Design and biomechanical testing of pedicle screw for osteoporotic incidents

被引:27
|
作者
Demir, Teyfik [1 ]
Camuscu, Necip [1 ]
Tureyen, Kudret
机构
[1] TOBB Univ Econ & Technol, TR-06560 Ankara, Turkey
关键词
Pedicle screw; pull-out; fusion and fatigue; PULL-OUT STRENGTH; FIXATION STRENGTH; CALCIUM-PHOSPHATE; AUGMENTATION; CEMENT;
D O I
10.1177/0954411911434680
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, geometrical features of pedicle screws have been modified and their performances are compared. Performance analysis has been made in terms of pull-out strength and torsional strength. The parameters investigated are core diameter, holes drilled normal to screw axis, angle between sequential holes and distance between holes. Three different core diameter have been studied, which are 4 mm (normal core diameter), 5 mm (medium core diameter) and 5.5 mm (high core diameter). Distance between sequential holes has been arranged such that there is either one hole per pitch or one hole per two pitches. Angle between sequential holes is either 90 degrees or 120 degrees. According to the test results, the screw, with medium core diameter (5 mm) containing one hole per two pitches with 90 degrees angle between sequential holes, has exhibited the optimum performance considering torsional strength and pull-out strength requirements. Its torsional strength is slightly higher than and, when Grade 40 polyurethane foam was used as bone simulating material, its pull-out strength is as good as, an undrilled normal core diameter screw, which is already being used in surgical operations. The fatigue performance of this best performed screw has also been found satisfactory according to the related standard. Its pull-out strength is also tested on a calf vertebra and a promising result has been obtained.
引用
收藏
页码:256 / 262
页数:7
相关论文
共 50 条
  • [21] The Biomechanical Properties of Pedicle Screw Fixation Combined With Trajectory Bone Cement Augmentation in Osteoporotic Vertebrae
    Fan, Haitao T.
    Zhang, Renjie J.
    Shen, Cailiang L.
    Dong, Fulong L.
    Li, Yong
    Song, Peiwen W.
    Gong, Chen
    Wang, Yijin J.
    CLINICAL SPINE SURGERY, 2016, 29 (02): : 78 - 85
  • [22] Augmentation of Pedicle Screw Stability With Calcium Sulfate Cement in Osteoporotic Sheep Biomechanical and Screw-bone Interfacial Evaluation
    Liu, Da
    Lei, Wei
    Wu, Zi-xiang
    Gao, Ming-xuan
    Wan, Shi-yong
    Fu, Suo-chao
    Shi, Lei
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2011, 24 (04): : 235 - 241
  • [23] BIOMECHANICAL TESTING OF A NEW DESIGN FOR SCHANZ PEDICLE SCREWS
    WILLETT, K
    HEARN, TC
    CUNCINS, AV
    JOURNAL OF ORTHOPAEDIC TRAUMA, 1993, 7 (04) : 375 - 380
  • [24] Biomechanical evaluation and preliminary clinical experience with an expansive pedicle screw design
    Cook, SD
    Salkeld, SL
    Whitecloud, TS
    Barbera, J
    JOURNAL OF SPINAL DISORDERS, 2000, 13 (03): : 230 - 236
  • [25] Biomechanical study of expandable pedicle screw fixation in severe osteoporotic bone comparing with conventional and cement-augmented pedicle screws
    Chen, Yi-Long
    Chen, Wen-Chuan
    Chou, Chi-Wei
    Chen, Jou-Wen
    Chang, Chia-Ming
    Lai, Yu-Shu
    Cheng, Cheng-Kung
    Wang, Shih-Tien
    MEDICAL ENGINEERING & PHYSICS, 2014, 36 (11) : 1416 - 1420
  • [26] Biomechanical comparison of pedicle screw augmented with different volumes of polymethylmethacrylate in osteoporotic and severely osteoporotic cadaveric lumbar vertebrae: an experimental study
    Liu, Da
    Zhang, Bo
    Xie, Qing-yun
    Kang, Xia
    Zhou, Jiang-jun
    Wang, Cai-ru
    Lei, Wei
    Zheng, Wei
    SPINE JOURNAL, 2016, 16 (09): : 1124 - 1132
  • [27] Pedicle screw placement in the lumbar spine: effect of trajectory and screw design on acute biomechanical purchase
    Wray, Steven
    Mimran, Ronnie
    Vadapalli, Sasidhar
    Shetye, Snehal S.
    McGilvray, Kirk C.
    Puttlitz, Christian M.
    JOURNAL OF NEUROSURGERY-SPINE, 2015, 22 (05) : 503 - 510
  • [28] Biomechanical comparison of different techniques in primary spinal surgery in osteoporotic cadaveric lumbar vertebrae: expansive pedicle screw versus polymethylmethacrylate-augmented pedicle screw
    Liu, Da
    Wu, Zi-xiang
    Pan, Xian-ming
    Fu, Suo-chao
    Gao, Ming-xuan
    Shi, Lei
    Lei, Wei
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2011, 131 (09) : 1227 - 1232
  • [29] Biomechanical comparison of different techniques in primary spinal surgery in osteoporotic cadaveric lumbar vertebrae: expansive pedicle screw versus polymethylmethacrylate-augmented pedicle screw
    Da Liu
    Zi-xiang Wu
    Xian-ming Pan
    Suo-chao Fu
    Ming-xuan Gao
    Lei Shi
    Wei Lei
    Archives of Orthopaedic and Trauma Surgery, 2011, 131 : 1227 - 1232
  • [30] Biomechanical Investigation of a Novel Revision Device in an Osteoporotic Model Pullout Strength of Pedicle Screw Anchor Versus Larger Screw Diameter
    Manon, Jacinto
    Hussain, Mir M.
    Harris, Jonathan
    Moldavsky, Mark
    La Marca, Frank
    Bucklen, Brandon S.
    CLINICAL SPINE SURGERY, 2017, 30 (06): : 265 - 271