Investigating ion recombination effects in a liquid-filled ionization chamber array used for IMRT QA measurements

被引:6
|
作者
Knill, Cory [1 ,2 ]
Snyder, Michael [1 ,2 ]
Rakowski, Joseph T. [1 ,2 ]
Zhuang, Ling [2 ]
Matuszak, Martha [3 ]
Burmeister, Jay [1 ,2 ]
机构
[1] Karmanos Canc Inst, Dept Radiat Oncol, Detroit, MI 48201 USA
[2] Wayne State Univ, Sch Med, Dept Radiat Oncol, Detroit, MI 48201 USA
[3] Univ Michigan Hlth Syst, Dept Radiat Oncol, Ann Arbor, MI 48109 USA
关键词
1000; SRS; ion recombination; Octavius; 4D; BODY RADIATION-THERAPY; COLLECTION EFFICIENCY; SMALL FIELDS; DOSIMETRY; BEAMS;
D O I
10.1118/1.4946822
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: PTW's Octavius 1000 SRS array performs IMRT quality assurance (QA) measurements with liquid-filled ionization chambers (LICs) to allow closer detector spacing and higher resolution, compared to air-filled QA devices. However, reduced ion mobility in LICs relative to air leads to increased ion recombination effects and reduced collection efficiencies that are dependent on Linac pulse frequency and pulse dose. These pulse parameters are variable during an IMRT delivery, which affects QA results. In this study, (1) 1000 SRS collection efficiencies were measured as a function of pulse frequency and pulse dose, (2) two methods were developed to correct changes in collection efficiencies during IMRT QA measurements, and the effects of these corrections on QA pass rates were compared. Methods: To obtain collection efficiencies, the OCTAVIUS 1000 SRS was used to measure open fields of varying pulse frequency, pulse dose, and beam energy with results normalized to air-filled chamber measurements. Changes in ratios of 1000 SRS to chamber measured dose were attributed to changing collection efficiencies, which were then correlated to pulse parameters using regression analysis. The usefulness of the derived corrections was then evaluated using 6 MV and 10FFF SBRT RapidArc plans delivered to the OCTAVIUS 4D system using a TrueBeam (Varian Medical Systems) linear accelerator equipped with a high definition multileaf collimator. For the first correction, MATLAB software was developed that calculates pulse frequency and pulse dose for each detector, using measurement and DICOM RT Plan files. Pulse information is converted to collection efficiency, and measurements are corrected by multiplying detector dose by ratios of calibration to measured collection efficiencies. For the second correction the MU/min in the daily 1000 SRS calibration was chosen to match the average MU/min of the volumetric modulated arc therapy plan. Effects of the two corrections on QA results were examined by performing 3D gamma analysis comparing predicted to measured dose, with and without corrections. Results: Collection efficiencies correlated linearly to pulse dose, while correlations with pulse frequency were less defined, generally increasing as pulse frequency decreased. After complex MATLAB corrections, average 3D gamma pass rates improved by [0.07%, 0.40%, 1.17%] for 6 MV and [0.29%, 1.40%, 4.57%] for 10FFF using [3%/3 mm, 2%/2 mm, 1%/1 mm] criteria. Maximum changes in gamma pass rates were [0.43%, 1.63%, 3.05%] for 6 MV and [1.00%, 4.80%, 11.2%] for 10FFF using [3%/3 mm, 2%/2 mm, 1%/1 mm] criteria. On average, pass rates of simple daily calibration corrections were within 1% of complex MATLAB corrections. Conclusions: OCTAVIUS 1000 SRS ion recombination effects have little effect on 6 MV measurements. However, the effect could potentially be clinically significant for higher pulse dose unflattened beams when using tighter gamma tolerances, especially when small aperture sizes are used, as is common for SRS/SBRT. In addition, ion recombination effects are strongly correlated to changing MU/min, therefore MU/min used in daily 1000 SRS calibrations should be matched to the expected average MU/min of the IMRT plan. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:2476 / 2484
页数:9
相关论文
共 50 条
  • [21] PHYSICAL ASPECTS OF A LIQUID-FILLED IONIZATION-CHAMBER WITH PULSED POLARIZING VOLTAGE
    VANHERK, M
    MEDICAL PHYSICS, 1991, 18 (04) : 692 - 702
  • [22] On-line beam monitoring and dose profile measurements of a 208Pb beam of 150 GeV/n with a liquid-filled ionization chamber array
    Wyrwoll, V.
