3D ultrasound navigation system for screw insertion in posterior spine surgery: a phantom study

被引:6
|
作者
Chan, Andrew [1 ]
Parent, Eric [2 ]
Mahood, Jim [3 ]
Lou, Edmond [1 ,3 ,4 ]
机构
[1] Univ Alberta, Dept Biomed Engn, 1098 Res Transit Facil, 8308-114 St, Edmonton, AB T6G 2V2, Canada
[2] Univ Alberta, Fac Rehabil Med, Dept Phys Therapy, 2-50 Corbett Hall, Edmonton, AB T6G 2G4, Canada
[3] Univ Alberta, 2D Walter C Mackenzie Hlth Sci Ctr, Dept Surg, 8440-112 St, Edmonton, AB T6G 2B7, Canada
[4] Univ Alberta, Dept Elect Engn, Donadeo ICE 11 263, 9211-116 St, Edmonton, AB T6G 1H9, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
3D ultrasound (3DUS); Image guidance; Motion tracking; Scoliosis; Image registration; ADOLESCENT IDIOPATHIC SCOLIOSIS; ACCURACY; REGISTRATION; CT; PLACEMENT; VERTEBRAE; FUSION;
D O I
10.1007/s11548-021-02516-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose Posterior spinal fusion surgery is required to correct severe idiopathic scoliosis. The surgery involves insertion of screws which requires high accuracy to prevent neurologic damage to the spinal cord. Although conventional CT navigation can reduce this risk, 3D-ultrasound-based navigation could achieve this without added ionizing radiation and usage of expensive and bulky equipment. This study aimed to evaluate the accuracy of a 3D ultrasound navigation system for posterior spine surgery. Methods A custom 3D ultrasound (3DUS) with model-to-surface registration algorithm was developed and integrated into a 3D navigation environment. A CT scan of an adolescent spine (T3-T11) was segmented and 3D printed for experiments. A probe with reflective markers was placed in vertebral pedicles 684 times in varying levels, positions in the capture space and orientation of vertebra, and the entrypoint and trajectory accuracies were measured. Results Among 684 probe placements in vertebral levels T3 to T11 in the phantom spine, 95.5% were within 1 mm and 5 degrees of accuracy, with an average accuracy of 0.4 +/- 0.4 mm and 2.1 +/- 0.9 degrees, requiring 8.8 s to process. Accuracies were statistically significantly affected by vertebral orientation and position in the capture volume, though this was still within the targeted accuracies of 1 mm and 5 degrees. Conclusion This preliminary ultrasound-based navigation system is accurate and fast enough for guiding placement of pedicle screws into the spine in posterior fusion surgery. The current results are limited to phantom spines, and future study in animal or human cadavers is needed to investigate soft tissue effects on registration accuracy.
引用
收藏
页码:271 / 281
页数:11
相关论文
共 50 条
  • [1] 3D ultrasound navigation system for screw insertion in posterior spine surgery: a phantom study
    Andrew Chan
    Eric Parent
    Jim Mahood
    Edmond Lou
    International Journal of Computer Assisted Radiology and Surgery, 2022, 17 : 271 - 281
  • [2] Use of 3D Navigation in Subaxial Cervical Spine Lateral Mass Screw Insertion
    Arab, Abdullah
    Alkherayf, Fahad
    Sachs, Adam
    Wai, Eugene K.
    JOURNAL OF NEUROLOGICAL SURGERY REPORTS, 2018, 79 (01) : e1 - e8
  • [3] 3D ultrasound navigation in syrinx surgery - a feasibility study
    Bonsanto, MM
    Metzner, R
    Aschoff, A
    Tronnier, V
    Kunze, S
    Wirtz, CR
    ACTA NEUROCHIRURGICA, 2005, 147 (05) : 533 - 541
  • [4] 3D ultrasound navigation in syrinx surgery – a feasibility study
    M. M. Bonsanto
    R. Metzner
    A. Aschoff
    V. Tronnier
    S. Kunze
    C. R. Wirtz
    Acta Neurochirurgica, 2005, 147 : 533 - 541
  • [5] 3D ultrasound navigation in syrinx surgery - a feasibility study - Comment
    Unsgaard, G
    ACTA NEUROCHIRURGICA, 2005, 147 (05) : 540 - 541
  • [6] Pitfalls of 2D and 3D navigation in spine surgery
    Arand, M.
    Kinzl, L.
    Gebhard, F.
    TRAUMA UND BERUFSKRANKHEIT, 2005, 7 : S311 - S316
  • [7] Positional and Orientational Accuracy of 3-D Ultrasound Navigation System on Vertebral Phantom Study
    Ottacher, Drake
    Chan, Andrew
    Parent, Eric
    Lou, Edmond
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2020, 69 (09) : 6412 - 6419
  • [8] A system for ultrasound based intraoperative navigation in spine surgery
    Weber, PK
    Schlegel, JC
    Meiche, J
    Peter, L
    Harland, U
    2001 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2001, : 1361 - 1364
  • [9] Factors influencing the accuracy of iliosacral screw insertion using 3D fluoroscopic navigation
    Takao, Masaki
    Hamada, Hidetoshi
    Sakai, Takashi
    Sugano, Nobuhiko
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2019, 139 (02) : 189 - 195
  • [10] Factors influencing the accuracy of iliosacral screw insertion using 3D fluoroscopic navigation
    Masaki Takao
    Hidetoshi Hamada
    Takashi Sakai
    Nobuhiko Sugano
    Archives of Orthopaedic and Trauma Surgery, 2019, 139 : 189 - 195