Optimal ROI setting on the anatomically normalized I-123 FP-CIT images using high-resolution SPECT

被引:6
|
作者
Nonokuma, Masanari [1 ]
Kuwabara, Yasuo [1 ]
Hida, Kosuke [1 ]
Tani, Tomonobu [1 ]
Takano, Koichi [1 ]
Yoshimitsu, Kengo [1 ]
机构
[1] Fukuoka Univ Hosp, Dept Radiol, Jonan Ku, 7-45-1 Nanakuma, Fukuoka 8140180, Japan
关键词
I-123; FP-CIT; SPECT; Specific binding ratio; Statistical parametric mapping; ROI; BRAIN IMAGES; STANDARDIZATION; QUANTIFICATION; I-123-FP-CIT;
D O I
10.1007/s12149-016-1107-6
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study is to establish the optimal regions of interest (ROIs) in anatomically normalized I-123 FP-CIT SPECT images for the quantification of dopamine transporter binding. The subjects comprised 16 normal controls and 14 Parkinsonian patients. All of the normal control subjects underwent I-123 FP-CIT SPECT and MRI. The SPECT device used in this study was a Toshiba GCA-9300R with triple head detectors. I-123 FP-CIT (148 MBq) was intravenously administered as a bolus, and the SPECT scan started 4 h after the administration. The data were collected over 20 min for each subject, and reconstructed using a 3D-OSEM algorithm. The data were analyzed using SPM8. I-123 FP-CIT SPECT images were anatomically normalized to the MNI space using an I-123 FP-CIT template, and then divided by the background counts automatically measured using the ROIs set for the cerebral cortices. In the normal control subjects, the specific binding ratios of the MRI-based ROIs were lowest in the caudate nucleus, while the ratios of the I-123 FP-CIT-based ROIs were almost the same throughout all three parts. In contrast, in Parkinsonian patients, the specific binding ratios of the I-123 FP-CIT-based ROIs revealed rostrocaudal decline, while those of the MRI-based ROIs were highest in the anterior putamen. We created an ROI template on the anatomically normalized MRI and I-123 FP-CIT images, and concluded that I-123 FP-CIT-based ROIs are more suitable for obtaining quantitative values than MRI-based ones.
引用
下载
收藏
页码:637 / 644
页数:8
相关论文
共 50 条
  • [41] Imaging of serotonin transporters with [123I]FP-CIT SPECT in the human hypothalamus
    Borgers, Anke J.
    Alkemade, Anneke
    Van de Giessen, Elsmarieke M.
    Drent, Madeleine L.
    Booij, Jan
    Bisschop, Peter H.
    Fliers, Eric
    EJNMMI RESEARCH, 2013, 3 : 1 - 7
  • [42] Drooling in Parkinson's Disease: A [123I]FP-CIT SPECT study
    Contreras, V. Suarez
    Tayyabah, Y.
    Caminiti, S.
    Politis, M.
    MOVEMENT DISORDERS, 2019, 34
  • [43] Japanese multicenter database of healthy controls for [123I]FP-CIT SPECT
    Matsuda, Hiroshi
    Murata, Miho
    Mukai, Yohei
    Sako, Kazuya
    Ono, Hidetoshi
    Toyama, Hiroshi
    Inui, Yoshitaka
    Taki, Yasuyuki
    Shimomura, Hideo
    Nagayama, Hiroshi
    Tateno, Amane
    Ono, Kenjiro
    Murakami, Hidetomo
    Kono, Atsushi
    Hirano, Shigeki
    Kuwabara, Satoshi
    Maikusa, Norihide
    Ogawa, Masayo
    Imabayashi, Etsuko
    Sato, Noriko
    Takano, Harumasa
    Hatazawa, Jun
    Takahashi, Ryosuke
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2018, 45 (08) : 1405 - 1416
  • [44] Acquired hepatocerebral degeneration and dopamine transporter imaging using [123I]-FP-CIT SPECT
    Kim, J.-M.
    Kim, Y. K.
    Kim, S. E.
    Jeon, B. S.
    MOVEMENT DISORDERS, 2007, 22 : S257 - S257
  • [45] [I-123]FP-CIT SPECT discriminates between groups of early nonmedicated patients with Parkinson's disease and controls.
    Booij, J
    Tissingh, G
    Bergmans, P
    Costa, DC
    Winogrodzka, A
    Janssen, AGM
    Wolters, EC
    Stoof, JC
    vanRoyen, EA
    JOURNAL OF NUCLEAR MEDICINE, 1997, 38 (05) : 518 - 518
  • [46] Quantification of [123I]FP-CIT SPECT brain images:: an accurate technique for measurement of the specific binding ratio
    Tossici-Bolt, Livia
    Hoffmann, Sandra M. A.
    Kemp, Paul M.
    Mehta, Rajnikant L.
    Fleming, John S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2006, 33 (12) : 1491 - 1499
  • [47] Quantification of [123I]FP-CIT SPECT brain images: an accurate technique for measurement of the specific binding ratio
    Livia Tossici-Bolt
    Sandra M. A. Hoffmann
    Paul M. Kemp
    Rajnikant L. Mehta
    John S. Fleming
    European Journal of Nuclear Medicine and Molecular Imaging, 2006, 33 : 1491 - 1499
  • [48] [123I]FP-CIT SPECT reconstruction: FBP versus OSEM with resolution recovery - a quantitative and qualitative comparison
    Teixeira, R.
    Castanheira, J.
    Silva, A.
    Vaz, S.
    Silva, M.
    Canudo, A.
    Oliveira, R.
    Correia, J.
    Delgado, H.
    Andrade, M.
    Freitas, B.
    Teixeira, S.
    Vieira, V.
    Machado, M.
    Constantino, C.
    Oliveira, F.
    Costa, D. C.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2021, 48 (SUPPL 1) : S496 - S497
  • [49] Unexpected (123I)FP-CIT SPECT findings: SWIDD, SWEDD and all DAT
    Balestrino Roberta
    Barone Paolo
    Filippi Massimo
    Erro Roberto
    Journal of Neurology, 2022, 269 : 758 - 770
  • [50] 123I FP-CIT SPECT AND TRANSCRANIAL SONOGRAPHY FINDINGS IN PATIENTS WITH MOVEMENT DISORDERS
    Bartova, P.
    Skoloudik, D.
    Kraft, O.
    Bernatek, J.
    Havel, M.
    Ressner, P.
    EUROPEAN JOURNAL OF NEUROLOGY, 2011, 18 : 63 - 63