Calibration of Photoelectric Pulsed Radiation Detectors

被引:0
|
作者
Kireev, S. G. [1 ]
Tumashevich, K. A. [1 ]
Shashkovskiy, S. G. [1 ]
Abakumov, A. V. [1 ]
Osin, N. A. [1 ]
机构
[1] Sci & Ind Enterprise Melitta Ltd, Moscow, Russia
关键词
xenon flash lamp; photodiode; calibration; continuous spectrum radiation;
D O I
10.1007/s11182-020-01943-6
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
A method for calibrating a photodiode intended for high-intensity radiation pulse detection using continuous lighting lamps is described. The method is based on the use of an attachable integrated circuit with a known time constant. Based on the volt sensitivity and the time constant obtained during calibration against a continuous lighting lamp, relationships of the power and radiation energy with the photodetector response are mathematically derived. The method has been tested using a xenon flash lamp with pulse duration at half maximum of approximate to 60 mu s.