Monte Carlo evaluation of hypothetical long axial field-of-view PET scanner using GE Discovery MI PET front-end architecture

被引:5
|
作者
Tiwari, Ashok [1 ,2 ]
Merrick, Michael [1 ,3 ]
Graves, Stephen A. [1 ,3 ,4 ]
Sunderland, John [1 ,2 ,4 ]
机构
[1] Univ Iowa Hosp & Clin, Dept Radiol, 200 Hawkins Dr, Iowa City, IA 52242 USA
[2] Univ Iowa, Dept Phys & Astron, Iowa City, IA USA
[3] Univ Iowa, Dept Biomed Engn, Iowa City, IA USA
[4] Univ Iowa Hosp & Clin, Dept Radiat Oncol, Iowa City, IA 52242 USA
关键词
Discovery MI; extended axial field-of-view; GATE Monte Carlo; NEMA NU 2-2018; PET; BODY; SIMULATION; GATE;
D O I
10.1002/mp.15422
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The development of total-body PET scanners is of growing interest in the PET community. Investigation into the imaging properties of a hypothetical extended axial field-of-view (AFOV) GE Healthcare SiPM-based Discovery MI (DMI) system architecture has not yet been performed. In this work, we assessed its potential as a whole-body scanner using Monte Carlo simulations. The aim of this work was to (1) develop and validate a Monte Carlo model of a four-ring scanner and (2) extend its AFOV up to 2 m to evaluate performance gain through NEMA-based evaluation. Methods The DMI four-ring geometry and its pulse digitization scheme were modeled within the GATE Monte Carlo platform using published literature. The GATE scanner model was validated by comparing results against published NEMA performance measurements. Following the validation of the four-ring model, the model was extended to simulate 8-, 20-, 30-, and 40-ring systems. Spatial resolution, sensitivity, NECR, and scatter fraction were characterized with modified NEMA NU-2 2018 standards; however, the image quality measurements were not acquired due to computational limitations. Spatial resolutions were simulated for all scanner ring configurations using point sources to examine the effects of parallax errors. NEMA count rates were estimated using a standard 70 cm scatter phantom and an extended version of scatter phantom of length 200 cm with (1-800) MBq of F-18 for all scanners. Sensitivity was evaluated using NEMA methods with a 70 cm standard and a 200 cm long line source. Results The average FWHM of the radial/tangential/axial spatial resolution reconstructed with filtered back-projection at 1 and 10 cm from the scanner center were 3.94/4.10/4.41 mm and 5.29/4.89/5.90 mm for the four-ring scanner. Sensitivity was determined to be 14.86 cps/kBq at the center of the FOV for the four-ring scanner using a 70 cm line source. Sensitivity enhancement up to 21-fold and 60-fold were observed for 1 and 2 m AFOV scanners compared to four-ring scanner using a 200 cm long line source. Spatial resolution simulations in a 2 m AFOV scanner suggest a maximum degradation of similar to 23.8% in the axial resolution compared to the four-ring scanner. However, the transverse resolution was found to be relatively constant when increasing the axial acceptance angle up to +/- 70 degrees. The peak NECR was 212.92 kcps at 22.70 kBq/ml with a scatter fraction of 38.9% for a four-ring scanner with a 70 cm scatter phantom. Comparison of peak NECR using the 200 cm long scatter phantom relative to the four-ring scanner resulted in a NECR gain of 15 for the 20-ring and 28 for the 40-ring geometry. Spatial resolution, sensitivity, and scatter fraction showed an agreement within similar to 7% compared with published measured values. Conclusions The four-ring DMI scanner simulation was successfully validated against published NEMA measurements. Sensitivity and NECR performance of extended 1 and 2 m AFOV scanners based upon the DMI architecture were subsequently simulated. Increases in sensitivity and count-rate performance are consistent with prior simulation studies utilizing extensions of the Siemens mCT architecture and published NEMA measurements with the uEXPLORER system.
引用
收藏
页码:1139 / 1152
页数:14
相关论文
共 50 条
  • [1] Monte Carlo simulation of 4-ring Discovery MI PET/CT scanner and its extended axial field-of-view to 2 m
    Tiwari, Ashok
    Merrick, Michael
    Graves, Stephen
    Sunderland, John
    JOURNAL OF NUCLEAR MEDICINE, 2021, 62
  • [2] Acceptance Testing of a Long Axial Field-of-View PET/ CT Scanner
    Willson, T.
    Chowdhury, A.
    Davis, N.
    Khan, J.
    Walford, O.
    Natarajan, H.
    McCool, D.
    Holman, B. F.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 : S723 - S723
  • [3] Scanner Design Considerations for Long Axial Field-of-View PET Systems
    Daube-Witherspoon, Margaret E.
    Cherry, Simon R.
    PET CLINICS, 2021, 16 (01) : 25 - 39
  • [4] Physical performance of a long axial field-of-view PET scanner prototype with sparse rings configuration: A Monte Carlo simulation study
    Zein, Sara A.
    Karakatsanis, Nicolas A.
    Issa, Mohammad
    Haj-Ali, Amin A.
    Nehmeh, Sadek A.
    MEDICAL PHYSICS, 2020, 47 (04) : 1949 - 1957
  • [5] Image Quality evaluation of analytic reconstruction algorithms for long axial field-of-view PET scanner
    Zhao, Yong
    Deng, Zilin
    Lv, Yang
    Hu, Debin
    Ding, Yu
    Dong, Yun
    JOURNAL OF NUCLEAR MEDICINE, 2019, 60
  • [6] CT recon optimisation on Long Axial Field-of-View PET/ CT scanner
    Ferreira, B.
    Edwards, J.
    Holman, B.
    Pritchard, A.
    Shirsavar, H.
    Leite, A.
    Nathan, M.
    Partiphun, A.
    Wilson, T.
    Wagner, T.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 : S708 - S709
  • [7] Monte Carlo simulation of the system performance of a long axial field-of-view PET based on monolithic LYSO detectors
    Abi-Akl, Maya
    Dadgar, Meysam
    Toufique, Yassine
    Bouhali, Othmane
    Vandenberghe, Stefaan
    EJNMMI PHYSICS, 2023, 10 (01)
  • [8] Monte Carlo simulation of the system performance of a long axial field-of-view PET based on monolithic LYSO detectors
    Maya Abi-Akl
    Meysam Dadgar
    Yassine Toufique
    Othmane Bouhali
    Stefaan Vandenberghe
    EJNMMI Physics, 10
  • [9] Evaluation of deep learning-based scatter correction on a long-axial field-of-view PET scanner
    Laurent, Baptiste
    Bousse, Alexandre
    Merlin, Thibaut
    Rominger, Axel
    Shi, Kuangyu
    Visvikis, Dimitris
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2025,
  • [10] Development and Evaluation of mini-EXPLORER: A Long Axial Field-of-View PET Scanner for Nonhuman Primate Imaging
    Berg, Eric
    Zhang, Xuezhu
    Bec, Julien
    Judenhofer, Martin S.
    Patel, Brijesh
    Peng, Qiyu
    Kapusta, Maciej
    Schmand, Matthias
    Casey, Michael E.
    Tarantal, Alice F.
    Qi, Jinyi
    Badawi, Ramsey D.
    Cherry, Simon R.
    JOURNAL OF NUCLEAR MEDICINE, 2018, 59 (06) : 993 - 998