Assessing the biomechanical properties of nitinol staples in normal, osteopenic and osteoporotic bone models: A finite element analysis

被引:8
|
作者
Curenton, Tanetta L. [1 ]
Davis, Brian L. [2 ]
Darnley, James E. [3 ]
Weiner, Scott D. [3 ]
Owusu-Danquah, Josiah S. [4 ]
机构
[1] Cleveland State Univ, Dept Chem & Biomed Engn, Cleveland, OH 44115 USA
[2] Cleveland State Univ, Dept Mech Engn, Cleveland, OH 44115 USA
[3] Summa Hlth Syst, Dept Orthopaed Surg, Akron, OH 44304 USA
[4] Cleveland State Univ, Dept Civil & Environm Engn, Cleveland, OH 44115 USA
基金
美国国家科学基金会;
关键词
Nitinol staples; Contact force; Fixation; Finite element simulations; Bone mechanical properties; FRACTURE FIXATION;
D O I
10.1016/j.injury.2021.08.006
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: Bone staples are internal fixation devices that are frequently used in the foot, ankle, and hand to provide stabilization. Fixation stability is vital after fusion or fracture surgeries to ensure proper bone healing. Patients undergoing surgeries that require fixation to keep bones aligned and stable may present with diminishing bone mechanical properties, and this may compromise the ability of the fixation hardware to maintain a stable construct. The purpose of this study was to investigate the mechanical performance of shape memory and superelastic nitinol bone staples with different bridge geometries in normal, osteopenic, and osteoporotic bone models. Contact forces and maximum principal stress and strain in the bone were recorded. Methods: Finite element simulations of a bone staple fixation procedure were performed to examine the initial and post-surgery contact force, as well as the maximum principal stress and strain of 15 mm bridge and 20 mm bridge staple-bone constructs. Results: Shape memory nitinol staples exhibited higher contact forces compared to superelastic nitinol staples. Nitinol bone staples with 20 mm bridge lengths displayed higher contact forces and lower stresses in all bone types, as well as lower strains in osteoporotic bone models compared to nitinol staples with a 15 mm bridge length. Conclusion: Nitinol bone staple constructs with 20 mm bridge length staples provide higher contact forces and display lower stresses in the bone than 15 mm bridge staple-bone constructs, which may be beneficial in bone with diminishing mechanical properties. Both superelastic and shape memory effect nitinol staples provide adequate compression and stress relief. However, if osteopenia is present, shape memory effect nitinol staples with a 20 mm bridge length may provide more stress relief and compression, if the bone anatomy allows. (C) 2021 Elsevier Ltd. All rights reserved.
引用
收藏
页码:2820 / 2826
页数:7
相关论文
共 50 条
  • [1] Biomechanical properties of normal and osteoporotic avian bone
    Knott, L
    Whitehead, CC
    Bailey, AJ
    JOURNAL OF BONE AND MINERAL RESEARCH, 1997, 12 (09) : P48 - P48
  • [2] Biomechanical analysis of sandwich vertebrae in osteoporotic patients: finite element analysis
    Huang, Shaolong
    Zhou, Chengqiang
    Zhang, Xu
    Tang, Zhongjian
    Liu, Liangyu
    Meng, Xiao
    Xue, Cheng
    Tang, Xianye
    FRONTIERS IN ENDOCRINOLOGY, 2023, 14
  • [3] Comparison of the biomechanical effects of lumbar disc degeneration on normal patients and osteoporotic patients: A finite element analysis
    Zhang, Xin-Ying
    Han, Ye
    MEDICAL ENGINEERING & PHYSICS, 2023, 112
  • [4] Biomechanical effects of metastasis in the osteoporotic lumbar spine: A Finite Element Analysis
    Giuseppe Salvatore
    Alessandra Berton
    Hugo Giambini
    Mauro Ciuffreda
    Pino Florio
    Umile Giuseppe Longo
    Vincenzo Denaro
    Andrew Thoreson
    Kai-Nan An
    BMC Musculoskeletal Disorders, 19
  • [5] Biomechanical effects of metastasis in the osteoporotic lumbar spine: A Finite Element Analysis
    Salvatore, Giuseppe
    Berton, Alessandra
    Giambini, Hugo
    Ciuffreda, Mauro
    Florio, Pino
    Longo, Umile Giuseppe
    Denaro, Vincenzo
    Thoreson, Andrew
    An, Kai-Nan
    BMC MUSCULOSKELETAL DISORDERS, 2018, 19
  • [6] The determination and evaluation of Nitinol constitutive models for finite element analysis
    McCummiskey, E.
    Dempster, W. M.
    Nash, D. H.
    Ashton, T. R.
    Stevenson, D. G.
    ADVANCES IN EXPERIMENTAL MECHANICS V, 2007, 7-8 : 81 - +
  • [7] Biomechanical assessment of different transforaminal lumbar interbody fusion constructs in normal and osteoporotic condition: a finite element analysis
    Liu, Chen
    Zhao, Mengmeng
    Zhang, Wencan
    Wang, Chongyi
    Hu, Bingtao
    Wang, Kaibin
    Xu, Wanlong
    Li, Le
    Si, Haipeng
    SPINE JOURNAL, 2024, 24 (06): : 1121 - 1131
  • [8] Biomechanical analysis of bioabsorbable suture anchors for rotator cuff repair using osteoporotic and normal bone models
    Yamauchi, Shohei
    Tsukada, Harehiko
    Sasaki, Eiji
    Sasaki, Shizuka
    Kimura, Yuka
    Yamamoto, Yuji
    Tsuda, Eiichi
    Ishibashi, Yasuyuki
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2022, 27 (01) : 115 - 121
  • [9] Finite element evaluation on biomechanical compatibility of osteoporotic vertebral augmentation with cancellous bone granules and bone cement
    Wu, Li-Jun
    Yang, Guo-Jing
    Zhan, Li-Cheng
    Yu, Bin-Feng
    Yiyong Shengwu Lixue/Journal of Medical Biomechanics, 2010, 25 (02): : 79 - 88
  • [10] Biomechanical tests of osteoporotic vertebral trabecular bone using finite element analysis based on micro-CT
    Woo, Dae Gon
    Lee, Tae Woo
    Ko, Chang Yong
    Kim, Han Sung
    Won, Ye-Yeon
    2006 INTERNATIONAL CONFERENCE ON BIOMEDICAL AND PHARMACEUTICAL ENGINEERING, VOLS 1 AND 2, 2006, : 231 - +