A novel anthropomorphic multimodality phantom for MRI-based radiotherapy quality assurance testing

被引:18
|
作者
Singhrao, Kamal [1 ]
Fu, Jie [1 ]
Wu, Holden H. [2 ]
Hu, Peng [2 ]
Kishan, Amar U. [1 ]
Chin, Robert K. [1 ]
Lewis, John H. [3 ]
机构
[1] Univ Calif Los Angeles, Dept Radiat Oncol, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, Dept Radiol, Los Angeles, CA 90095 USA
[3] Cedars Sinai Med Ctr, Dept Radiat Oncol, Los Angeles, CA 90048 USA
关键词
anthropomorphic; MRI; multimodality; phantom; QA; radiotherapy; RADIATION-THERAPY; RELAXATION-TIMES; PROSTATE RADIOTHERAPY; IN-VIVO; CT; TISSUE; SYSTEM; SCANS; 1.5T;
D O I
10.1002/mp.14027
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Increased utilization of magnetic resonance imaging (MRI) in radiotherapy has caused a growing need for phantoms that provide tissue-like contrast in both computed tomography (CT) and MRI images. Such phantoms can be used to compare MRI-based processes with CT-based clinical standards. Here, we develop and demonstrate the clinical utility of a three-dimensional (3D)-printed anthropomorphic pelvis phantom containing materials capable of T-1, T-2, and electron density matching for a clinically relevant set of soft tissues and bone. Methods The phantom design was based on a male pelvic anatomy template with thin boundaries separating tissue types. Slots were included to allow insertion of various dosimeters. The phantom structure was created using a 3D printer. The tissue compartments were filled with carrageenan-based materials designed to match the T-1 and T-2 relaxation times and electron densities of the corresponding tissues. CT and MRI images of the phantom were acquired and used to compare phantom T-1 and T-2 relaxation times and electron densities to literature-reported values for human tissue. To demonstrate clinical utility, the phantom was used for end-to-end testing of an MRI-only treatment simulation and planning workflow. Based on a T-2-weighted MRI image, synthetic CT (sCT) images were created using a statistical decomposition algorithm (MRIPlanner, Spectronic Research AB, Sweden) and used for dose calculation of volumetric-modulated arc therapy (VMAT) and seven-field intensity-modulated radiation therapy (IMRT) prostate plans. The plans were delivered on a Truebeam STX (Varian Medical Systems, Palo Alto, CA), with film and a 0.3 cc ion chamber used to measure the delivered dose. Doses calculated on the CT and sCTs were compared using common dose volume histogram metrics. Results T-1 and T-2 relaxation time and electron density measurements for the muscle, prostate, and bone agreed well with literature-reported in vivo measurements. Film analysis resulted in a 99.7% gamma pass rate (3.0%, 3.0 mm) for both plans. The ion chamber-measured dose discrepancies at the isocenter were 0.36% and 1.67% for the IMRT and VMAT plans, respectively. The differences in PTV D98% and D95% between plans calculated on the CT and 1.5T/3.0 T-derived sCT images were under 3%. Conclusion The developed phantom provides tissue-like contrast on MRI and CT and can be used to validate MRI-based processes through comparison with standard CT-based processes.
引用
收藏
页码:1443 / 1451
页数:9
相关论文
共 50 条
  • [1] A Novel Multimodal Anthropomorphic Physical Pelvis Phantom for MRI-Based Radiotherapy
    Singhrao, K.
    Fu, J.
    Tenn, S.
    Wu, H.
    Hu, P.
    Cao, M.
    Kishan, A.
    Chin, R.
    Lewis, J.
    MEDICAL PHYSICS, 2019, 46 (06) : E256 - E256
  • [2] Multimodality image correlation quality assurance anthropomorphic phantom
    Mutic, S
    Bosch, WR
    Drzymala, RE
    Chao, KSC
    Low, DA
    Lewis, JS
    Cutler, PD
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 1172 - 1175
  • [3] Dynamic anthropomorphic thorax phantom for quality assurance of motion management in radiotherapy
    Abdollahi, Sara
    Mowlavi, Ali Asghar
    Yazdi, Mohammad Hadi Hadizadeh
    Ceberg, Sofie
    Aznar, Marianne Camille
    Tabrizi, Fatemeh Varshoee
    Salek, Roham
    Guckenberger, Matthias
    Tanadini-Lang, Stephanie
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2024, 30
  • [4] An anthropomorphic multimodality (CT/MRI) head phantom prototype for end-to-end tests in ion radiotherapy
    Gallas, Raya R.
    Huenemohr, Nora
    Runz, Armin
    Niebuhr, Nina I.
    Jaekel, Oliver
    Greilich, Steffen
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2015, 25 (04): : 391 - 399
  • [5] An anthropomorphic phantom for quality assurance and training in gynaecological brachytherapy
    de Almeida, CE
    Rodriguez, M
    Vianello, E
    Ferreira, IH
    Sibata, C
    RADIOTHERAPY AND ONCOLOGY, 2002, 63 (01) : 75 - 81
  • [6] MRI-Based Radiotherapy
    Mcdonald, F.
    JOURNAL OF THORACIC ONCOLOGY, 2019, 14 (10) : S53 - S53
  • [7] Helical Tomotherapy Patient Specific Quality Assurance in An Anthropomorphic Phantom
    Neeley, M.
    Harris, C.
    Charara, Y.
    Ramsey, C.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [8] Development of an anthropomorphic multimodality pelvis phantom for PET/MRI- and CT-based RT planning
    Homolka, N.
    Pfaffenberger, A.
    Beuthien-Baumann, B.
    Mann, P.
    Schneider, V.
    Johnen, W.
    Runz, A.
    Echner, G.
    Hoffmann, A. L.
    Troost, E.
    Koerber, S. A.
    Seco, J.
    Gillmann, C.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S246 - S246
  • [9] The Radiological Physics Center's Anthropomorphic Quality Assurance Phantom Program
    Amador, C.
    Hernandez, N.
    Molineu, A.
    Alvarez, P.
    Followill, D.
    MEDICAL PHYSICS, 2012, 39 (06) : 3744 - 3744
  • [10] Design of an anthropomorphic intensity modulated radiation therapy quality assurance phantom
    Radford, DA
    Followill, DS
    Hanson, WF
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 197 - 200