Introduction of ROSA robotic-arm system for total knee arthroplasty is associated with a minimal learning curve for operative time

被引:26
|
作者
Bolam, Scott M. [1 ,2 ]
Tay, Mei Lin [2 ,3 ]
Zaidi, Faseeh [2 ,4 ]
Sidaginamale, Raghavendra P. [5 ]
Hanlon, Michael [1 ]
Munro, Jacob T. [1 ,2 ]
Monk, A. Paul [1 ,2 ,4 ]
机构
[1] Auckland City Hosp, Dept Orthopaed Surg, Auckland, New Zealand
[2] Univ Auckland, Fac Med & Hlth Sci, Dept Surg, Bldg 502 20185 Pk Rd, Auckland 1023, New Zealand
[3] North Shore Hosp, Dept Orthopaed Surg, Auckland, New Zealand
[4] Univ Auckland, Auckland Bioengn Inst, Auckland, New Zealand
[5] Univ Hosp North Tees, Dept Orthopaed, Stockton On Tees, England
关键词
Learning curve; Operative time; Robotically assisted system; Personalised robotic TKA; ROSA knee system; REVISION;
D O I
10.1186/s40634-022-00524-5
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose The introduction of robotics for total knee arthroplasty (TKA) into the operating theatre is often associated with a learning curve and is potentially associated with additional complications. The purpose of this study was to determine the learning curve of robotic-assisted (RA) TKA within a multi-surgeon team. Methods This prospective cohort study included 83 consecutive conventional jig-based TKAs compared with 53 RA TKAs using the Robotic Surgical Assistant (ROSA) system (Zimmer Biomet, Warsaw, Indiana, USA) for knee osteoarthritis performed by three high-volume (> 100 TKA per year) orthopaedic surgeons. Baseline characteristics including age, BMI, sex and pre-operative Kellgren-Lawrence graded and Hip-Knee-Ankle Axis were well-matched between the conventional and RA TKA groups. Cumulative summation (CUSUM) analysis was used to assess learning curves for operative times for each surgeon. Peri-operative and delayed complications (infection, periprosthetic fracture, thromboembolism, and compromised wound healing) and revisions were reviewed. Results The CUSUM analysis for operative time demonstrated an inflexion point after 5, 6 and 15 cases for each of the three surgeons, or 8.7 cases on average. There were no significant differences (p = 0.53) in operative times between the RA TKA learning (before inflexion point) and proficiency (after inflexion point) phases. Similarly, the operative times of the RA TKA group did not differ significantly (p = 0.92) from the conventional TKA group. There was no discernible learning curve for the accuracy of component planning using the RA TKA system. The average length of post-operative follow-up was 21.3 +/- 9.0 months. There was one revision for instability in the conventional TKA group and none in the RA TKA group. There were no significant difference (p > 0.99) in post-operative complication rates between the conventional TKA and RA TKA groups. Conclusions The introduction of the RA TKA system was associated with a learning curve for operative time of 8.7 cases. Operative times between the RA TKA and conventional TKA group were similar. The short learning curve implies this RA TKA system can be adopted relatively quickly into a surgical team with minimal risks to patients.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] Learning curve for active robotic total knee arthroplasty
    Mahure, Siddharth A.
    Teo, Greg Michael
    Kissin, Yair D.
    Stulberg, Bernard N.
    Kreuzer, Stefan
    Long, William J.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2022, 30 (08) : 2666 - 2676
  • [22] Robotic-Arm Assisted Total Knee Arthroplasty: Cost Savings Demonstrated at One Year
    Ong, Kevin L.
    Coppolecchia, Andrea
    Chen, Zhongming
    Watson, Heather N.
    Jacofsky, David
    Mont, Michael A.
    CLINICOECONOMICS AND OUTCOMES RESEARCH, 2022, 14 : 309 - 318
  • [23] Accuracy in the Execution of Pre-operative Plan for Limb Alignment and Implant Positioning in Robotic-arm Assisted Total Knee Arthroplasty and Manual Total Knee Arthroplasty: A Prospective Observational Study
    Jai Thilak
    Balu C. Babu
    Mohan Thadi
    Vipin Mohan
    T. Arun Kumar
    Prajwal P. Mane
    Greeshma C. Ravindran
    Indian Journal of Orthopaedics, 2021, 55 : 953 - 960
  • [24] Accuracy in the Execution of Pre-operative Plan for Limb Alignment and Implant Positioning in Robotic-arm Assisted Total Knee Arthroplasty and Manual Total Knee Arthroplasty: A Prospective Observational Study
    Thilak, Jai
    Babu, Balu C.
    Thadi, Mohan
    Mohan, Vipin
    Kumar, T. Arun
    Mane, Prajwal P.
    Ravindran, Greeshma C.
    INDIAN JOURNAL OF ORTHOPAEDICS, 2021, 55 (04) : 953 - 960
  • [25] Semiactive robotic-arm system versus patient-specific instrumentation in primary total knee arthroplasty: Efficacy and accuracy
    He, Rui
    Sun, Maolin
    Xiong, Ran
    Yang, Junjun
    Guo, Lin
    Yang, Liu
    ASIAN JOURNAL OF SURGERY, 2023, 46 (02) : 742 - 750
  • [26] The learning curve to ROSA: cases needed to match the surgery time between a robotic-assisted and a manual primary total knee arthroplasty
    Eustathios Kenanidis
    Panagiotis Boutos
    Olga Sitsiani
    Eleftherios Tsiridis
    European Journal of Orthopaedic Surgery & Traumatology, 2023, 33 : 3357 - 3363
  • [27] The learning curve to ROSA: cases needed to match the surgery time between a robotic-assisted and a manual primary total knee arthroplasty
    Kenanidis, Eustathios
    Boutos, Panagiotis
    Sitsiani, Olga
    Tsiridis, Eleftherios
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2023, 33 (08): : 3357 - 3363
  • [28] The Learning Curve of Reaching the Planned Limb Alignment in Robotic-Arm-Assisted Total Knee Arthroplasty
    Pannu, Tejbir S.
    Villa, Jesus M.
    Schultz, Isaac
    Barsoum, Wael K.
    Higuera, Carlos A.
    Patel, Preetesh D.
    JOURNAL OF KNEE SURGERY, 2023, 36 (06) : 682 - 688
  • [29] MAKO Robotic-Arm Assisted Total Knee Arthroplasty: Surgical Technique From the Office to the Operating Room
    Yohe, Nicholas
    Chen, Zhongming
    Mont, Michael A.
    Sultan, Assem A.
    SURGICAL TECHNOLOGY INTERNATIONAL-INTERNATIONAL DEVELOPMENTS IN SURGERY AND SURGICAL RESEARCH, 2021, 39
  • [30] Robotic-arm assisted total knee arthroplasty is associated with improved early functional recovery and reduced time to hospital discharge compared with conventional jig-based total knee arthroplasty A PROSPECTIVE COHORT STUDY
    Kayani, B.
    Konan, S.
    Tahmassebi, J.
    Pietrzak, J. R. T.
    Haddad, F. S.