A visualization method for a wide range of rising temperature induced by high-intensity focused ultrasound using a tissue-mimicking phantom

被引:3
|
作者
Takagi, Ryo [1 ]
Yoshinaka, Kiyoshi [1 ]
Washio, Toshikatsu [1 ]
Koseki, Yoshihiko [1 ]
机构
[1] Natl Inst Adv Ind Sci & Technol, Hlth & Med Res Inst, Med Devices Res Grp, Tsukuba, Ibaraki 3058564, Japan
关键词
High-intensity focused ultrasound; noninvasive thermometry; ultrasound phantom; tissue-mimicking phantom; principal component analysis (PCA); pressure measurement; CANCER; ABLATION; OUTCOMES;
D O I
10.1080/02656736.2021.2012603
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose High-intensity focused ultrasound (HIFU) treatment requires prior evaluation of the HIFU transducer output. A method using micro-capsulated thermochromic liquid crystal (MTLC) to evaluate the temperature distribution in the media during HIFU exposure has been previously developed. However, the color-coded temperature range of commercial MTLC is approximately 10 degrees C, which is insufficient for temperature measurement for HIFU exposure. We created two layers of tissue-mimicking phantoms with different color-coded temperature ranges, and a new visualization method was developed by utilizing the axisymmetric pressure distribution of a HIFU focus. Methods A two-layer phantom with two sensitivity ranges was created. The HIFU transducer was set to align the focal point to the boundary between the two layers. Images of the upper and lower layers were flipped along the boundary between the two layers such that they overlapped with each other, assuming the pressure distribution of HIFU to be axisymmetric. Results The experimental and simulation results were compared to evaluate the accuracy of the phantom temperature measurement. The experimental time profile of the temperature and spatial distribution around the HIFU focus matched well with that of the simulation. However, there is room for improvement in the accuracy in the axial direction of HIFU focus. Conclusion Users can apply our proposed method in clinical practice to promptly assess the output of the HIFU transducer before treatment.
引用
收藏
页码:22 / 33
页数:12
相关论文
共 50 条
  • [1] Prediction and Measurement of Temperature Rise Induced by High Intensity Focused Ultrasound in a Tissue-Mimicking Phantom
    Kang Il Lee
    Journal of the Korean Physical Society, 2018, 72 : 1313 - 1319
  • [2] Prediction and Measurement of Temperature Rise Induced by High Intensity Focused Ultrasound in a Tissue-Mimicking Phantom
    Lee, Kang Il
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2018, 72 (11) : 1313 - 1319
  • [3] Boiling activity induced by high intensity focused ultrasound in a tissue-mimicking phantom
    Kang, Hwi Suk
    Lee, Kang Il
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2020, 59 (02)
  • [4] Coagulation and ablation patterns of high-intensity focused ultrasound on a tissue-mimicking phantom and cadaveric skin
    Hee-Jin Kim
    Han Gu Kim
    Zhenlong Zheng
    Hyoun Jun Park
    Jeung Hyun Yoon
    Wook Oh
    Cheol Woo Lee
    Sung Bin Cho
    Lasers in Medical Science, 2015, 30 : 2251 - 2258
  • [5] Coagulation and ablation patterns of high-intensity focused ultrasound on a tissue-mimicking phantom and cadaveric skin
    Kim, Hee-Jin
    Kim, Han Gu
    Zheng, Zhenlong
    Park, Hyoun Jun
    Yoon, Jeung Hyun
    Oh, Wook
    Lee, Cheol Woo
    Cho, Sung Bin
    LASERS IN MEDICAL SCIENCE, 2015, 30 (09) : 2251 - 2258
  • [6] Development and Characterization of a Tissue-Mimicking Material for High-Intensity Focused Ultrasound
    King, Randy L.
    Liu, Yunbo
    Maruvada, Subha
    Herman, Bruce A.
    Wear, Keith A.
    Harris, Gerald R.
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2011, 58 (07) : 1397 - 1405
  • [7] AN IMPROVED TISSUE-MIMICKING POLYACRYLAMIDE HYDROGEL PHANTOM FOR VISUALIZING THERMAL LESIONS WITH HIGH-INTENSITY FOCUSED ULTRASOUND
    Guntur, Sitaramanjaneya Reddy
    Choi, Min Joo
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2014, 40 (11): : 2680 - 2691
  • [8] Analysis of Temperature Rise Induced by High-Intensity Focused Ultrasound in Tissue-Mimicking Gel Considering Cavitation Bubbles
    Asai, Ayumu
    Okano, Hiroki
    Yoshizawa, Shin
    Umemura, Shin-ichiro
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2013, 52 (07)
  • [9] SIMULTANEOUS EVALUATION OF THERMAL AND NON-THERMAL EFFECTS OF HIGH-INTENSITY FOCUSED ULTRASOUND ON A TISSUE-MIMICKING PHANTOM
    Kim, Yong Tae
    Ma, Donghee
    Sim, Jai Kyoung
    Kim, Se-Hwa
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2018, 44 (08): : 1799 - 1809
  • [10] Prediction and Measurement of the Size of Thermal Lesion Induced by High Intensity Focused Ultrasound in a Tissue-Mimicking Phantom
    Lee, Kang Il
    Choi, Min Joo
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2009, 48 (02)