Clinical Application Evaluation of Two Fourth-Generation Human Immunodeficiency Virus (HIV) Screening Assays in West China Hospital

被引:7
|
作者
Liu, Yongming [1 ]
Li, Dongdong [1 ]
Wang, Tingting [1 ]
Yan, Kening [1 ]
Zhu, Siyuan [1 ]
Yang, Tingfu [1 ]
Luo, Lan [1 ]
Tao, Chuanmin [1 ]
机构
[1] Sichuan Univ, West China Hosp, Dept Lab Med, Div Clin Microbiol, Chengdu 610041, Sichuan, Peoples R China
基金
中国国家自然科学基金;
关键词
HIV; screening; specificity; sensitivity; evaluation; B SURFACE-ANTIGEN; PERFORMANCE; INFECTION; ANTIBODIES; SEROCONVERSION; IMMUNOASSAYS; DIAGNOSIS; TESTS; NEED; HCV;
D O I
10.1002/jcla.21743
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BackgroundFourth-generation human immunodeficiency virus (HIV) screening assays have been used in many laboratories. The Elecsys (R) HIV combi PT assay is a new kind of fourth-generation HIV screening assay developed to allow earlier detection of seroconversion. MethodsA total of 271,845 routine specimens were detected using the Elecsys (R) HIV combi assay and Elecsys (R) HIV combi PT assay from September 2010 to December 2012 in a large university hospital. Repeatedly, reactive screening samples were confirmed according to recommended confirmatory algorithms. ResultsThe false-positive rate and positive predictive value (PPV) of two assays are 0.08 and 78.35%, respectively, for the Elecsys (R) HIV combi assay and 0.07 and 82.21% for the Elecsys (R) HIV combi PT assay. Ninety-four percent cases with cutoff index ratio <15.0 were false-positive. When we set the specificity as 95.0 and 99.0%, PPV could increase to 98.7, 99.6, 98.8, and 99.7%, and sensitivity reduced to 99.2, 98.4, 98.5, and 96.8% for the Elecsys (R) HIV combi assay and the Elecsys (R) HIV combi PT assay, respectively. ConclusionsThe Elecsys (R) HIV combi PT assay shows a better performance in specificity than the Elecsys (R) HIV combi assay. Most weakly reactive results were false-positive, this means it still need to be improved and it will need laboratory personnel to communicate with the clinical doctor and patients more properly about the result of the assay.
引用
收藏
页码:146 / 152
页数:7
相关论文
共 50 条
  • [31] Assessment of the Ability of a Fourth-Generation Immunoassay for Human Immunodeficiency Virus (HIV) Antibody and p24 Antigen To Detect both Acute and Recent HIV Infections in a High-Risk Setting
    Pandori, Mark W.
    Hackett, John, Jr.
    Louie, Brian
    Vallari, Ana
    Dowling, Teri
    Liska, Sally
    Klausner, Jeffrey D.
    JOURNAL OF CLINICAL MICROBIOLOGY, 2009, 47 (08) : 2639 - 2642
  • [32] Acute Infections, Cost per Infection and Turnaround Time in Three United States Hospital Laboratories Using Fourth-Generation Antigen-Antibody Human Immunodeficiency Virus Immunoassays
    Wesolowski, Laura G.
    Nasrullah, Muazzam
    Coombs, Robert W.
    Rosenberg, Eric
    Ethridge, Steven F.
    Hutchinson, Angela B.
    Dragavon, Joan
    Rychert, Jennifer
    Nolte, Frederick S.
    Madory, James E.
    Werner, Barbara G.
    OPEN FORUM INFECTIOUS DISEASES, 2016, 3 (01):
  • [33] Evaluation of a rapid and simple fourth-generation HIV screening assay for qualitative detection of HIV p24 antigen and/or antibodies to HIV-1 and HIV-2
    Beelaert, G.
    Fransen, K.
    JOURNAL OF VIROLOGICAL METHODS, 2010, 168 (1-2) : 218 - 222
  • [34] Human immunodeficiency virus (HIV) antigen-antibody combination assays: Evaluation of HIV Seroconversion sensitivity and subtype detection
    Weber, B
    JOURNAL OF CLINICAL MICROBIOLOGY, 2002, 40 (11) : 4402 - 4403
  • [35] An evaluation of human immunodeficiency virus oral screening test awareness and preferences in the West region of Cameroon
    Nkenfou, Celine N.
    Kembou, Japhette T.
    Djikeng, Appolinaire
    Domkam, Irenee
    Tchuinkam, Timoleon
    JOURNAL OF INFECTION AND PUBLIC HEALTH, 2015, 8 (03) : 254 - 259
  • [36] Multicenter evaluation of a new automated fourth- generation human immunodeficiency virus screening assay with a sensitive antigen detection module and high specificity
    Weber, B
    Gürtler, L
    Thorstensson, R
    Michl, U
    Mühlbacher, A
    Bürgisser, P
    Villaescusa, R
    Eiras, A
    Gabriel, C
    Stekel, H
    Tanprasert, S
    Oota, S
    Silvestre, MJ
    Marques, C
    Ladeira, M
    Rabenau, H
    Berger, A
    Schmitt, U
    Melchior, W
    JOURNAL OF CLINICAL MICROBIOLOGY, 2002, 40 (06) : 1938 - 1946
  • [37] CLINICAL-EVALUATION OF ABBOTT AND WELLCOME ENZYME LINKED IMMUNOSORBENT ASSAYS FOR DETECTION OF SERUM ANTIBODIES TO HUMAN-IMMUNODEFICIENCY-VIRUS (HIV)
    EVANS, RP
    SHANSON, DC
    MORTIMER, PP
    JOURNAL OF CLINICAL PATHOLOGY, 1987, 40 (05) : 552 - 555
  • [38] EVALUATION OF 3 2ND-GENERATION AND 3 CONFIRMATORY ASSAYS FOR ANTIBODIES TO HUMAN IMMUNODEFICIENCY VIRUS
    LELIE, PN
    REESINK, HW
    HUISMAN, H
    VOX SANGUINIS, 1988, 54 (02) : 84 - 91
  • [39] Cytomegalovirus Viremia and Clinical Outcomes in Kenyan Children Diagnosed With Human Immunodeficiency Virus (HIV) in Hospital
    Wamalwa, Dalton
    Njuguna, Irene
    Maleche-Obimbo, Elizabeth
    Begnel, Emily
    Chebet, Daisy J.
    Onyango, Judith A.
    Cranmer, Lisa Marie
    Huang, Meei-Li
    Richardson, Barbra A.
    Boeckh, Michael
    John-Stewart, Grace
    Slyker, Jennifer
    CLINICAL INFECTIOUS DISEASES, 2022, 74 (07) : 1237 - 1246
  • [40] Seven human immunodeficiency virus (HIV) antigen-antibody combination assays: Evaluation of HIV seroconversion sensitivity and subtype detection
    Ly, TD
    Martin, L
    Daghfal, D
    Sandridge, A
    West, D
    Bristow, R
    Chalouas, L
    Qiu, XX
    Lou, SC
    Hunt, JC
    Schochetman, G
    Devare, SG
    JOURNAL OF CLINICAL MICROBIOLOGY, 2001, 39 (09) : 3122 - 3128