Comparison of electrospray LC-MS, LC-MS with Cold EI and GC-MS with Cold EI for sample identification

被引:25
|
作者
Tsizin, Svetlana [1 ]
Bokka, Ramesh [1 ]
Keshet, Uri [1 ]
Alon, Tal [1 ]
Fialkov, Alexander B. [1 ]
Tal, Noam [1 ]
Amirav, Aviv [1 ]
机构
[1] Tel Aviv Univ, Sch Chem, IL-69978 Tel Aviv, Israel
基金
以色列科学基金会;
关键词
ESI-LC-MS; Electron ionization LC-MS; LC-MS with Cold EI; GC-MS with Cold EI; Supersonic molecular beams; SUPERSONIC MOLECULAR-BEAMS; CHROMATOGRAPHY-MASS-SPECTROMETRY; ATMOSPHERIC-PRESSURE PHOTOIONIZATION; ION-SOURCE; IONIZATION; INTERFACE;
D O I
10.1016/j.ijms.2017.09.006
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
Electrospray LC-MS, LC-MS with Cold EI and GC-MS with Cold EI (Electron Ionization (EI) of vibrationally cold molecules in supersonic molecular beams (SMB)) are compared in their sample identification capabilities. Two test mixtures were used: (a) acetaminophen, caffeine, sulfadimethoxine, terfenadine and reserpine; and (b) octafluoronaphthalene (OFN), pyrene, cholesterol and agidol 40. The major strength of EI in both LC-MS with Cold EI and GC-MS with Cold EI is its compatibility with library based sample identification that provides samples names and structures at the isomer level. GC-MS with Cold EI alleviates two majordownsides of GC-MS with standard EI as it provides enhanced molecular ions that are often weak or missing in standard EI and, it significantly extends the range of compounds that are amenable for analysis. Consequently, reserpine and agidol 40 that are incompatible with standard GC-MS analyses are analyzed by GC-MS with Cold EI and yield effective library based identification. In cases involving sample compounds that are not in the library, ESI-LC-MS provides elemental formula information (when combined with high resolution MS) while GC-MS with Cold EI also provides elemental formula when TAMI software is used for the quadrupole mass accuracy improvement combined with isotope abundance analysis. We found that both LC-MS with Cold EI and GC-MS with Cold EI have broader range of small molecules amenable for analysis than ESI-LC-MS that fails to analyze relatively non-polar compounds and exhibits highly non-uniform and compound dependent ionization yields that span over four orders of magnitude. The total ion mass chromatograms S/N of ESI-LC-MS and LC-MS with Cold EI are comparable while it is better (>100 times per on-column amount) for GC-MS with Cold EI. (C) 2017 Elsevier B.V. All rights reserved.
引用
下载
收藏
页码:119 / 125
页数:7
相关论文
共 50 条
  • [31] LC-MS and GC-MS based metabolomics platform for cancer research
    Wenyun Lu
    Sisi Zhang
    Xin Teng
    Eugene Melamud
    Mitchell A Lazar
    Eileen White
    Joshua D Rabinowitz
    Cancer & Metabolism, 2 (Suppl 1)
  • [32] Detection and Confirmation of Ginsenosides in Horse Urine by GC-MS and LC-MS
    Chung, E. W.
    Kwok, W. H.
    Leung, D. K. K.
    Leung, G. N. W.
    Tang, F. P. W.
    Wan, T. S. M.
    Stewart, B. D.
    Watkins, K. L.
    CHROMATOGRAPHIA, 2009, 69 (9-10) : 923 - 932
  • [33] Electronic cigarette solution and vapor analysis with GC-MS and LC-MS
    Lazernik, Michael
    Cole, Preston
    Jones, Martin
    Shamp, Brian
    Dormedy, Derek
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2017, 253
  • [34] Derivatization of steroids in biological samples for GC-MS and LC-MS analyses
    Marcos, Josep
    Pozo, Oscar J.
    BIOANALYSIS, 2015, 7 (19) : 2515 - 2536
  • [35] LC-MS and GC-MS based metabolomics platform for cancer research
    Lu, Wenyun
    Zhang, Sisi
    Teng, Xin
    Melamud, Eugene
    Lazar, Mitchell A.
    White, Eileen
    Rabinowitz, Joshua D.
    CANCER & METABOLISM, 2014, 2
  • [36] LC-MS sample prep
    不详
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2011, 30 (08) : VIII - VIII
  • [37] Determination of 355 Pesticides in Lemon and Lemon Juice by LC-MS/MS and GC-MS/MS
    Aslantas, Sule
    Golge, Ozgur
    Gonzalez-Curbelo, Miguel Angel
    Kabak, Bulent
    FOODS, 2023, 12 (09)
  • [38] Multiresidue pesticide analysis of botanical dietary supplements by GC-MS/MS and LC-MS/MS
    Shi, Feng
    Wong, Jon W.
    Zhang, Kai
    Hayward, Douglas G.
    Begley, Timothy H.
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2010, 240
  • [39] Determination of Nine Fentanyl Drugs in Hair Samples by GC-MS/MS and LC-MS/MS
    Wei, Qi
    Su, Fu Hai
    ACS OMEGA, 2022, 7 (23): : 19176 - 19182
  • [40] Comparison of GC-MS and LC-MS methods for the analysis of antioxidant phenolic acids in herbs
    Kivilompolo, Maarit
    Oburka, Vit
    Hyotylainen, Tuulia
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2007, 388 (04) : 881 - 887