Objectives: To study the agreement of serum cobalamin results obtained with three different widely used analyzers. Design and methods: Comparative measurement of 100 serum samples was performed using the Abbott Architect, the Bayer Centaur, and the Roche Elecsys 2010 analyzer. Results: Relatively close numeric correlation of the results was found with Pearson's r>0.95 for all three comparisons; however, the mean concentration of the highest reading and the lowest reading method differed by 16%. Conclusions: Relatively good agreement of results from three automated serum cobalamin assays was observed. (C) 2007 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.