A visualization method for a wide range of rising temperature induced by high-intensity focused ultrasound using a tissue-mimicking phantom

被引:3
|
作者
Takagi, Ryo [1 ]
Yoshinaka, Kiyoshi [1 ]
Washio, Toshikatsu [1 ]
Koseki, Yoshihiko [1 ]
机构
[1] Natl Inst Adv Ind Sci & Technol, Hlth & Med Res Inst, Med Devices Res Grp, Tsukuba, Ibaraki 3058564, Japan
关键词
High-intensity focused ultrasound; noninvasive thermometry; ultrasound phantom; tissue-mimicking phantom; principal component analysis (PCA); pressure measurement; CANCER; ABLATION; OUTCOMES;
D O I
10.1080/02656736.2021.2012603
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose High-intensity focused ultrasound (HIFU) treatment requires prior evaluation of the HIFU transducer output. A method using micro-capsulated thermochromic liquid crystal (MTLC) to evaluate the temperature distribution in the media during HIFU exposure has been previously developed. However, the color-coded temperature range of commercial MTLC is approximately 10 degrees C, which is insufficient for temperature measurement for HIFU exposure. We created two layers of tissue-mimicking phantoms with different color-coded temperature ranges, and a new visualization method was developed by utilizing the axisymmetric pressure distribution of a HIFU focus. Methods A two-layer phantom with two sensitivity ranges was created. The HIFU transducer was set to align the focal point to the boundary between the two layers. Images of the upper and lower layers were flipped along the boundary between the two layers such that they overlapped with each other, assuming the pressure distribution of HIFU to be axisymmetric. Results The experimental and simulation results were compared to evaluate the accuracy of the phantom temperature measurement. The experimental time profile of the temperature and spatial distribution around the HIFU focus matched well with that of the simulation. However, there is room for improvement in the accuracy in the axial direction of HIFU focus. Conclusion Users can apply our proposed method in clinical practice to promptly assess the output of the HIFU transducer before treatment.
引用
收藏
页码:22 / 33
页数:12
相关论文
共 50 条
  • [41] An extended focused assessment with sonography in trauma ultrasound tissue-mimicking phantom for developing automated diagnostic technologies
    Hernandez-Torres, Sofia I.
    Bedolla, Carlos
    Berard, David
    Snider, Eric J.
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2023, 11
  • [42] Characterization of Focal Location During High-Intensity Focused Ultrasound Ablation in a Tissue Phantom Using Remote Thermocouple Arrays
    Devarakonda, Surendra
    Dibaji, Seyed Ahmad Reza
    Hariharan, Prasanna
    Myers, Matthew R.
    Banerjee, Rupak K.
    JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME, 2016, 10 (02):
  • [43] Absolute Ultrasound Perfusion Parameter Quantification of a Tissue-Mimicking Phantom Using Bolus Tracking
    Mezl, Martin
    Jirik, Radovan
    Harabis, Vratislav
    Kolar, Radim
    Standara, Michal
    Nylund, Kim
    Gilja, Odd Helge
    Taxt, Torfinn
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2015, 62 (05) : 983 - 987
  • [44] TOMOGRAPHIC RECONSTRUCTION OF TISSUE PROPERTIES AND TEMPERATURE INCREASE FOR HIGH-INTENSITY FOCUSED ULTRASOUND APPLICATIONS
    Yin, Lu
    Gudur, Madhu Sudhan Reddy
    Hsiao, Yi-Sing
    Kumon, Ronald E.
    Deng, Cheri X.
    Jiang, Huabei
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2013, 39 (10): : 1760 - 1770
  • [45] Quality Assurance for Ultrasound Scanners using a durable tissue-mimicking Phantom and radial MTF
    Kaar, Marcus
    Semturs, Friedrich
    Figl, Michael
    Hoffmann, Rainer
    Hummel, Johann
    MEDICAL IMAGING 2014: ULTRASONIC IMAGING AND TOMOGRAPHY, 2014, 9040
  • [46] High-Intensity Focused Ultrasound: Heating and Destruction of Biological Tissue
    T. A. Andreeva
    A. E. Berkovich
    N. Y. Bykov
    S. V. Kozyrev
    A. Ya. Lukin
    Technical Physics, 2020, 65 : 1455 - 1466
  • [47] Noninvasive surgery of prostate tissue by high-intensity focused ultrasound
    Sanghvi, NT
    Fry, FJ
    Bihrle, R
    Foster, RS
    Phillips, MH
    Syrus, J
    Zaitsev, AV
    Hennige, CW
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1996, 43 (06) : 1099 - 1110
  • [48] High-Intensity Focused Ultrasound Effectively Reduces Adipose Tissue
    Fatemi, Afschin
    SEMINARS IN CUTANEOUS MEDICINE AND SURGERY, 2009, 28 (04) : 257 - 262
  • [49] NONLINEAR EFFECT IN FAT TISSUE FOR HIGH-INTENSITY FOCUSED ULTRASOUND
    Wang, Shou-bei
    Li, Ke
    Zhou, Bo-fan
    Ji, Xiang
    PROCEEDINGS OF 2016 SYMPOSIUM ON PIEZOELECTRICITY, ACOUSTIC WAVES, AND DEVICE APPLICATIONS (SPAWDA), 2016, : 127 - 130
  • [50] INVESTIGATION INTO THE MECHANISMS OF TISSUE ATOMIZATION BY HIGH-INTENSITY FOCUSED ULTRASOUND
    Simon, Julianna C.
    Sapozhnikov, Oleg A.
    Wang, Yak-Nam
    Khokhlova, Vera A.
    Crum, Lawrence A.
    Bailey, Michael R.
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2015, 41 (05): : 1372 - 1385