Objective The objective of this study was to evaluate the technical feasibility, safety, and efficacy of the new device Amplatzer duct occluder II additional sizes (ADO II AS) for closure of patent ductus arteriosus (PDA). Background: Transcatheter device closure is the standard care for PDA. Currently available technology is not designed for closure of small PDA in young children. Methods From April to July 2011, seven children (five females, median age 1.5 years, median weight 11.4 kg) underwent PDA closure with the ADO II AS. Six had isolated PDA, one had PDA associated with preductal coarctation. We evaluated early and short-term results. Results All but one PDA was closed via an antegrade approach. Mean fluoroscopy and procedural times were 8.0 +/- 3.9 and 49.8 +/- 27.9 min, respectively. No complications occurred. Immediate trivial residual shunt was present in one patient. In all devices, the retention disks laid flat against the walls of the pulmonary artery and aorta, without protrusion into the vessel lumen. The echocardiography performed after 24 hr did not show any residual shunt. At a median follow-up of 2 months, the PDAs were completely occluded without obstruction of the pulmonary arteries or aorta. Conclusions The new device ADO II AS was safely deployed with complete resolution of the PDA shunt. The lower profile and symmetry of this device allows for venous or arterial approach and smaller delivery catheter size. The ADO II AS might be a preferable alternative for closure of small-moderate PDAs. (C) 2012 Wiley Periodicals, Inc.
引用
收藏
页码:1169 / 1174
页数:6
相关论文
共 50 条
  • [1] Is the new Occlutech duct occluder an appropriate device for transcatheter closure of patent ductus arteriosus?
    Godart, Francois
    Houeijeh, Ali
    Domanski, Olivia
    Guillaume, Marie-Paule
    Brard, Melanie
    Lucron, Hugues
    [J]. INTERNATIONAL JOURNAL OF CARDIOLOGY, 2018, 261 : 54 - 57
  • [2] The Amplatzer Duct Occluder II: A New Device for Percutaneous Ductus Arteriosus Closure
    Saliba, Zakhia
    El-Rassi, Issam
    Abi-Warde, Marie-Therese
    Chehab, Ghassan
    Daou, Linda
    Khater, Daniele
    Gerbaka, Bernard
    Jebara, Victor
    [J]. JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2009, 22 (06) : 496 - 502
  • [3] Closure of Patent Ductus Arteriosus in Children, Small Infants, and Premature Babies with Amplatzer Duct Occluder II Additional Sizes: Multicenter Study
    Sungur, Metin
    Karakurt, Cemsit
    Ozbarlas, Nazan
    Baspinar, Osman
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 82 (02) : 245 - 252
  • [4] Closure of the patent ductus arteriosus with the Amplatzer Duct Occluder II: a clinical experience
    Karagoz, Tevfik
    Akin, Alper
    Ertugrul, Ilker
    Aykan, Hayrettin Hakan
    Alehan, Dursun
    Ozer, Sema
    Ozkutlu, Suheyla
    [J]. ACTA CARDIOLOGICA, 2012, 67 (06) : 675 - 680
  • [5] Transcatheter closure of patent ductus arteriosus using the AMPLATZER™ duct occluder II (ADO II)
    Gruenstein, Daniel H.
    Ebeid, Makram
    Radtke, Wolfgang
    Moore, Phillip
    Holzer, Ralf
    Justino, Henri
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2017, 89 (06) : 1118 - 1128
  • [6] Transcatheter closure of patent ductus arteriosus with the Amplatzer Duct Occluder.
    Saliba, Z
    Aggoun, Y
    Hausse, AO
    Acar, P
    Bonnet, D
    Fraisse, A
    Sidi, D
    Kachaner, J
    Bonhoeffer, P
    [J]. ARCHIVES DES MALADIES DU COEUR ET DES VAISSEAUX, 2000, 93 (05): : 533 - 538
  • [7] Transcatheter Closure of Patent Ductus Arteriosus in Children with the Occlutech Duct Occluder
    Bilici, Meki
    Demir, Fikri
    Akin, Alper
    Ture, Mehmet
    Balik, Hasan
    Kuyumcu, Mahir
    [J]. PEDIATRIC CARDIOLOGY, 2017, 38 (08) : 1598 - 1605
  • [8] Transcatheter Closure of Patent Ductus Arteriosus in Children with the Occlutech Duct Occluder
    Meki Bilici
    Fikri Demir
    Alper Akın
    Mehmet Türe
    Hasan Balık
    Mahir Kuyumcu
    [J]. Pediatric Cardiology, 2017, 38 : 1598 - 1605
  • [9] Transcatheter Closure of Large Sized Patent Ductus Arteriosus Using the Amplatzer Duct Occluder Device
    Shabbir, Muhammad
    Akhtar, Khurram
    Maadullah
    Ahmed, Waqar
    [J]. JCPSP-JOURNAL OF THE COLLEGE OF PHYSICIANS AND SURGEONS PAKISTAN, 2009, 19 (05): : 275 - 278
  • [10] Transcatheter Closure of a Patent Ductus Arteriosus Using a Piccolo Duct Occluder
    Deb, Pranjit
    Benerjee, Anindya
    Som, Tapas
    Barik Sr, Ramachandra
    [J]. CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (08)