摘要
Extracorporeal blood flow devices such as artificial ventricular assistance devices, heart-lung machines, and etc. are widely used in clinic medicine for many years. However, the thrombus formation is still a major concern for these devices. Hence, thrombus real-time monitoring is very essential in these devices. Currently, there are few studies based on optical, ultrasonic and electrical measurement principles for the real-time thrombus detection in the extracorporeal blood flow circulation. The red blood cells (RBCs) concentration has been measured by using the optical measurement method. However, the optical measurement for thrombus formation is impossible in the case of various biocompatible coating in the channel wall. Additionally, the ultrasonic method has some inherent limitation such as the inability to differentiate air emboli from thrombus. The electrical method seems better alternative to the both methods. Asakura et al. reported thrombus formation detection using permittivity measurement. However, the permittivity based detection has frequency sensitivity. Determining a suitable frequency range for each application is a difficult task as it depends upon the samples and the nature of the progress of the thrombus formation process. Cole-Cole analysis is the method to analyzer properties of blood on whole frequency range by using four parameters of the relaxation frequency, the resistance at zero frequency, the resistance at infinite frequency, and shape factor. In this study, thrombus formation results of the thrombus formation experiment in flowing condition were reported. To compare with the present blood coagulability measurement, Activated Clotting Time (ACT) result was also reported. Additionally, temporal change in coagulation factor fibrinogen quantity was represented to consider its effects on thrombus formation results.
引用
收藏
页码:61 / 64
页数:4
相关论文
共 50 条
  • [1] Noninvasive and Online Detection of Thrombus in Extracorporeal Circulation Based on Electrical Impedance Tomography
    Zhang, Jia-Hui
    Gao, Gui-Feng
    Wu, Yang
    Zhou, Tong
    Liu, Kai
    Yao, Jia-Feng
    PROGRESS IN BIOCHEMISTRY AND BIOPHYSICS, 2023, 50 (06) : 1480 - 1488
  • [2] Continuous Monitoring of Thrombus Formation by Finite Element Methods (FEMs) Coupled With Electrical Impedance Spectroscopy
    Gao, Bihan
    Liu, Kai
    Lin, Tao
    Sun, Bo
    Wu, Hongtao
    Yao, Jiafeng
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2023, 72 : 1 - 8
  • [3] Cole-Cole analysis of thrombus formation in an extracorporeal blood flow circulation using electrical measurement
    Huu, D. Nguyen
    Kikuchi, D.
    Maruyama, O.
    Sapkota, A.
    Takei, M.
    FLOW MEASUREMENT AND INSTRUMENTATION, 2017, 53 : 172 - 179
  • [4] Quantitative detection and evaluation of thrombus formation based on electrical impedance spectroscopy
    Li, Jianping
    Wan, Nen
    Wen, Jianming
    Cheng, Guangming
    He, Lidong
    Cheng, Li
    BIOSENSORS & BIOELECTRONICS, 2019, 141
  • [5] DIET AND THROMBUS FORMATION - QUANTITATIVE STUDIES USING AN EXTRACORPOREAL CIRCULATION IN PIGS
    MUSTARD, JF
    DOWNIE, HG
    ROWSELL, HC
    MURPHY, EA
    JOURNAL OF CLINICAL INVESTIGATION, 1963, 42 (11): : 1783 - +
  • [6] Development of a Portable Electrical Impedance Tomography Device for Online Thrombus Detection in Extracorporeal-Circulation Equipment
    Chen, Huaijin
    Yao, Jiafeng
    Yang, Lu
    Liu, Kai
    Chen, Bai
    Li, Jianping
    Takei, Masahiro
    IEEE SENSORS JOURNAL, 2021, 21 (03) : 3653 - 3659
  • [7] Monitoring of froth stability using electrical impedance spectroscopy
    Hu, Shenggen
    Ofori, Philip
    Firth, Bruce
    INTERNATIONAL JOURNAL OF MINERAL PROCESSING, 2009, 92 (1-2) : 15 - 21
  • [8] Electrical Impedance Spectroscopy on Flowing Blood to predict White Thrombus Formation in Artificial Microchannels
    Affanni, A.
    Specogna, R.
    Trevisan, F.
    2012 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC), 2012, : 1477 - 1480
  • [9] Red blood cells aggregability measurement of coagulating blood in extracorporeal circulation system with multiple-frequency electrical impedance spectroscopy
    Li, Jianping
    Sapkota, Achyut
    Kikuchi, Daisuke
    Sakota, Daisuke
    Maruyama, Osamu
    Takei, Masahiro
    BIOSENSORS & BIOELECTRONICS, 2018, 112 : 79 - 85
  • [10] Real-time monitoring of scale formation in reverse osmosis using electrical impedance spectroscopy
    Hu, Zhixin
    Antony, Alice
    Leslie, Greg
    Le-Clech, Pierre
    JOURNAL OF MEMBRANE SCIENCE, 2014, 453 : 320 - 327