Influence of pinhole size and fiber-optic imaging bundle on resolution in confocal endoscopic imaging system

被引:0
|
作者
Ge, HY [1 ]
Ren, QS [1 ]
Wang, BH [1 ]
Li, WR [1 ]
机构
[1] Shanghai Jiao Tong Univ, Inst Laser Med & Bio Photon, Shanghai 200030, Peoples R China
关键词
confocal imaging; high resolution; point spread function (PSF); half width at half maximum (HWHM);
D O I
10.1117/12.575698
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A novel confocal endoscopic imaging system is designed for imaging thick sample. The lateral and axial detected image intensity is deduced and calculated based on point spread function (PSF). The analysis of influence of pinhole and fiber on resolution is given. Our researches show that the lateral resolution improvement is much more sensitive to pinhole size than the axial resolution. Confocal system has narrower half width at half maximum (HWHM) and higher lateral and axial resolution for smaller pinhole. At last a criterion for pinhole size is introduced first of all to ensure confocal imaging and maximize signal-to-noise ratio, which is very helpful in the practical design of confocal endoscopic imaging system.
引用
收藏
页码:506 / 513
页数:8
相关论文
共 50 条
  • [31] Fiber-optic imaging for in-situ chemical measurements
    Carter, JC
    Egan, WJ
    Nair, RB
    Murphy, CJ
    Morgan, SL
    Angel, SM
    [J]. CHEMICAL, BIOCHEMICAL, AND ENVIRONMENTAL FIBER SENSORS X, 1999, 3540 : 210 - 221
  • [32] Endoscopic imaging of quantum gases through a fiber bundle
    Benedicto-Orenes, Daniel
    Kowalczyk, Anna
    Bongs, Kai
    Barontini, Giovanni
    [J]. OPTICS EXPRESS, 2017, 25 (17): : 19701 - 19710
  • [33] High resolution imaging of impacted CFRP composites with a fiber-optic laser-ultrasound scanner
    Pelivanov, Ivan
    Ambrozinski, Lukasz
    Khomenko, Anton
    Koricho, Ermias G.
    Cloud, Gary L.
    Haq, Mahmoodul
    O'Donnell, Matthew
    [J]. PHOTOACOUSTICS, 2016, 4 (02): : 55 - 64
  • [34] In Vivo Quantitative Microvasculature Phenotype Imaging of Healthy and Malignant Tissues Using a Fiber-Optic Confocal Laser Microprobe
    Lin, Ken Young
    Maricevich, Marco
    Bardeesy, Nabeel
    Weissleder, Ralph
    Mahmood, Umar
    [J]. TRANSLATIONAL ONCOLOGY, 2008, 1 (02): : 84 - 94
  • [35] Spatial compound imaging for fiber-bundle optic microscopy
    Cheon, Gyeong Woo
    Cha, Jaepyeong
    Kang, Jin U.
    [J]. OPTICAL FIBERS AND SENSORS FOR MEDICAL DIAGNOSTICS AND TREATMENT APPLICATIONS XIV, 2014, 8938
  • [36] Spatial imaging identification in a fiber-bundle confocal fluorescence microspectrometer
    Li, YU
    Sasaki, SY
    Inoue, T
    Harata, A
    Ogawa, T
    [J]. APPLIED SPECTROSCOPY, 1998, 52 (08) : 1111 - 1114
  • [37] Scintillating Balloon-Enabled Fiber-Optic System for Radionuclide Imaging of Atherosclerotic Plaques
    Zaman, Raiyan T.
    Kosuge, Hisanori
    Carpenter, Colin
    Sun, Conroy
    McConnell, Michael V.
    Xing, Lei
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (05) : 771 - 777
  • [38] Feasibility study on the development of a fiber-optic gamma imaging probe
    Seunghan Hong
    Wook Jae Yoo
    Sang Hun Shin
    Hyesu Jeon
    Jae Seok Jang
    Guwon Kwon
    Dong Eun Lee
    Kyoung Won Jang
    Bongsoo Lee
    [J]. Journal of the Korean Physical Society, 2015, 67 : 175 - 179
  • [39] Fiber-Optic Cardiorespiratory Monitoring and Triggering in Magnetic Resonance Imaging
    Ladrova, Martina
    Nedoma, Jan
    Martinek, Radek
    Behbehani, Khosrow
    Kahankova, Radana
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2022, 71
  • [40] Floor Pressure Imaging for Fall Detection with Fiber-Optic Sensors
    Feng, Guodong
    Mai, Jiechao
    Ban, Zhen
    Guo, Xuemei
    Wang, Guoli
    [J]. IEEE PERVASIVE COMPUTING, 2016, 15 (02) : 40 - 47