Design and analysis of magnetorheological damper based ankle-foot prosthesis prototype

被引:6
|
作者
Negi, Sachin [1 ]
Sagar, Ujjwal [2 ]
Nautiyal, Vijay Kumar [3 ]
Sharma, Neeraj [1 ]
机构
[1] Indian Inst Technol BHU Varanasi, Sch Biomed Engn, Varanasi, Uttar Pradesh, India
[2] MKU Ltd, Electroopt Unit, Kanpur, Uttar Pradesh, India
[3] Nautiyal Orthot Prosthet Rehabil Clin, Dehra Dun, Uttarakhand, India
关键词
Electromyography; Ankle-foot prosthesis; Inertial measurement units; Magnetorheological damper;
D O I
10.1108/IR-04-2021-0084
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Purpose This paper aims to design and analyze a controlled magnetorheological damper-based ankle-foot prosthesis prototype. Design/methodology/approach The ankle-foot prostheses prototype is proposed using the lightweight three dimensional (3 D)-printed parts, MR damper and digital servomotor. Initially, the computer-aided design (CAD) model of the prosthetic foot, leaf spring, retention spring and the various connecting parts required to connect the pylon and damper actuator assemblies are designed using CAD software. Later, the fused deposition modeling 3 D printer-based technique prints a prosthetic foot and other connecting parts using Acrylonitrile Butadiene Styrene filament. The prototype consists of two control parts: the first part controls the MR actuator that absorbs the impacts during walking. The second part is the control of the electric actuator intended to generate the dorsiflexion and plantar flexion movements. Finally, the prototype is tested on a transtibial amputee under the supervision of a prosthetist. Findings The ANalysis SYStems software-based analysis has shown that the prosthetic foot has a factor of safety values between 4.7 and 8.7 for heel strike, mid-swing and toe-off; hence, it is safe from mechanical failure. The designed MR damper-based ankle-foot prosthesis prototype is tested on an amputee for a level-ground walk; he felt comfortable compared to his passive prosthesis. Originality/value The design of an MR damper-based prosthesis prototype offers a better dynamic range for locomotion than passive prostheses. It reduces the injuries and provides relief to the transtibial amputees.
引用
收藏
页码:240 / 248
页数:9
相关论文
共 50 条
  • [41] A Novel Design of Electro-hydraulic Driven Active Powered Ankle-Foot Prosthesis
    Huang, Qitao
    Li, Bowen
    Jia, Fei
    Wang, Peng
    INTELLIGENT ROBOTICS AND APPLICATIONS, ICIRA 2021, PT I, 2021, 13013 : 622 - 630
  • [42] DEVELOPMENT OF BIOMIMETIC POWERED ANKLE-FOOT PROSTHESIS WITH COMPLIANT JOINTS AND SEGMENTED FOOT
    Zhu, Jinying
    Yuan, Kebin
    Wang, Qining
    Wang, Long
    EMERGING TRENDS IN MOBILE ROBOTICS, 2010, : 1039 - 1046
  • [43] Powered Ankle-Foot Prosthesis Improves Walking Metabolic Economy
    Au, Samuel K.
    Weber, Jeff
    Herr, Hugh
    IEEE TRANSACTIONS ON ROBOTICS, 2009, 25 (01) : 51 - 66
  • [44] Implementation and Testing of Ankle-Foot Prosthesis With a New Compensated Controller
    Gao, Fei
    Liu, Yannan
    Liao, Wei-Hsin
    IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2019, 24 (04) : 1775 - 1784
  • [45] A PASSIVE TWO-DEGREE-OF-FREEDOM ANKLE-FOOT PROSTHESIS
    Elley, Colin
    Nelson, Carl A.
    2018 DESIGN OF MEDICAL DEVICES CONFERENCE, 2018,
  • [46] An Ankle-Foot Prosthesis Emulator Capable of Modulating Center of Pressure
    Chiu, Vincent L.
    Voloshina, Alexandra S.
    Collins, Steven H.
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2020, 67 (01) : 166 - 176
  • [47] Development of a Bimodal Ankle-Foot Prosthesis for Walking and Standing/Swaying
    Hansen, Andrew H.
    Nickel, Eric A.
    JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME, 2013, 7 (03):
  • [48] A Universal Ankle-Foot Prosthesis Emulator for Human Locomotion Experiments
    Caputo, Joshua M.
    Collins, Steven H.
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2014, 136 (03):
  • [49] Ankle-foot orthosis with an oil damper versus nonarticulated ankle-foot orthosis in the gait of patients with subacute stroke: a randomized controlled trial
    Yamamoto, Sumiko
    Motojima, Naoyuki
    Kobayashi, Yosuke
    Osada, Yuji
    Tanaka, Souji
    Daryabor, Aliyeh
    JOURNAL OF NEUROENGINEERING AND REHABILITATION, 2022, 19 (01)
  • [50] Biomechanics of the ankle-foot system during stair ambulation: Implications for design of advanced ankle-foot prostheses
    Sinitski, Emily H.
    Hansen, Andrew H.
    Wilken, Jason M.
    JOURNAL OF BIOMECHANICS, 2012, 45 (03) : 588 - 594