Radiation dose and fluoroscopy time of conventional versus fluoroscopic computer assisted pedicle screw implantation

被引:17
|
作者
Linhardt, O
Perlick, L
Lüring, C
Stern, U
Plitz, W
Grifka, J
机构
[1] Univ Regensburg, Orthopad Klin, D-93074 Bad Abbach, Germany
[2] Univ Regensburg, Inst Rontgendiagnost Klinikum, D-93074 Bad Abbach, Germany
[3] Univ Munich, Labor Biomech & Expt Orthopad, Munich, Germany
来源
关键词
pedicle screws; computer-assisted surgery; fluoroscopy; radiacopy dose; fluoroscopy time;
D O I
10.1055/s-2005-836489
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Aim: Goal of the current study was to compare radiation dose and fluoroscopy time of fluoroscopic computer assisted pedicle screw implantation versus the conventional technique. Method: For each of 10 specimens two pedicle screws were placed using conventional technique (group 1) and two screws were inserted with fluoroscopic navigation system (group 2) contralateraly. Results: For implantation of two pedicle screws the mean radiation dose was 0.041 mSv in group 1 and 0.029 mSv in group 2. Fluoroscopy time was 34 seconds in group 1 and 25 seconds in group 2. The differences of radiation dose and fluoroscopy time for group 1 and, 2 were statistically significant (radiation dose p = 0.00044, fluoroscopy time p = 0.00039). Conclusion: We achieved significantly lower radiation dose and fluoroscopy time with fluoroscopic computer assisted pedicle screw implantation compared with the conventional technique. Concerning exposure to radiation for patients and personnels fluoroscopic navigated screw insertion is to favour.
引用
收藏
页码:175 / 178
页数:4
相关论文
共 50 条
  • [41] COMPUTER-ASSISTED PEDICLE SCREW FIXATION - A FEASIBILITY STUDY
    AMIOT, LP
    LABELLE, H
    DEGUISE, JA
    SATI, M
    BRODEUR, P
    RIVARD, CH
    SPINE, 1995, 20 (10) : 1208 - 1212
  • [42] Comparison of Contrast Volume, Radiation Dose, Fluoroscopy Time, and Procedure Time in Previously Published Studies of Rotational Versus Conventional Coronary Angiography
    Loomba, Rohit S.
    Rios, Rodrigo
    Buelow, Matthew
    Eagam, Mamata
    Aggarwal, Saurabh
    Arora, Rohit R.
    AMERICAN JOURNAL OF CARDIOLOGY, 2015, 116 (01): : 43 - 49
  • [43] Development of a computer-aided training system for pedicle screw implantation
    Fang, Jing-Jing
    Yang, Chyun-Yu
    Lin, Ruey-Mo
    Biomedical Engineering - Applications, Basis and Communications, 2007, 19 (05) : 323 - 329
  • [44] DEVELOPMENT OF A COMPUTER-AIDED TRAINING SYSTEM FOR PEDICLE SCREW IMPLANTATION
    Fang, Jing-Jing
    Yang, Chyun-Yu
    Lin, Ruey-Mo
    BIOMEDICAL ENGINEERING-APPLICATIONS BASIS COMMUNICATIONS, 2007, 19 (05): : 323 - 329
  • [45] DIGITAL FLUOROSCOPY WITH A CONVENTIONAL FLUOROSCOPIC ROOM AND A NUCLEAR-MEDICINE COMPUTER-SYSTEM
    OCONNOR, MK
    MOLLOY, M
    MAHER, K
    MALONE, JF
    BRITISH JOURNAL OF RADIOLOGY, 1984, 57 (679): : 553 - 560
  • [46] Robot-assisted and fluoroscopy-guided pedicle screw placement: a systematic review
    Marcus, Hani J.
    Cundy, Thomas P.
    Nandi, Dipankar
    Yang, Guang-Zhong
    Darzi, Ara
    EUROPEAN SPINE JOURNAL, 2014, 23 (02) : 291 - 297
  • [47] Robot-assisted and fluoroscopy-guided pedicle screw placement: a systematic review
    Hani J. Marcus
    Thomas P. Cundy
    Dipankar Nandi
    Guang-Zhong Yang
    Ara Darzi
    European Spine Journal, 2014, 23 : 291 - 297
  • [48] TiRobot‑assisted versus conventional fluoroscopy-assisted percutaneous sacroiliac screw fixation for pelvic ring injuries: a meta‑analysis
    Chunpeng Zhao
    Gang Zhu
    Yu Wang
    Xinbao Wu
    Journal of Orthopaedic Surgery and Research, 17
  • [49] Air kerma versus fluoroscopy time as an indication of radiation dose in cardiac interventions
    Mercuri, M.
    Moran, G. R.
    Gauthier, L.
    Sheth, T.
    Velianou, J. L.
    Nataraian, M. K.
    CANADIAN JOURNAL OF CARDIOLOGY, 2007, 23 : 234C - 235C
  • [50] Pedicle screw placement in the thoracic spine: a randomized comparison study of computer-assisted navigation and conventional techniques
    Wu Han
    Gao Zhong-li
    Wang Jin-cheng
    Li Ying-pu
    Xia Peng
    Jiang Rui
    CHINESE JOURNAL OF TRAUMATOLOGY, 2010, 13 (04) : 201 - 205