Basic imaging properties of a new screen-film system for chest radiography

被引:3
|
作者
Higashida, Y
Murakami, Y
Yoshida, A
Kastuda, N
Moribe, N
Bussaka, H
Hamada, T
Yoshida, Y
Hidetoshi, M
Takahashi, M
机构
[1] Dept. of Radiological Technology, Kumamoto University, College of Medical Science, Kumamoto 862
[2] Department of Radiology, O-ita Medical University, O-ita 879-56
[3] Dept. Radiological Sci. and Technol., Hiroshima Pref. Coll. Hlth. Welf., Mihara Hiroshima 723
[4] Department of Radiology, Kumamoto University Hospital, Kumamoto 860
关键词
screen-film system; chest radiography; resolution property; noise Wiener spectrum; radiological visibility;
D O I
10.1118/1.597709
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To evaluate the potential clinical usefulness of a new screen-film system (advanced screen-film system; AD system) for chest radiography, its fundamental imaging properties compared with a conventional screen-film system (HR-4/HR-S) were investigated. The basic imaging properties were evaluated by measuring characteristic (H&D) curves, relative speeds, MTFs (modulation transfer functions), WS (Wiener spectra), and x-ray attenuations of screens. The detail visibilities and pathological details of various diseases in chest radiographs of patients were evaluated subjectively. The film gradient of the AD system was slightly lower at low radiographic density, and higher at high density, as compared with a conventional screen-film system. The screen speed of the AD system was 212% greater than that of the conventional system and the film speed was 53% that of the conventional film. As the result the total speed of the AD system was slightly higher compared with the conventional system. The spatial resolution of the AD system was comparable to or slightly lower than that of the conventional system. The noise level of the AD system was considerably lower than that of the conventional system at low (D=0.5) and middle (D=1.0) radiographic density levels. However, it was high at high radiographic density (D=1.8). The radiographic densities in the underpenetrated areas with the AD system were greater than those of the conventional system when the lung densities are matched comparable. Improvement in noise level with the AD system at low and middle density levels may be useful for detection of various diseases in chest radiographs. (C) 1996 American Association of Physicists in Medicine.
引用
收藏
页码:1351 / 1357
页数:7
相关论文
共 50 条
  • [31] Experiences with a new film-screen system in pediatric chest radiography.
    Leenen, A
    Brandt, GA
    Riebel, T
    Marciniak, H
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1996, 165 (04): : 349 - 352
  • [32] COMPARISON OF IMAGING PROPERTIES OF A COMPUTED RADIOGRAPHY SYSTEM AND SCREEN FILM SYSTEMS
    SANADA, S
    DOI, K
    XU, XW
    YIN, FF
    GIGER, ML
    MACMAHON, H
    MEDICAL PHYSICS, 1991, 18 (03) : 414 - 420
  • [33] AN EVALUATION OF SCREEN-FILM COMBINATIONS FOR DETAIL SKELETAL RADIOGRAPHY
    DESMET, AA
    RITTER, EM
    FRITZ, SL
    MARTIN, NL
    CHANG, CHJ
    TEMPLETON, AW
    RADIOLOGY, 1982, 143 (01) : 259 - 260
  • [35] DETERMINATION OF THE SPEED OF MEDICAL RADIOGRAPHY SCREEN-FILM SYSTEMS
    VERDUN, FR
    BOCHUD, FO
    VALLEY, JF
    THANH, OL
    RADIATION PROTECTION DOSIMETRY, 1995, 57 (1-4) : 257 - 261
  • [36] REDUCTION OF ABSORBED DOSE IN RADIOGRAPHY OF BREAST - EXPERIENCE WITH A NEW SCREEN-FILM COMBINATION
    ANDERSSON, I
    ANDREN, L
    NILSSON, M
    PETTERSSON, C
    ACTA RADIOLOGICA-DIAGNOSIS, 1977, 18 (02): : 264 - 268
  • [37] Chest radiography: Depiction of normal anatomy and pathologic structures with selenium-based digital radiography versus conventional screen-film radiography
    Woodard, PK
    Slone, RM
    Gierada, DS
    Reiker, GG
    Pilgram, TK
    Jost, RG
    RADIOLOGY, 1997, 203 (01) : 197 - 201
  • [38] LARGE-SCREEN IMAGE INTENSIFIER PHOTOFLUOROGRAPHY COMPARED WITH FULL-SIZE SCREEN-FILM TECHNIQUE IN CHEST RADIOGRAPHY
    MANNINEN, H
    RYTKONEN, H
    SOIMAKALLIO, S
    TERHO, EO
    HENTUNEN, J
    ACTA RADIOLOGICA-DIAGNOSIS, 1985, 26 (05): : 525 - 533
  • [39] Technical aspects of screen-film radiography, film processing, and quality control
    Gray, JE
    RADIOGRAPHICS, 1997, 17 (01) : 177 - 187
  • [40] DEVELOPMENT OF NEW IMAGING-SYSTEM FOR CHEST RADIOGRAPHY - INVESTIGATION OF BASIC IMAGING PROPERTIES AND IMPROVEMENT OF SIGNAL-DETECTION
    HIGASHIDA, Y
    KARIYA, K
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1991, 30 (9A): : 2139 - 2143