Comparison of two software versions of a commercially available computer-aided detection (CAD) system for detecting breast cancer

被引:9
|
作者
Kim, Seung Ja [1 ]
Moon, Woo Kyung [1 ]
Kim, Soo-Yeon [1 ]
Chang, Jung Min [1 ]
Kim, Sun Mi [2 ]
Cho, Nariya [1 ]
机构
[1] Seoul Natl Univ, Boramae Med Ctr, Seoul Metropolitan Govt, Dept Radiol, Seoul, South Korea
[2] Bundang Seoul Natl Univ Hosp, Seoul, South Korea
关键词
Breast neoplasms; computer-aided diagnosis; digital mammography; comparison study; calcification; FIELD DIGITAL MAMMOGRAMS; SCREEN-FILM MAMMOGRAMS; MASS DETECTION; INITIAL-EXPERIENCE; RECALL RATES; SENSITIVITY; PERFORMANCE; SIZE; INFORMATION; IMPROVEMENT;
D O I
10.3109/02841851003709490
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The performance of the computer-aided detection (CAD) system can be determined by the sensitivity and false-positive marks rate, therefore these factors should be improved by upgrading the software version of the CAD system. Purpose: To compare retrospectively the performances of two software versions of a commercially available CAD system when applied to full-field digital mammograms for the detection of breast cancers in a screening group. Material and Methods: Versions 3.1 and 8.3 of a CAD software system (ImageChecker, R2 Technology) were applied to the full-field digital mammograms of 130 women (age range 36-80, mean age 53 years) with 130 breast cancers detected by screening. Results: The overall sensitivities of the version 3.1 and 8.3 CAD systems were 92.3% (120 of 130) and 96.2% (125 of 130) (P = 0.025), respectively, and sensitivities for masses were 78.3% (36 of 46) and 89.1% (41 of 46) (P = 0.024) and for microcalcifications 100% (84 of 84) and 100% (84 of 84), respectively. Version 8.3 correctly marked five lesions of invasive ductal carcinoma that were missed by version 3.1. Average numbers of false-positive marks per image were 0.38 (0.15 for calcifications, 0.23 for masses) for version 3.1 and 0.46 (0.13 for calcifications, 0.33 for masses) for version 8.3 (P = 0.1420). Conclusion: The newer version 8.3 of the CAD system showed better overall sensitivity for the detection of breast cancer than version 3.1 due to its improved sensitivity for masses when applied to full-field digital mammograms.
引用
收藏
页码:482 / 490
页数:9
相关论文
共 50 条
  • [1] Comparison of three commercially available FDA approved computer-aided detection (CAD) systems
    Nelson, MT
    Lechner, MC
    RADIOLOGY, 2002, 225 : 600 - 600
  • [2] Reproducibility of a commercially available computer aided detection (CAD) System
    Malich, A
    Azhari, T
    Freesmeyer, MG
    Fleck, M
    Kaiser, WA
    RADIOLOGY, 1999, 213P : 324 - 324
  • [3] Earlier detection of breast cancer using a computer-aided detection (CAD) system
    Sallam, MY
    Cantor, AB
    Schilling, KJ
    Woods, KS
    RADIOLOGY, 2002, 222 (02) : 590 - 590
  • [4] Tumour detection rate of a new commercially available computer-aided detection system
    Ansgar Malich
    Christiane Marx
    Mirjam Facius
    Thomas Boehm
    Marlies Fleck
    Werner A. Kaiser
    European Radiology, 2001, 11 : 2454 - 2459
  • [5] Tumour detection rate of a new commercially available computer-aided detection system
    Malich, A
    Marx, C
    Facius, M
    Boehm, T
    Fleck, M
    Kaiser, WA
    EUROPEAN RADIOLOGY, 2001, 11 (12) : 2454 - 2459
  • [6] Reproducibility of prompts in computer-aided detection (CAD) of breast cancer
    Taylor, CG
    Champness, J
    Reddy, M
    Taylor, P
    Potts, HWW
    Given-Wilson, R
    CLINICAL RADIOLOGY, 2003, 58 (09) : 733 - 738
  • [7] Computer-aided auscultation of murmurs in children: evaluation of commercially available software
    Lee, Cecilia
    Rankin, Kathryn N.
    Zuo, Kevin J.
    Mackie, Andrew S.
    CARDIOLOGY IN THE YOUNG, 2016, 26 (07) : 1359 - 1364
  • [8] Computer-aided diagnostic system for breast cancer by detecting microcalcifications
    Lee, CS
    Kim, JK
    Park, HW
    IMAGE DISPLAY - MEDICAL IMAGING 1998, 1998, 3335 : 615 - 626
  • [9] Breast Cancer Screening: Potential Role Of Computer-Aided Detection (CAD)
    Feig, Stephen A.
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2002, 1 (02) : 127 - 131
  • [10] Impact of breast density on the performance of a computer-aided detection (CAD) system
    Brem, RF
    Hoffmeister, J
    Jindal, G
    BREAST CANCER RESEARCH AND TREATMENT, 2003, 82 : S20 - S20