A modified TEW approach to scatter correction for In-111 and Tc-99m dual-isotope small-animal SPECT

被引:6
|
作者
Prior, Paul [1 ,2 ]
Timmins, Rachel [1 ,2 ]
Petryk, Julia [2 ]
Strydhorst, Jared [1 ,2 ]
Duan, Yin [3 ]
Wei, Lihui [3 ]
Wells, R. Glenn [1 ,2 ]
机构
[1] Carleton Univ, Dept Phys, Ottawa, ON K1S 5B6, Canada
[2] Ottawa Heart Inst, Div Cardiol, Ottawa, ON K1Y 4W7, Canada
[3] Nordion Inc, 447 March Rd, Ottawa, ON K2K 1X8, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
small animal SPECT/CT; cross talk correction; scatter correction; PHANTOM EXPERIMENTS; RECONSTRUCTION; INFECTION;
D O I
10.1118/1.4962469
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In dual-isotope (Tc-99m/In-111) small-animal single-photon emission computed tomography (SPECT), quantitative accuracy of Tc-99m activity measurements is degraded due to the detection of Compton-scattered photons in the Tc-99m photopeak window, which originate from the In-111 emissions (cross talk) and from the Tc-99m emission (self-scatter). The standard triple-energy window (TEW) estimates the total scatter (self-scatter and cross talk) using one scatter window on either side of the Tc-99m photopeak window, but the estimate is biased due to the presence of unscattered photons in the scatter windows. The authors present a modified TEW method to correct for total scatter that compensates for this bias and evaluate the method in phantoms and in vivo. Methods: The number of unscattered Tc-99m and In-111 photons present in each scatter-window projection is estimated based on the number of photons detected in the photopeak of each isotope, using the isotope-dependent energy resolution of the detector. The camera-head-specific energy resolutions for the 140 keV Tc-99m and 171 keV In-111 emissions were determined experimentally by separately sampling the energy spectra of each isotope. Each sampled spectrum was fit with a Linear + Gaussian function. The fitted Gaussian functions were integrated across each energy window to determine the proportion of unscattered photons from each emission detected in the scatter windows. The method was first tested and compared to the standard TEW in phantoms containing Tc-99m: In-111 activity ratios between 0.15 and 6.90. True activities were determined using a dose calibrator, and SPECT activities were estimated from CT-attenuation-corrected images with and without scatter-correction. The method was then tested in vivo in six rats using In-111-liposome and Tc-99m-tetrofosmin to generate cross talk in the area of the myocardium. The myocardium was manually segmented using the SPECT and CT images, and partial-volume correction was performed using a template-based approach. The rat heart was counted in a well-counter to determine the true activity. Results: In the phantoms without correction for Compton-scatter, Tc-99m activity quantification errors as high as 85% were observed. The standard TEW method quantified Tc-99m activity with an average accuracy of -9.0%+/- 0.7%, while the modified TEW was accurate within 5% of truth in phantoms with Tc-99m: In-111 activity ratios >= 0.52. Without scatter-correction, In-111 activity was quantified with an average accuracy of 4.1%, and there was no dependence of accuracy on the activity ratio. In rat myocardia, uncorrected images were overestimated by an average of 23%+/- 5%, and the standard TEW had an accuracy of -13.8%+/- 1.6%, while the modified TEW yielded an accuracy of -4.0%+/- 1.6%. Conclusions: Cross talk and self-scatter were shown to produce quantification errors in phantoms as well as in vivo. The standard TEW provided inaccurate results due to the inclusion of unscattered photons in the scatter windows. The modified TEW improved the scatter estimate and reduced the quantification errors in phantoms and in vivo. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:5503 / 5513
页数:11
相关论文
共 50 条
  • [21] WINDOW SELECTION FOR DUAL PHOTOPEAK WINDOW SCATTER CORRECTION IN TC-99M IMAGING
    DEVRIES, DJ
    KING, MA
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (06) : 2771 - 2778
  • [22] CARDIAC PHANTOM EVALUATION OF SIMULTANEOUSLY ACQUIRED DUAL-ISOTOPE REST TL-201 STRESS TC-99M SPECT IMAGES
    LOWE, VJ
    GREER, KL
    HANSON, MW
    JASZCZAK, RJ
    COLEMAN, RE
    JOURNAL OF NUCLEAR MEDICINE, 1993, 34 (11) : 1998 - 2006
  • [23] Evaluation of Monte Carlo-based compensation for scatter and crosstalk in simultaneous In-111/Tc-99m SPECT-CT imaging of infection
    Cervo, Morgan
    Moore, Stephen C.
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3804 - 3806
  • [24] Simultaneous Tc-99m PYP/Tl-201 dual-isotope SPECT myocardial imaging in patients with suspected cardiac amyloidosis
    Tamarappoo, Balaji
    Otaki, Yuka
    Manabe, Osamu
    Hyun, Mark
    Cantu, Stephanie
    Arnson, Yoav
    Gransar, Heidi
    Hayes, Sean W.
    Friedman, John D.
    Thomson, Louise
    Slomka, Piotr
    Dey, Damini
    Vescio, Robert
    Patel, Jignesh
    Berman, Daniel S.
    JOURNAL OF NUCLEAR CARDIOLOGY, 2020, 27 (01) : 28 - 37
  • [25] Simultaneous Tc-99m PYP/Tl-201 dual-isotope SPECT myocardial imaging in patients with suspected cardiac amyloidosis
    Balaji Tamarappoo
    Yuka Otaki
    Osamu Manabe
    Mark Hyun
    Stephanie Cantu
    Yoav Arnson
    Heidi Gransar
    Sean W. Hayes
    John D. Friedman
    Louise Thomson
    Piotr Slomka
    Damini Dey
    Robert Vescio
    Jignesh Patel
    Daniel S. Berman
    Journal of Nuclear Cardiology, 2020, 27 : 28 - 37
  • [26] Influence of spill-over for 99mTc images and the effect of scatter correction for dual-isotope simultaneous acquisition with 99mTc and 18F using small-animal SPECT-PET/CT system
    Shibutani, Takayuki
    Onoguchi, Masahisa
    Kanno, Takayuki
    Kinuya, Seigo
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2024, 47 (01) : 135 - 142
  • [27] Influence of spill-over for 99mTc images and the effect of scatter correction for dual-isotope simultaneous acquisition with 99mTc and 18F using small-animal SPECT-PET/CT system
    Takayuki Shibutani
    Masahisa Onoguchi
    Takayuki Kanno
    Seigo Kinuya
    Physical and Engineering Sciences in Medicine, 2024, 47 : 135 - 142
  • [28] Enhanced cross-talk correction technique for simultaneous dual-isotope imaging: A TL-201/Tc-99m myocardial perfusion SPECT dog study
    Knesaurek, K
    Machac, J
    MEDICAL PHYSICS, 1997, 24 (12) : 1914 - 1923
  • [29] Utility of crosstalk correction for accurate estimation of myocardial perfusion and viability in dual-isotope simultaneous TC-99M tetrofosmin/F-18 fluorodeoxyglucose SPECT.
    Fukuchi, K
    Katafuchi, T
    Fukushima, K
    Toba, M
    Oka, H
    Hayashida, K
    Ishida, Y
    JOURNAL OF NUCLEAR MEDICINE, 1999, 40 (05) : 177P - 177P
  • [30] Attenuation and scatter correction in brain SPECT with Tc-99m labeled receptor-specific radiopharmaceuticals
    Jang, S
    Mozley, PD
    Barraclough, ED
    1999 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1999, : 1569 - 1573