Cone beam computed tomography and image registration based on target area for stereotactic body radiation therapy of lung cancer

被引:1
|
作者
Gui, Long-Gang [1 ]
Shi, Miao [1 ]
Li, Jun [1 ]
机构
[1] Northern Jiangsu Peoples Hosp, Radiotherapy Ctr, Yangzhou 225001, Jiangsu, Peoples R China
关键词
Lung cancer; SBRT; CBCT; image registration; dosimetric parameter; INTENSITY-MODULATED RADIOTHERAPY; CLINICAL-OUTCOMES; DIAGNOSTIC-VALUE; CT; IRRADIATION;
D O I
10.1080/16878507.2021.1981754
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
This paper aims to compare the dosimetric parameters of the original plan and the simulated plan to explore the clinical value of cone beam CT and planned CT image registration based on the target area. The results showed that Dmax, Dmin, dmean and V60 of PTV in the original plan were (108.3 +/- 1.1)%, (97.2 +/- 0.3)%, (102.6 +/- 0.6)% and (95.2 +/- 0.6)% respectively. The Dmax of spinal cord was (9.6 +/- 1.8) gy, the V20 of lung was (5.2 +/- 1.9)%, and the dmean of lung was (- 3.7 +/- 1.5) gy; however, when the target area was translated in the three directions of the patient's back, right side and foot, the dosimetric parameters of the target area (Dmax, Dmin and V60), dmean and V20 of lung and Dmax of spinal cord showed slight differences, which all met the limit requirements of target area and dangerous organs (OARs) (RTOG) 0915 report. Therefore, target based image registration is feasible. If the registration translation error is less than or equal to 1 cm, there is no need to re move the treatment bed according to the registration error.
引用
收藏
页码:404 / 411
页数:8
相关论文
共 50 条
  • [41] Cone Beam Computed Tomography (CBCT) Verification of Stereotactic Body Radiation Therapy (SBRT): A Prospective Analysis on the Concordance Between Radiation Therapists and Radiation Oncologists
    Yeung, R.
    Sidney, J.
    Ferris, R.
    Soo, J.
    Liu, M. C.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E540 - E540
  • [42] Patient-specific synthetic magnetic resonance imaging generation from cone beam computed tomography for image guidance in liver stereotactic body radiation therapy
    Zhang, Zeyu
    Jiang, Zhuoran
    Zhong, Hualiang
    Lu, Ke
    Yin, Fang-Fang
    Ren, Lei
    PRECISION RADIATION ONCOLOGY, 2022, 6 (02): : 110 - 118
  • [43] Dosimetric Uncertainties Resulting From Interfractional Anatomic Variations for Patients Receiving Pancreas Stereotactic Body Radiation Therapy and Cone Beam Computed Tomography Image Guidance
    Niedzielski, Joshua S.
    Liu, Yufei
    Ng, Sylvia S. W.
    Martin, Rachael M.
    Perles, Luis A.
    Beddar, Sam
    Rebueno, Neal
    Koay, Eugene J.
    Taniguchi, Cullen
    Holliday, Emma B.
    Das, Prajnan
    Smith, Grace L.
    Minsky, Bruce D.
    Ludmir, Ethan B.
    Herman, Joseph M.
    Koong, Albert
    Sawakuchi, Gabriel O.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 111 (05): : 1298 - 1309
  • [44] Intrafraction cone beam computed tomography verification of breath hold during liver stereotactic radiation therapy
    Brown, Elizabeth
    Muscat, Erika
    O'Connor, Patrick
    Liu, Howard
    Lee, Yoo-Young
    Pryor, David
    JOURNAL OF MEDICAL RADIATION SCIENCES, 2021, 68 (01) : 52 - 59
  • [45] Quantitative evaluation of cone-beam computed tomography in target volume definition for offline image-guided radiation therapy of prostate cancer
    Wang, Weihu
    Wu, Qiuwen
    Yan, Di
    RADIOTHERAPY AND ONCOLOGY, 2010, 94 (01) : 71 - 75
  • [46] Cone-beam computed tomography imaging in stereotactic body radiotherapy allows for more than target localization
    Bakri, Bonnie
    Verbakel, Wilko
    Slotman, Ben
    Dahele, Max
    JOURNAL OF RADIOSURGERY AND SBRT, 2013, 2 (02): : 141 - 145
  • [47] Optimization of four-dimensional cone-beam computed tomography in image-guided radiation therapy of the lung
    Ahmad, Moiz
    Pan, Tinsu
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [48] Evaluation of Lung Toxicity Risk with Computed Tomography Ventilation Functional Image for Lung Stereotactic Body Radiation Therapy and Three-Dimensional Conformal Radiation Therapy
    Otsuka, M.
    Monzen, H.
    Kadoya, N.
    Inada, M., IV
    Matsumoto, K.
    Nishimura, Y.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E708 - E708
  • [49] Evaluation of Image Enhancement Method on Target Registration Using Cone Beam CT in Radiation Therapy
    Yan, Hui
    Lei, Ren
    Wu, Jackie
    Di, Fu
    Yin, Fang-Fang
    CLINICAL MEDICINE INSIGHTS-ONCOLOGY, 2008, 2 : 289 - 299
  • [50] Comparison between manual and automatic image registration in image-guided radiation therapy using megavoltage cone-beam computed tomography with an imaging beam line for prostate cancer
    Hashido T.
    Nakasone S.
    Fukao M.
    Ota S.
    Inoue S.
    Radiological Physics and Technology, 2018, 11 (4) : 392 - 405