    Delfs, B.
    Lapp, M.
    Poppinga, D.
    Martinez, P. Fernandez
    Kastriotou, M.
    Alia, R. Garcia
    Roed, K.
    Gerbershagen, A.
    Looe, H. K.
    Poppe, B.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2021, 987
  • [23] On-line beam monitoring and dose profile measurements of a 208Pb beam of 150 GeV/n with a liquid-filled ionization chamber array
    Wyrwoll, V.
    Delfs, B.
    Lapp, M.
    Poppinga, D.
    Fernandéz Martinéz, P.
    Kastriotou, M.
    García Alía, R.
    Røed, K.
    Gerbershagen, A.
    Looe, H.K.
    Poppe, B.
    Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2021, 987
  • [24] NanoLIC: A Liquid-Filled Ionization Chamber for Experimental Verification of Dose Enhancement with Gold Nanoparticles
    Jimenez-Ortega, E.
    Gonzalez-Castano, D.
    Lopez-Valverde, J.
    Gomez, F.
    Leal Plaza, A.
    MEDICAL PHYSICS, 2022, 49 (06) : E327 - E327
  • [25] Verification of high-dose-rate brachytherapy treatment planning dose distribution using liquid-filled ionization chamber array
    Yoosuf, A. B. Mohamed
    Jeevanandam, Prakash
    Whitten, Glenn
    Workman, Geraldine
    McGarry, Conor K.
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2018, 10 (02) : 142 - 154
  • [26] Development and first tests of a 0.5 mm3 volume liquid-filled ionization chamber
    Gonzalez-Castano, D. M.
    Gomez, F.
    Pena, J.
    Rosello, J. V.
    Brualla, L.
    Pazos, A.
    Zapata, M.
    Iglesias, A.
    Sanchez-Doblado, F.
    Pombar, M.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S114 - S114
  • [27] The stability of liquid-filled matrix ionization chamber electronic portal imaging devices for dosimetry purposes
    Louwe, RJW
    Tielenburg, R
    van Ingen, KM
    Mijnheer, BJ
    van Herk, MB
    MEDICAL PHYSICS, 2004, 31 (04) : 819 - 827
  • [28] NEW HIGH RESOLUTION 2D LIQUID FILLED IONIZATION CHAMBER ARRAY FOR SMALL FIELD MEASUREMENTS
    Brualla Gonzalez, L.
    Vicedo, A.
    Gonzalez-Castano, D. M.
    Gago Arias, A.
    Pardo Montero, J.
    Pazos Alvarez, A.
    Bermudez, M.
    Gomez, F.
    Rosello, J.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S146 - S146
  • [29] A two-dimensional liquid-filled ionization chamber array prototype for small-field verification: characterization and first clinical tests
    Brualla-Gonzalez, Luis
    Gomez, Faustino
    Vicedo, Aurora
    Gonzalez-Castano, Diego M.
    Gago-Arias, Araceli
    Pazos, Antonio
    Zapata, Martin
    Rosello, Joan V.
    Pardo-Montero, Juan
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (16): : 5221 - 5234
  • [30] CHARACTERIZATION OF THE TEMPERATURE, POLARIZATION, AND ION RECOMBINATION FOR THE MICROLION LIQUID IONIZATION CHAMBER
    Gutierrez, A.
    Stathakis, S.
    Papanikolaou, N.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S169 - S169