Comparison of two software versions of a commercially available computer-aided detection (CAD) system for detecting breast cancer

被引:9
|
作者
Kim, Seung Ja [1 ]
Moon, Woo Kyung [1 ]
Kim, Soo-Yeon [1 ]
Chang, Jung Min [1 ]
Kim, Sun Mi [2 ]
Cho, Nariya [1 ]
机构
[1] Seoul Natl Univ, Boramae Med Ctr, Seoul Metropolitan Govt, Dept Radiol, Seoul, South Korea
[2] Bundang Seoul Natl Univ Hosp, Seoul, South Korea
关键词
Breast neoplasms; computer-aided diagnosis; digital mammography; comparison study; calcification; FIELD DIGITAL MAMMOGRAMS; SCREEN-FILM MAMMOGRAMS; MASS DETECTION; INITIAL-EXPERIENCE; RECALL RATES; SENSITIVITY; PERFORMANCE; SIZE; INFORMATION; IMPROVEMENT;
D O I
10.3109/02841851003709490
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The performance of the computer-aided detection (CAD) system can be determined by the sensitivity and false-positive marks rate, therefore these factors should be improved by upgrading the software version of the CAD system. Purpose: To compare retrospectively the performances of two software versions of a commercially available CAD system when applied to full-field digital mammograms for the detection of breast cancers in a screening group. Material and Methods: Versions 3.1 and 8.3 of a CAD software system (ImageChecker, R2 Technology) were applied to the full-field digital mammograms of 130 women (age range 36-80, mean age 53 years) with 130 breast cancers detected by screening. Results: The overall sensitivities of the version 3.1 and 8.3 CAD systems were 92.3% (120 of 130) and 96.2% (125 of 130) (P = 0.025), respectively, and sensitivities for masses were 78.3% (36 of 46) and 89.1% (41 of 46) (P = 0.024) and for microcalcifications 100% (84 of 84) and 100% (84 of 84), respectively. Version 8.3 correctly marked five lesions of invasive ductal carcinoma that were missed by version 3.1. Average numbers of false-positive marks per image were 0.38 (0.15 for calcifications, 0.23 for masses) for version 3.1 and 0.46 (0.13 for calcifications, 0.33 for masses) for version 8.3 (P = 0.1420). Conclusion: The newer version 8.3 of the CAD system showed better overall sensitivity for the detection of breast cancer than version 3.1 due to its improved sensitivity for masses when applied to full-field digital mammograms.
引用
收藏
页码:482 / 490
页数:9
相关论文
共 50 条
  • [21] Computer-aided detection and diagnosis of breast cancer
    Collins, Michael J.
    Hoffmeister, Jeffrey
    Worrell, Steven W.
    SEMINARS IN ULTRASOUND CT AND MRI, 2006, 27 (04) : 351 - 355
  • [22] Computer-aided detection and diagnosis of breast cancer
    Vyborny, CJ
    Giger, ML
    Nishikawa, RM
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2000, 38 (04) : 725 - +
  • [23] Computer-aided detection of breast cancer - Respond
    Helvie, MA
    RADIOLOGY, 2004, 233 (02) : 617 - 617
  • [24] Diagnostic performance of ultrasound with computer-aided diagnostic system in detecting breast cancer
    Song, Pengjie
    Zhang, Li
    Bai, Longmei
    Wang, Qing
    Wang, Yanlei
    HELIYON, 2023, 9 (10)
  • [25] Diagnostic performance of a commercially available computer-aided diagnosis system for automatic detection of pulmonary nodules: Comparison with single and double reading
    Wormanns, D
    Beyer, F
    Diederich, S
    Ludwig, K
    Heindel, W
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2004, 176 (07): : 953 - 958
  • [26] Comparison of diagnostic performance of computer-aided diagnosis (CAD) of ultrasonography and scintimammography for breast cancer
    Hwang, K.
    Lee, H.
    Om, K.
    Yoon, M.
    Choe, W.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2006, 33 : S269 - S269
  • [27] Computer-Aided Detection System for Breast Cancer Based on GMM and SVM
    El-Sokary, N.
    Arafa, A. A.
    Asad, A. H.
    Hefny, H. A.
    ARAB JOURNAL OF NUCLEAR SCIENCES AND APPLICATIONS, 2019, 52 (02): : 142 - 150
  • [28] MS: Diagnostic performance of a commercially available computer-aided diagnosis system for automatic detection of pulmonary nodules: comparison with single and double reading
    Debatin, J
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2004, 176 (07): : 930 - 930
  • [29] The efficacy of using computer-aided detection (CAD) for detection of breast cancer in mammography screening: a systematic review
    Henriksen, Emilie L.
    Carlsen, Jonathan F.
    Vejborg, Ilse M. M.
    Nielsen, Michael B.
    Lauridsen, Carsten A.
    ACTA RADIOLOGICA, 2019, 60 (01) : 13 - 18
  • [30] Breast Cancer: Computer-aided Detection with Digital Breast Tomosynthesis
    Morra, Lia
    Sacchetto, Daniela
    Durando, Manuela
    Agliozzo, Silvano
    Carbonaro, Luca Alessandro
    Delsanto, Silvia
    Pesce, Barbara
    Persano, Diego
    Mariscotti, Giovanna
    Marra, Vincenzo
    Fonio, Paolo
    Bert, Alberto