Design of a Semi-Implantable Hearing Device for Direct Acoustic Cochlear Stimulation

被引:25
|
作者
Bernhard, Hans [1 ]
Stieger, Christof [2 ]
Perriard, Yves [3 ]
机构
[1] Helbling Tech Bern AG, Med Device Dept, CH-3097 Liebefeld, Switzerland
[2] Univ Bern, Inselspital, Artorg Ctr Biomed Engn, Artificial Hearing Res Dept,Dept ENT Head & Neck, CH-3010 Bern, Switzerland
[3] Ecole Polytech Fed Lausanne, Integrated Actuators Lab, CH-2002 Neuchatel, Switzerland
关键词
Actuator design; balanced armature actuator; hearing implant; middle ear transfer function; mixed hearing loss; HUMAN MIDDLE-EAR;
D O I
10.1109/TBME.2010.2087756
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new hearing therapy based on direct acoustic cochlear stimulation was developed for the treatment of severe to profound mixed hearing loss. The device efficacy was validated in an initial clinical trial with four patients. This semi-implantable investigational device consists of an externally worn audio processor, a percutaneous connector, and an implantable microactuator. The actuator is placed in the mastoid bone, right behind the external auditory canal. It generates vibrations that are directly coupled to the inner ear fluids and that, therefore, bypass the external and the middle ear. The system is able to provide an equivalent sound pressure level of 125 dB over the frequency range between 125 and 8000 Hz. The hermetically sealed actuator is designed to provide maximal output power by keeping its dimensions small enough to enable implantation. A network model is used to simulate the dynamic characteristics of the actuator to adjust its transfer function to the characteristics of the middle ear. The geometry of the different actuator components is optimized using finite-element modeling.
引用
收藏
页码:420 / 428
页数:9
相关论文
共 50 条
  • [11] Semi-implantable transcutaneous bone conduction hearing devices
    Siegert, R.
    Kanderske, J.
    HNO, 2014, 62 (07) : 502 - 508
  • [12] Vibrant semi-implantable hearing device with digital sound processing - Effective gain and speech perception
    Snik, AFM
    Cremers, CWRJ
    ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2001, 127 (12) : 1433 - 1437
  • [13] Mixed-signal IC for a semi-implantable hearing aid
    Case Western Reserve Univ, Cleveland, United States
    Proc IEEE Int Symp Circuits Syst, (I-186 - I-189):
  • [14] CT pre-operative planning of a new semi-implantable bone conduction hearing device
    Eric K. C. Law
    Kunwar S. S. Bhatia
    Willis S. S. Tsang
    Michael C. F. Tong
    Lin Shi
    European Radiology, 2016, 26 : 1686 - 1695
  • [15] A mixed-signal IC for a semi-implantable hearing aid
    Amantia, PJ
    Kane, MJ
    Kimball, EA
    Garverick, SL
    ISCAS '99: PROCEEDINGS OF THE 1999 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS, VOL 1: VLSI, 1999, : 186 - 189
  • [16] CT pre-operative planning of a new semi-implantable bone conduction hearing device
    Law, Eric K. C.
    Bhatia, Kunwar S. S.
    Tsang, Willis S. S.
    Tong, Michael C. F.
    Shi, Lin
    EUROPEAN RADIOLOGY, 2016, 26 (06) : 1686 - 1695
  • [17] Long-term results using a piezoelectric semi-implantable middle ear hearing device - The Rion Device E-type
    Yanagihara, N
    Sato, H
    Hinohira, Y
    Gyo, K
    Hori, K
    OTOLARYNGOLOGIC CLINICS OF NORTH AMERICA, 2001, 34 (02) : 389 - +
  • [18] Functional results with active middle ear implant or semi-implantable bone conduction device in patients with comparable hearing loss
    Spiegel, Jennifer L.
    Weiss, Bernhard G.
    Bertlich, Mattis
    Stoycheva, Ivelina
    Canis, Martin
    Ihler, Friedrich
    INTERNATIONAL JOURNAL OF AUDIOLOGY, 2022, 61 (10) : 859 - 867
  • [19] Direct acoustic cochlear stimulation for therapy of severe to profound mixed hearing loss. Codacs™ Direct Acoustic Cochlear Implant System
    Lenarz, T.
    Schwab, B.
    Maier, H.
    Kludt, E.
    HNO, 2014, 62 (07) : 481 - 489
  • [20] Preoperative Headband Assessment for Semi-Implantable Bone Conduction Hearing Devices in Conductive Hearing Loss: Is it Useful or Misleading?
    Rainsbury, James W.
    Williams, Blair A.
    Gulliver, Mark
    Morris, David P.
    OTOLOGY & NEUROTOLOGY, 2015, 36 (02) : e58 - e62