Image quality and radiation dose in planar imaging - Image quality figure of merits from the CDRAD phantom

被引:9
|
作者
Konst, Bente [1 ]
Weedon-Fekjaer, Harald [2 ]
Bath, Magnus [3 ,4 ]
机构
[1] Vestfold Hosp Trust, Dept Radiol, Tonsberg, Norway
[2] Oslo Univ Hosp, Oslo Ctr Biostat & Epidemiol, Res Support Serv, Oslo, Norway
[3] Sahlgrens Univ Hosp, Dept Med Phys & Biomed Engn, Gothenburg, Sweden
[4] Univ Gothenburg, Sahlgrenska Acad, Inst Clin Sci, Dept Radiat Phys, Gothenburg, Sweden
来源
关键词
CD-curve; contrast detail phantom; figure of merit; IQF; planar imaging; precision; LOW-CONTRAST DETAIL; DIGITAL CHEST RADIOGRAPHY; DETECTOR; SYSTEMS; PERFORMANCE; RESOLUTION;
D O I
10.1002/acm2.12649
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A contrast-detail phantom such as CDRAD is frequently used for quality assurance, optimization of image quality, and several other purposes. However, it is often used without considering the uncertainty of the results. The aim of this study was to assess two figure of merits (FOM) originating from CDRAD regarding the variations of the FOMs by dose utilized to create the x-ray image. The probability of overlapping (assessing an image acquired at a lower dose as better than an image acquired at a higher dose) was determined. Methods: The CDRAD phantom located underneath 12, 20, and 26 cm PMMA was imaged 16 times at five dose levels using an x-ray system with a flat-panel detector. All images were analyzed by CDRAD Analyser, version 1.1, which calculated the FOM inverse image quality figure (IQF(inv)) and gave contrast detail curves for each image. Inherent properties of the CDRAD phantom were used to derive a new FOM h, which describes the size of the hole with the same diameter and depth that is just visible. Data were analyzed using heteroscedastic regression of mean and variance by dose. To ease interpretation, probabilities for overlaps were calculated assuming normal distribution, with associated bootstrap confidence intervals. Results: The proportion of total variability in IQF(inv), explained by the dose (R-2), was 91%, 85%, and 93% for 12, 20, and 26 cm PMMA. Corresponding results for h were 91%, 89%, and 95%. The overlap probability for different mAs levels was 1% for 0.8 vs 1.2 mAs, 5% for 1.2 vs 1.6 mAs, 10% for 1.6 vs 2.0 mAs, and 10% for 2.0 mAs vs 2.5 mAs for 12 cm PMMA. For 20 cm PMMA, it was 0.5% for 10 vs 16 mAs, 13% for 16 vs 20 mAs, 14% for 20 vs 25 mAs, and 14% for 25 vs 32 mAs. For 26 cm PMMA, the probability varied from 0% to 6% for various mAs levels. Even though the estimated probability for overlap was small, the 95% confidence interval (CI) showed relatively large uncertainties. For 12 cm PMMA, the associated CI for 0.8 vs 1.2 mAs was 0.1-3.2%, and the CI for 1.2 vs 1.6 mAs was 2.1-7.8%. Conclusions: Inverse image quality figure and h are about equally related to dose level. The FOM h, which describes the size of a hole that should be seen in the image, may be a more intuitive FOM than IQFinv. However, considering the probabilities for overlap and their confidence intervals, the FOMs deduced from the CDRAD phantom are not sensitive to dose. Hence, CDRAD may not be an optimal phantom to differentiate between images acquired at different dose levels.
引用
收藏
页码:151 / 159
页数:9
相关论文
共 50 条
  • [1] Image quality and radiation dose in spinal surgery: a comparison of three imaging systems in phantom
    Casiraghi, Margherita
    Piliero, Maria Antonietta
    Bellesi, Luca
    Dufour, Mathieu
    Chatterjea, Anindita
    Scarone, Pietro
    Presilla, Stefano
    JOURNAL OF MEDICAL IMAGING, 2023, 10 (03)
  • [2] Radiation dose and image quality optimisation in medical imaging
    Hogg, Peter
    JOURNAL OF MEDICAL RADIATION SCIENCES, 2015, 62 (03) : 236 - 237
  • [3] A phantom study of image quality versus radiation dose for digital radiography
    Tung, C. J.
    Tsai, H. Y.
    Shi, M. Y.
    Huang, T. T.
    Yang, C. H.
    Chen, I. J.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (01): : 602 - 605
  • [4] Development of a computed tomography phantom to evaluate radiation dose and image quality
    Gomez Cardona, Daniel
    Estrada Leon, Edgar Santiago
    Zuluaga Santamaria, Alejandro
    Benitez, Luis
    Ramirez Giraldo, Juan Carlos
    2013 PAN AMERICAN HEALTH CARE EXCHANGES (PAHCE), 2013,
  • [5] Optimising image quality and radiation dose for neonatal incubator imaging
    Tugwell-Allsup, J.
    Morris, R. W.
    Hibbs, R.
    England, A.
    RADIOGRAPHY, 2020, 26 (04) : E258 - E263
  • [6] Editorial: Radiation dose and image quality
    Manning, David
    RADIOGRAPHY, 2014, 20 (04) : 291 - 292
  • [7] CT radiation dose and image quality
    Payne, JT
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2005, 43 (06) : 953 - +
  • [8] Mammography radiation dose and image quality
    Jacobson, DR
    RADIATION PROTECTION DOSIMETRY, 1998, 80 (1-3) : 295 - 297
  • [9] Tradeoffs in image quality and radiation dose
    McNitt-Gray, M
    MEDICAL PHYSICS, 2004, 31 (06) : 1802 - 1802
  • [10] Average glandular dose and phantom image quality in mammography
    Oliveira, M.
    Noguelra, M. S.
    Guedes, E.
    Andrade, M. C.
    Peixoto, J. E.
    Joana, G. S.
    Castro, J. G.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (01): : 574 - 577