:61
|
作者
Fisher, J. [1 ]
机构
[1] Univ Leeds, Inst Med & Biol Engn, Sch Mech Engn, Leeds LS2 9JT, W Yorkshire, England
来源
基金
英国工程与自然科学研究理事会;
关键词
WEAR DEBRIS; CUP ANGLE; MICROSEPARATION; REPLACEMENTS; BEARINGS; FRICTION; LUBRICATION;
D O I
10.1302/0301-620X.93B8.26936
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Bioengineering reasons for increased wear and failure of metal-on-metal (MoM) bearings in hip prostheses have been described. Low wear occurs in MoM hips when the centre of the femoral head is concentric with the centre of the acetabular component and the implants are correctly positioned. Translational or rotational malpositioning of the components can lead to the contact-patch of the femoral component being displaced to the rim of the acetabular component, resulting in a ten- to 100-fold increase in wear and metal ion levels. This may cause adverse tissue reactions, loosening of components and failure of the prosthesis.
引用
下载
收藏
页码:1001 / 1004
页数:4
相关论文
共 50 条
  • [31] Predictivity and Fate of Metal Ion Release From Metal-On-Metal Total Hip Prostheses
    Nicolli, Annamaria
    Bisinella, Gianluca
    Padovani, Giovanni
    Vitella, Antonio
    Chiara, Federica
    Trevisan, Andrea
    JOURNAL OF ARTHROPLASTY, 2014, 29 (09): : 1763 - 1767
  • [32] Metal release from hip prostheses: biological monitoring of cobalt and chromium in patients with metal-on-metal hip arthroplasty
    Torra, Merce
    Sandalinas, Silvia
    Martin, Mayte
    Nogue, Santiago
    TRACE ELEMENTS AND ELECTROLYTES, 2017, 34 (01) : 1 - 4
  • [33] METAL-ON-METAL HIP IMPLANTS BMJ's reply to DePuy's criticisms of article on safety of metal-on-metal hip implants
    Cohen, Deborah
    BRITISH MEDICAL JOURNAL, 2012, 344
  • [34] Investigation of Taper Failure in a Contemporary Metal-on-Metal Hip Arthroplasty System Through Examination of Unused and Explanted Prostheses
    Langton, David
    Ahmed, Imran
    Avery, Peter
    Bone, Martin
    Cooke, Nicholas
    Deehan, David
    Duffy, Patrick
    Foguet, Pedro
    Green, Stephen
    Holland, James
    Jafri, Anwar
    Longstaff, Lee
    Lord, James
    Loughead, Jonathan
    Meek, R. M. Dominic
    Murray, Helen
    Nanu, Ananda
    Nargol, A. V. F.
    Scholes, Susan Claire
    Sidaginamale, Raghavendra Prasad
    Waller, Susan
    Joyce, Thomas
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2017, 99 (05): : 427 - 436
  • [35] The relationship between activity and ions in patients with metal-on-metal bearing hip prostheses
    Heisel, C
    Silva, M
    Skipor, AK
    Jacobs, JJ
    Schmalzried, TP
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (04): : 781 - 787
  • [36] Cobalt toxicity - an emerging clinical problem in patients with metal-on-metal hip prostheses?
    Mao, Xinzhan
    Wong, Andrew A.
    Crawford, Ross W.
    MEDICAL JOURNAL OF AUSTRALIA, 2011, 194 (12) : 649 - 651
  • [37] Clinical and Wear Analyses of 9 Large Metal-on-Metal Total Hip Prostheses
    Koper, M. C.
    Mathijssen, N. M. C.
    Witt, F.
    Morlock, M. M.
    Vehmeijer, S. B. W.
    PLOS ONE, 2016, 11 (10):
  • [38] The risk of cardiac failure following metal-on-metal hip arthroplasty
    Sabah, S. A.
    Moon, J. C.
    Jenkins-Jones, S.
    Morgan, C. L. I.
    Currie, C. J.
    Wilkinson, J. M.
    Porter, M.
    Captur, G.
    Henckel, J.
    Chaturvedi, N.
    Kay, P.
    Skinner, J. A.
    Hart, A. H.
    Manisty, C.
    BONE & JOINT JOURNAL, 2018, 100B (01): : 20 - 27
  • [39] Failure rates of stemmed metal-on-metal hip replacements reply
    Smith, Alison J.
    Dieppe, Paul
    Vernon, Kelly
    Porter, Martyn
    Blom, Ashley W.
    LANCET, 2012, 380 (9837): : 106 - 106
  • [40] Metal-on-Metal Hip Resurfacing
    Quesada, Mario J.
    Marker, David R.
    Mont, Michael A.
    JOURNAL OF ARTHROPLASTY, 2008, 23 (07): : 69 - 73