Press-fit versus threaded acetabular cups in total hip arthroplasty : Functional and radiological results after five years

被引:0
|
作者
Ellenrieder, Martin [1 ]
Bader, Rainer [1 ]
Bergschmidt, Philipp [1 ]
Mittelmeier, Wolfram [1 ]
机构
[1] Univ Med Rostock, Dept Orthopaed, Doberaner Str 142, D-18057 Rostock, Germany
来源
ACTA ORTHOPAEDICA BELGICA | 2016年 / 82卷 / 01期
关键词
hip; prosthesis; cup; cementless; outcome; CROSS-LINKED POLYETHYLENE; FEMORAL COMPONENTS; RETRIEVAL ANALYSIS; STRESS-ANALYSIS; WEAR; REPLACEMENT; STABILITY; PROSTHESIS; FIXATION; STEM;
D O I
暂无
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Prospectively the outcome after total hip replacement with a new threaded acetabular cup design was compared to an established press-fit cup. After 1, 2 and 5 years, the 36-item Short Form Health Survey, Western Ontario and McMaster University Osteoarthritis Index and Harris Hip Score revealed no significant differences between the two groups (each group : n = 42 patients), except for a higher Harris Hip Score in the threaded cup group after five years (p = 0.02). After five years, one threaded cup had a mild radiolucent line without further signs of loosening. All other cups of both groups (98.6%) showed a full osseous integration. The cup inclination angle ranged from 41-58 degrees (threaded cups) to 39-77 degrees (press-fit cups). The new threaded cup provides equivalent clinical outcomes and osseous integration but more precise implant positioning compared to the press-fit design. No complications typically ascribed to threaded cups (acetabular fractures, bone resorption, nerve impairment) occurred.
引用
收藏
页码:60 / 67
页数:8
相关论文
共 50 条
  • [31] Nonmodular flexible press-fit cup in primary total hip arthroplasty - 15-year followup
    Berli, Bernhard J.
    Ping, Gu
    Dick, Walter
    Morscher, Erwin W.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (461) : 114 - 121
  • [32] Catastrophic wear in uncemented acetabular cups after Total Hip Arthroplasty. A case series
    Jasqui Remba, Salomon
    Esquivel Gomez, Ricardo
    Gutierrez Ambriz, Luis
    Portman Santos, Daniel
    Baley Amiga, Isaac
    [J]. INTERNATIONAL JOURNAL OF SURGERY CASE REPORTS, 2021, 81
  • [33] Primary total hip arthroplasty with a flattened press-fit acetabular component in osteoarthritis and inflammatory arthritis: a prospective study on 416 hips with 6-10 years follow-up
    Zwartele, Rob E.
    Olsthoorn, Paul G. M.
    Poll, Rudolf G.
    Brand, Ronald
    Doets, H. Cornelis
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2008, 128 (12) : 1379 - 1386
  • [34] BONE IMAGING AFTER CEMENTLESS PRESS FIT TOTAL HIP-ARTHROPLASTY
    LETTE, J
    EYBALIN, MC
    ROY, A
    CERINO, M
    PRENOVAULT, J
    DESJARDINS, A
    DUCHESNE, R
    BAH, C
    FALLAHA, M
    LEVASSEUR, A
    [J]. CLINICAL NUCLEAR MEDICINE, 1989, 14 (08) : 614 - 622
  • [35] Clinical and radiographic outcomes after total shoulder arthroplasty with an anatomic press-fit short stem
    Szerlip, Benjamin W.
    Morris, Brent J.
    Laughlin, Mitzi S.
    Kilian, Christopher M.
    Edwards, T. Bradley
    [J]. JOURNAL OF SHOULDER AND ELBOW SURGERY, 2018, 27 (01) : 10 - 16
  • [36] Results of osteosynthesis versus acute total hip arthroplasty for acetabular fractures in the elderly
    Lannes, Xavier
    Steinmetz, Sylvain
    Moerenhout, Kevin
    Borens, Olivier
    [J]. SWISS MEDICAL WEEKLY, 2019, 149 : 55S - 56S
  • [37] Changes in periacetabular bone mineral density five years after resurfacing hip arthroplasty versus conventional total hip arthroplasty
    Gerhardt, Davey M. J. M.
    Smolders, Jose M. H.
    Roovers, Elisabeth A.
    Rijnders, Ton A. J. M.
    van Susante, Job L. C.
    [J]. HIP INTERNATIONAL, 2019, 29 (02) : 153 - 160
  • [38] Randomised, prospective study comparing cemented and cementless total knee replacement - Results of press-fit condylar total knee replacement at five years
    McCaskie, AW
    Deehan, DJ
    Green, TP
    Lock, KR
    Thompson, JR
    Harper, WM
    Gregg, PJ
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (06): : 971 - 975
  • [39] Revision Total Hip Arthroplasty Using a Tapered, Press-Fit Cement less Revision Stem in Elderly Patients
    Chang, Jun-Dong
    Kim, Tae-Young
    Rao, Murari Bhaskar
    Lee, Sang-Soo
    Kim, In-Sung
    [J]. JOURNAL OF ARTHROPLASTY, 2011, 26 (07): : 1045 - 1049
  • [40] Jumbo cups for revision of acetabular defects after total hip arthroplasty: a retrospective review of a case series
    Christian Wedemeyer
    Carl Neuerburg
    Hansjoerg Heep
    Fabian von Knoch
    Marius von Knoch
    Franz Löer
    Guido Saxler
    [J]. Archives of Orthopaedic and Trauma Surgery, 2008, 128 : 545 - 550