An apparatus with a horizontal capillary tube intended for measurement of the surface tension of supercooled liquids

被引:3
|
作者
Vins, Vaclav [1 ]
Hosek, Jan [1 ]
Hykl, Jiri [1 ]
Hruby, Jan [1 ]
机构
[1] Inst Thermomech AS CR, Vvi, Prague 18200 8, Czech Republic
关键词
WATER;
D O I
10.1051/epjconf/20159202108
中图分类号
O3 [力学];
学科分类号
08 ; 0801 ;
摘要
New experimental apparatus for measurement of the surface tension of liquids under the metastable supercooled state has been designed and assembled in the study. The measuring technique is similar to the method employed by P.T. Hacker [NACA TN 2510] in 1951. A short liquid thread of the liquid sample was sucked inside a horizontal capillary tube partly placed in a temperature-controlled glass chamber. One end of the capillary tube was connected to a setup with inert gas which allowed for precise tuning of the gas overpressure in order of hundreds of Pa. The open end of the capillary tube was precisely grinded and polished before the measurement in order to assure planarity and perpendicularity of the outer surface. The liquid meniscus at the open end was illuminated by a laser beam and observed by a digital camera. Application of an increasing overpressure of the inert gas at the inner meniscus of the liquid thread caused variation of the outer meniscus such that it gradually changed from concave to flat and subsequently convex shape. The surface tension at the temperature of the inner meniscus could be evaluated from the overpressure corresponding to exactly planar outer meniscus. Detailed description of the new setup together with results of the preliminary tests is provided in the study.
引用
收藏
页数:6
相关论文