Transcutaneous monitoring as a replacement for arterial PCO2 monitoring during nocturnal non-invasive ventilation

被引:91
|
作者
Storre, Jan Hendrik [1 ]
Magnet, Friederike Sophie [1 ]
Dreher, Michael [1 ]
Windisch, Wolfram [1 ]
机构
[1] Univ Hosp Freiburg, Dept Pneumol, D-79106 Freiburg, Germany
关键词
Alveolar ventilation; Carbon dioxide; Monitoring; Nocturnal ventilation; Sleep; Transcutaneous pressure of carbone dioxid; STABLE HYPERCAPNIC COPD; POSITIVE-PRESSURE VENTILATION; CHRONIC RESPIRATORY-FAILURE; RANDOMIZED CROSSOVER TRIAL; CARBON-DIOXIDE; MECHANICAL VENTILATION; CRITICALLY-ILL; CLINICAL-EVALUATION; NASAL VENTILATION; HIGH-INTENSITY;
D O I
10.1016/j.rmed.2010.10.007
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Continuous, non-invasive assessment of alveolar ventilation achieved by transcutaneous PCO2 (PtcCO(2)) monitoring is clearly superior to intermittent, invasive blood gas analyses in patients receiving nocturnal non-invasive positive pressure ventilation (NPPV), but the reliability and accuracy of PtcCO(2)-monitoring is still disputed. The present study was aimed at investigating the capability of modern PtcCO(2)-monitoring to reliably assess alveolar ventilation during nocturnal NPPV. Methods: Capillary blood gas measurements (11pm, 2am, 5am and 7am) and 8 h of continuous PtcCO(2)-monitoring using three of the latest generation devices (SenTec Digital Monitor, Radiometer TCM4-TINA and Radiometer TOSCA500) were performed during polysomnography-proven sleep studies in 24 patients receiving NPPV (15 with COPD, 9 with restrictive disorders). Results: The technical calibration drift for SenTec DM, TCM4-TINA and TOSCA500 was 0.1, -0.4 and -0.5 mmHg/h, respectively. Bland-Altman method comparison of PaCO2/drift-uncorrected PtcCO(2) revealed a mean bias (limits of agreement) of 1.0 (-4.7 to 6.7), -1.5 (-15.6 to 12.5) and 0.8 (-6.8 to 8.3) mmHg, respectively. Continuous overnight PtcCO(2)-monitoring detected variations in alveolar ventilation, with median ranges of 12.3 (10.7-14.5) mmHg for SenTec DM, 14.5 (12.5-17.0) mmHg for TCM4-TINA and 11.5 (11.0-13.0) mmHg for TOSCA500 (RM-ANOVA, p < 0.001). The four capillary PaCO2 values ranged by a median of 6.3 (4.7-9.7) mmHg. Conclusions: Modern PtcCO(2)-monitoring is reliable, accurate and robust. Since PtcCO(2)-monitoring is also non-invasive, does not disrupt sleep quality and provides a more complete picture of alveolar ventilation than intermittent capillary PaCO2, PtcCO(2)-monitoring should become the preferred technique for assessing alveolar ventilation during nocturnal NPPV. Trial Registration: DRKS00000433 at http://apps.who.int/trialsearch/default.aspx. (C) 2010 Elsevier Ltd. All rights reserved.
引用
收藏
页码:143 / 150
页数:8
相关论文
共 50 条
  • [1] Validity of transcutaneous PCO2 in monitoring chronic hypoventilation treated with non-invasive ventilation
    Aarrestad, Sigurd
    Tollefsen, Elin
    Kleiven, Anne Louise
    Qvarfort, Magnus
    Janssens, Jean-Paul
    Skjonsberg, Ole Henning
    [J]. RESPIRATORY MEDICINE, 2016, 112 : 112 - 118
  • [2] Agreement between arterial and transcutaneous PCO2 in patients undergoing non-invasive ventilation
    Kelly, Anne-Maree
    Klim, Sharon
    [J]. RESPIRATORY MEDICINE, 2011, 105 (02) : 226 - 229
  • [3] Transcutaneous Pco2 monitoring during initiation of noninvasive ventilation
    Storre, Jan H.
    Steurer, Boris
    Kabitz, Hans-Joachim
    Dreher, Michael
    Windisch, Wolfram
    [J]. CHEST, 2007, 132 (06) : 1810 - 1816
  • [4] Non-invasive (transcutaneous) monitoring of PCO2 (TcPCO2) in older adults
    Janssens, JP
    Laszlo, A
    Uldry, C
    Titelion, V
    Picaud, C
    Michel, JP
    [J]. GERONTOLOGY, 2005, 51 (03) : 174 - 178
  • [5] Transcutaneous PCO2 monitoring is accurate for assessing PaCO2 during nocturnal noninvasive ventilation
    Storre, J. H.
    Dreher, M.
    Windisch, W.
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2010, 181
  • [6] Continuous non-invasive PCO2 monitoring in weaning patients: Transcutaneous is advantageous over end-tidal PCO2
    Schwarz, Sarah B.
    Windisch, Wolfram
    Magnet, Friederike S.
    Schmoor, Claudia
    Karagiannidis, Christian
    Callegari, Jens
    Huttmann, Sophie E.
    Storre, Jan H.
    [J]. RESPIROLOGY, 2017, 22 (08) : 1579 - 1584
  • [7] Monitoring of pCO2 during ventilation
    Magnet, F. S.
    Windisch, W.
    Storre, J. H.
    [J]. MEDIZINISCHE KLINIK-INTENSIVMEDIZIN UND NOTFALLMEDIZIN, 2016, 111 (03) : 202 - 207
  • [8] Non-invasive intraoperative monitoring of blood pressure and arterial pCO2 during surgical anesthesia in mice
    Thal, Serge C.
    Plesnila, Nikolaus
    [J]. JOURNAL OF NEUROSCIENCE METHODS, 2007, 159 (02) : 261 - 267
  • [9] NON INVASIVE CUTANEOUS PCO2 MONITORING
    RENAUD, S
    WILLARD, D
    MESSER, J
    EBERHARD, P
    [J]. INTENSIVE CARE MEDICINE, 1980, 6 (01) : 39 - 39
  • [10] NON-INVASIVE CONJUNCTIVAL PCO2 AND PO2 MONITORING DURING HYPERVENTILATION AND HYPOVENTILATION
    KRAM, HB
    APPEL, PL
    FINK, S
    SHOEMAKER, WC
    [J]. RESUSCITATION, 1988, 16 (04) : 263 - 269