Initial experience of mitral valve repair using the Carpentier-Edwards Physio II annuloplasty ring

被引:15
|
作者
Vohra, Hunaid A. [1 ]
Whistance, Robert N. [1 ]
Bezuska, Laurynas [1 ]
Livesey, Steve A. [1 ]
机构
[1] Southampton Univ Hosp NHS Trust, Wessex Cardiothorac Ctr, Southampton, Hants, England
关键词
Ring; Mitral; Annuloplasty; RECONSTRUCTION; DISEASE; MOTION; VALVULOPLASTY; REGURGITATION; DYNAMICS; OUTCOMES;
D O I
10.1016/j.ejcts.2010.10.004
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: The Carpentier-Edwards (CE) Physio II ring is a new prosthetic ring designed to accommodate the changing pathology seen in the spectrum of degenerative valve disease, particularly the larger anterior leaflet in repair of the Barlow valve. The aim of our study was to assess the safety and efficacy of mitral valve (MV) repair with the CE Physio II ring. Methods: Between April 2009 and March 2010, 100 patients underwent MV repair using the Physio II ring. Median age of patients was 70 years (54-85 years). The left ventricle (LV) was moderately (30-50%; n = 21) or severely (< 30%; n = 6) impaired in 27 patients preoperatively. Mitral regurgitation (MR) was due to degenerative disease in 87 patients (bileaflet prolapse: 34 patients). Mean logistic EuroSCORE was 10.07 +/- 8.9 and mean follow-up was 6.3 +/- 2.4 months. Results: Seventeen patients were non-elective (eight emergencies), five were re-do operations and 23 Maze +/- pulmonary vein isolations, and 14 tricuspid annuloplasties were performed. Neo-chordae were inserted in 50 patients (50%), whereas sliding annuloplasty was performed only in three patients. The median ring size was 32 mm (range 26-40 mm). On-table trans-oesophageal echocardiography (TOE) showed trivial/no MR in 87 patients, and mild in 13 patients, and there were no cases of systolic anterior motion (SAM). There were two re-explorations for bleeding and two patients required haemofiltration. There were no strokes or deep sternal wound infections (DSWIs). There was one hospital death (1%). At discharge, mean left ventricular end-diastolic (LVEDD) was 4.8 +/- 0.7 cm compared with 5.5 +/- 0.8 cm preoperatively (p = 0.03) and mean left ventricular endsystolic (LVESD) was 3.3 +/- 0.5 cm as compared with 3.6 +/- 0.8 preoperatively (p = 0.4). There was no MR in 87 patients and mild MR in 13 patients. The mean mitral valve area (MVA) was 2.8 +/- 0.7 cm(2). The mean systolic pulmonary artery pressure (SPAP) was 26.6 +/- 7.3 mmHg as compared with 50.9 +/- 17.2 mmHg preoperatively (p = 0.02). During follow-up, there were no thrombo-embolic complications, re-operation, endocarditis or deaths. Conclusions: MV repair with the Physio II ring has excellent short-term results, including subgroups with large anterior mitral valve leaflet (AMVL). Moreover, the dimensional ratios of the ring may allow it to be used for MV repair for degenerative MV disease, irrespective of anterior leaflet size. Crown Copyright (C) 2010 European Association for Cardio-Thoracic Surgery. Published by Elsevier B.V. All rights reserved.
引用
收藏
页码:881 / 885
页数:5
相关论文
共 50 条
  • [1] Mitral valve repair by Carpentier-Edwards physio annuloplasty ring.
    Kurosawa H.
    Nakano M.
    Kawase M.
    Kasegawa H.
    Nakano K.
    Eishi K.
    [J]. The Japanese Journal of Thoracic and Cardiovascular Surgery, 1999, 47 (8): : 355 - 360
  • [2] The Carpentier-Edwards Classic and Physio Annuloplasty Rings in Repair of Degenerative Mitral Valve Disease: A Retrospective Study
    Sidiki, Abubakari, I
    Faybushevich, Alexandr G.
    Lishchuk, Alexandr N.
    Koltunov, Alexandr N.
    Roshchina, Ekaterina A.
    [J]. JOURNAL OF THE SAUDI HEART ASSOCIATION, 2020, 32 (02) : 224 - 232
  • [3] Midterm results of mitral valve repair with the Carpentier-Edwards rigid ring
    Sakamoto Y.
    Hashimoto K.
    Okuyama H.
    Ishii S.
    Taguchi S.
    Kagawa H.
    Yoshitake M.
    [J]. The Japanese Journal of Thoracic and Cardiovascular Surgery, 2006, 54 (2) : 56 - 60
  • [4] Mitral valve annuloplasty: Comparison of the mural annulus shortening suture (MASS) with the Carpentier-Edwards prosthetic ring
    Detter, C
    Aybek, T
    Kupilik, N
    Fischlein, T
    Moritz, A
    [J]. JOURNAL OF HEART VALVE DISEASE, 2000, 9 (04): : 478 - 486
  • [5] Clinical Outcomes after Mitral Valve Repair with the Physio II Annuloplasty Ring
    Noack, Thilo
    Sieg, Franz
    Cuartas, Mateo Marin
    Spampinato, Ricardo
    Holzhey, David
    Seeburger, Joerg
    Borger, Michael Andrew
    [J]. THORACIC AND CARDIOVASCULAR SURGEON, 2022, 70 (02): : 100 - 105
  • [6] Comparison of the novel Medtentia double helix mitral annuloplasty system with the Carpentier-Edwards Physio annuloplasty ring: morphological and functional long-term outcome in a mitral valve insufficiency sheep model
    Konerding, Moritz A.
    Simpanen, Jarmo
    Ihlberg, Leo
    Aittomaeki, Juha
    Werkkala, Kalervo
    Delventhal, Vera
    Ackermann, Maximilian
    [J]. JOURNAL OF CARDIOTHORACIC SURGERY, 2013, 8
  • [7] Comparison of the novel Medtentia double helix mitral annuloplasty system with the Carpentier-Edwards Physio annuloplasty ring: morphological and functional long-term outcome in a mitral valve insufficiency sheep model
    Moritz A Konerding
    Jarmo Simpanen
    Leo Ihlberg
    Juha Aittomäki
    Kalervo Werkkala
    Vera Delventhal
    Maximilian Ackermann
    [J]. Journal of Cardiothoracic Surgery, 8
  • [8] De Vega annuloplasty and Carpentier-Edwards ring annuloplasty for secondary tricuspid regurgitation
    Matsuyama, K
    Matsumoto, M
    Sugita, T
    Nishizawa, J
    Tokuda, Y
    Matsuo, T
    Ueda, Y
    [J]. JOURNAL OF HEART VALVE DISEASE, 2001, 10 (04): : 520 - 524
  • [9] Long-term outcomes after mitral ring annuloplasty for degenerative mitral regurgitation: Duran ring versus carpentier-edwards ring
    Chung, Cheol Hyun
    Kim, Joon Bum
    Choo, Suk Jung
    Kim, Kyung Sun
    Song, Hyun
    Song, Meong Gun
    Song, Jae-Kwan
    Kang, Duk-Hyun
    Lee, Jae Won
    [J]. JOURNAL OF HEART VALVE DISEASE, 2007, 16 (05): : 536 - 544
  • [10] Long-term outcomes after mitral ring annuloplasty for degenerative mitral regurgitation: Duran ring versus carpentier-edwards ring - Meeting discussion
    Northrup, William F., III
    Duran, Carlos
    Kim, Joon Bum
    Okada, Yukikatsu
    [J]. JOURNAL OF HEART VALVE DISEASE, 2007, 16 (05): : 544 - 545