A novel anthropomorphic multimodality phantom for MRI-based radiotherapy quality assurance testing

被引:18
|
作者
Singhrao, Kamal [1 ]
Fu, Jie [1 ]
Wu, Holden H. [2 ]
Hu, Peng [2 ]
Kishan, Amar U. [1 ]
Chin, Robert K. [1 ]
Lewis, John H. [3 ]
机构
[1] Univ Calif Los Angeles, Dept Radiat Oncol, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, Dept Radiol, Los Angeles, CA 90095 USA
[3] Cedars Sinai Med Ctr, Dept Radiat Oncol, Los Angeles, CA 90048 USA
关键词
anthropomorphic; MRI; multimodality; phantom; QA; radiotherapy; RADIATION-THERAPY; RELAXATION-TIMES; PROSTATE RADIOTHERAPY; IN-VIVO; CT; TISSUE; SYSTEM; SCANS; 1.5T;
D O I
10.1002/mp.14027
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Increased utilization of magnetic resonance imaging (MRI) in radiotherapy has caused a growing need for phantoms that provide tissue-like contrast in both computed tomography (CT) and MRI images. Such phantoms can be used to compare MRI-based processes with CT-based clinical standards. Here, we develop and demonstrate the clinical utility of a three-dimensional (3D)-printed anthropomorphic pelvis phantom containing materials capable of T-1, T-2, and electron density matching for a clinically relevant set of soft tissues and bone. Methods The phantom design was based on a male pelvic anatomy template with thin boundaries separating tissue types. Slots were included to allow insertion of various dosimeters. The phantom structure was created using a 3D printer. The tissue compartments were filled with carrageenan-based materials designed to match the T-1 and T-2 relaxation times and electron densities of the corresponding tissues. CT and MRI images of the phantom were acquired and used to compare phantom T-1 and T-2 relaxation times and electron densities to literature-reported values for human tissue. To demonstrate clinical utility, the phantom was used for end-to-end testing of an MRI-only treatment simulation and planning workflow. Based on a T-2-weighted MRI image, synthetic CT (sCT) images were created using a statistical decomposition algorithm (MRIPlanner, Spectronic Research AB, Sweden) and used for dose calculation of volumetric-modulated arc therapy (VMAT) and seven-field intensity-modulated radiation therapy (IMRT) prostate plans. The plans were delivered on a Truebeam STX (Varian Medical Systems, Palo Alto, CA), with film and a 0.3 cc ion chamber used to measure the delivered dose. Doses calculated on the CT and sCTs were compared using common dose volume histogram metrics. Results T-1 and T-2 relaxation time and electron density measurements for the muscle, prostate, and bone agreed well with literature-reported in vivo measurements. Film analysis resulted in a 99.7% gamma pass rate (3.0%, 3.0 mm) for both plans. The ion chamber-measured dose discrepancies at the isocenter were 0.36% and 1.67% for the IMRT and VMAT plans, respectively. The differences in PTV D98% and D95% between plans calculated on the CT and 1.5T/3.0 T-derived sCT images were under 3%. Conclusion The developed phantom provides tissue-like contrast on MRI and CT and can be used to validate MRI-based processes through comparison with standard CT-based processes.
引用
收藏
页码:1443 / 1451
页数:9
相关论文
共 50 条
  • [11] MRI-based radiotherapy planning
    Largent, A.
    Nunes, J. -C.
    Lafond, C.
    Perichon, N.
    Castelli, J.
    Rolland, Y.
    Acosta, O.
    de Crevoisier, R.
    CANCER RADIOTHERAPIE, 2017, 21 (08): : 788 - 798
  • [12] Phantom for Quality assurance of the Isocenter in High precision radiotherapy
    Hofmann, R.
    Weber, U.
    Jentzsch, M.
    Wittig, A.
    Engenhart-Cabillic, R.
    Zink, K.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2012, 188 : 83 - 83
  • [13] A Novel Integrated Quality Assurance Phantom for Radiographic and Nonradiographic Radiotherapy Localization and Positioning Systems
    Fowler, T.
    Yu, Amy S.
    MEDICAL PHYSICS, 2017, 44 (06) : 3060 - 3060
  • [14] Commissioning an Anthropomorphic Spine and Lung Phantom for Remote Quality Assurance of Spinal Radiosurgery
    Caruthers, D.
    Ibbott, G.
    Shiu, A.
    Chang, E.
    White, R.
    Followill, D.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [15] A novel phantom design for brachytherapy quality assurance
    Gholami, S.
    Mirzaei, H. R.
    Artaee, A. Jabbary
    Jaberi, R.
    Nedaie, H. A.
    Mandavi, S. Rabi
    Bolookat, E. Rajah
    Meigooni, A. S.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2016, 14 (01): : 67 - 71
  • [16] Development and Testing of a Quality Assurance Phantom for Microwave Imaging
    Alsbou, N.
    Ahmad, S.
    Ali, I.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [17] A deformable phantom for quality assurance in 4D radiotherapy
    Margeanu, M.
    Heath, E.
    Stroian, G.
    Seuntjens, J.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S80 - S81
  • [18] A programmable motion phantom for quality assurance of motion management in radiotherapy
    L. Dunn
    T. Kron
    P. N. Johnston
    L. N. McDermott
    M. L. Taylor
    J. Callahan
    R. D. Franich
    Australasian Physical & Engineering Sciences in Medicine, 2012, 35 : 93 - 100
  • [19] A programmable motion phantom for quality assurance of motion management in radiotherapy
    Dunn, L.
    Kron, T.
    Johnston, P. N.
    McDermott, L. N.
    Taylor, M. L.
    Callahan, J.
    Franich, R. D.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (01) : 93 - 100
  • [20] A novel-integrated quality assurance phantom for radiographic and nonradiographic radiotherapy localization and positioning systems
    Yu, Amy S.
    Fowler, Tyler L.
    Dubrowski, Piotr
    MEDICAL PHYSICS, 2018, 45 (07) : 2857 - 2863